510(k) Report

510(k) · Cleared

Cleared

ICONIX XBraid TT with Needles, NanoTack TT Suture Anchor

K173074510(k) SpecialCleared Apr 4, 2018187-day reviewClass II21 CFR 888.3040

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker · MBI Fastener, Fixation, Nondegradable, Soft Tissue

Cleared on Apr 4, 2018 as substantially equivalent to K151201 (Stryker, 2015) and 1 other predicate after a 187-day review. Its family tree reaches back 10 generations to 1996. 5 later devices descend from it.

Decided
Apr 4, 2018
Review
187 days
Class
Class II
Product code
MBI
Recalls
None
Predicates
2

Its longest line of descent runs back 10 generations to K961555 (1996). 30 ancestors in all. 5 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2007: 6 other ancestors2012: 1 other ancestor2006: 2 other ancestors2004: 1 other ancestor2001: 3 other ancestors2005: 2 other ancestors2000: 2 other ancestors1998: 1 other ancestor1997: 1 other ancestor2023: 2 descendants cleared2024: 2 descendants cleared2025: 1 descendant clearedReference device: K162310 · XBraid TT Suture Tape · Stryker · 2016K961555 · Acufex Tag Bioabsorbable Anchors · Acufex Microsurgical, Inc. · 1996K031685 · Tag Rod Bioraptor Suture Anchor 7209317, 7209318 · Smith & Nephew, Inc. · 2003 · recalledK053344 · Bioraptor Suture Anchor, Twinfix 2.8 MM Suture Anchor, Twinfix 3.5 MM Suture Anchor · Smith & Nephew, Inc. · 2006K061863 · Arthrex Corkscrew Suture Anchor, Tak, and Pushlock Family · Arthrex, Inc. · 2006K071177 · Arthrex Bio- Composite Suture Anchors: Bio-Composite Pushlock, Bio-Composite Tak, Bio-Composite Corkscrew · Arthrex, Inc. · 2007K082810 · Arthrex Bio-Composite Suture Anchors: Arthrex Bio-Composite Pushlock, Bio-Composite Tak and Bio-Composite Corkscrew · Arthrex, Inc. · 2009K101823 · Arthrex Swivelock Anchors · Arthrex, Inc. · 2011 · recalledK110545 · Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor (Ngftsa PK) · Smith & Nephew, Inc. · 2011K133671 · Iconix All Suture Anchors · Stryker Corp. · 2014K151201 · Stryker ICONIX with Needles · Stryker · 20151995200020052010201520202025
  1. Root · 10 generations backK961555 Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc. · 1996
  2. 6 generations between
  3. 3 generations backK110545 Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor…Smith & Nephew, Inc. · 2011
  4. 2 generations backK133671 Iconix All Suture AnchorsStryker Corp. · 2014
  5. 1 generation backK151201 Stryker ICONIX with NeedlesStryker · 2015
  6. This deviceK173074 ICONIX XBraid TT with Needles, NanoTack TT Suture AnchorStryker · 2018
  7. Descendant · cites it directly · 3 of its ownK231078 All-Suture AnchorRiverpoint Medical, LLC · 2023
  8. Descendant · 2 generations on · 1 of its ownK233468 Iconix AnchorRiverpoint Medical · 2023
  9. Descendant · 3 generations on · 0 of its ownK252144 Iconix Speed Anchor; Iconix Speed HA+ AnchorRiverpoint Medical · 2025
  10. Descendant · 2 generations on · 0 of its ownK241577 Iconix® AnchorRiverpoint Medical · 2024
  11. Descendant · cites it directly · 0 of its ownK240407 ICONIX All-Suture Anchor; ICONIX TT All-Suture Anchor; ICONIX with…Stryker Endoscopy · 2024
Show all 10 generations
  1. Root · 10 generations backK961555 Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc. · 1996
  2. 9 generations backK031685 Tag Rod Bioraptor Suture Anchor 7209317, 7209318Smith & Nephew, Inc. · 2003 · Recalled
  3. 8 generations backK053344 Bioraptor Suture Anchor, Twinfix 2.8 MM Suture Anchor, Twinfix 3.5…Smith & Nephew, Inc. · 2006
  4. 7 generations backK061863 Arthrex Corkscrew Suture Anchor, Tak, and Pushlock FamilyArthrex, Inc. · 2006
  5. 6 generations backK071177 Arthrex Bio- Composite Suture Anchors: Bio-Composite Pushlock…Arthrex, Inc. · 2007
  6. 5 generations backK082810 Arthrex Bio-Composite Suture Anchors: Arthrex Bio-Composite…Arthrex, Inc. · 2009
  7. 4 generations backK101823 Arthrex Swivelock AnchorsArthrex, Inc. · 2011 · Recalled
  8. 3 generations backK110545 Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor…Smith & Nephew, Inc. · 2011
  9. 2 generations backK133671 Iconix All Suture AnchorsStryker Corp. · 2014
  10. 1 generation backK151201 Stryker ICONIX with NeedlesStryker · 2015
  11. This deviceK173074 ICONIX XBraid TT with Needles, NanoTack TT Suture AnchorStryker · 2018

Most influential descendants

Showing 5 of 5 · See all
  • K231078All-Suture AnchorRiverpoint Medical, LLC · 2023 · cites it directly3descendants
  • K233468Iconix AnchorRiverpoint Medical · 2023 · 2 generations on1descendant
  • K252144Iconix Speed Anchor; Iconix Speed HA+ AnchorRiverpoint Medical · 2025 · 3 generations on0descendants
  • K241577Iconix® AnchorRiverpoint Medical · 2024 · 2 generations on0descendants
  • K240407ICONIX All-Suture Anchor; ICONIX TT All-Suture Anchor; ICONIX with Needles All-Suture…Stryker Endoscopy · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K961555Acufex Tag Bioabsorbable AnchorsAcufex Microsurgical, Inc.1996—
−9K031685Tag Rod Bioraptor Suture Anchor 7209317, 7209318Smith & Nephew, Inc.20035
−8K053344Bioraptor Suture Anchor, Twinfix 2.8 MM Suture Anchor, Twinfix 3.5 MM Suture…Smith & Nephew, Inc.2006—
−7K061863Arthrex Corkscrew Suture Anchor, Tak, and Pushlock FamilyArthrex, Inc.2006—
−6K071177Arthrex Bio- Composite Suture Anchors: Bio-Composite Pushlock, Bio-Composite…Arthrex, Inc.2007—
−5K082810Arthrex Bio-Composite Suture Anchors: Arthrex Bio-Composite Pushlock…Arthrex, Inc.2009—
−4K101823Arthrex Swivelock AnchorsArthrex, Inc.20116
−3K110545Smith & Nephew Next Generation Fully Threaded Peek Suture Anchor (Ngftsa PK)Smith & Nephew, Inc.2011—
−2K133671Iconix All Suture AnchorsStryker Corp.2014—
−1K151201Stryker ICONIX with NeedlesStryker2015—
0 (this device)K173074ICONIX XBraid TT with Needles, NanoTack TT Suture AnchorStryker2018—
+1K231078All-Suture AnchorRiverpoint Medical, LLC2023—
+2K233468Iconix AnchorRiverpoint Medical2023—
+3K252144Iconix Speed Anchor; Iconix Speed HA+ AnchorRiverpoint Medical2025—
+2K241577Iconix® AnchorRiverpoint Medical2024—
+1K240407ICONIX All-Suture Anchor; ICONIX TT All-Suture Anchor; ICONIX with Needles…Stryker Endoscopy2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
200020
200130
200410
200520
200620
200760
201210
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202320
202420
202510
FIG. 1 — Lineage of K173074, 1996–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBI device

  1. #1 K203268 Arthrex FiberTak Suture Anchorscore 21.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K151201Stryker ICONIX with Needles
Summary page 5
…(21 CFR 888.3040) Regulatory Class: II Product Code: MBI III. PREDICATE AND REFERENCE DEVICES Predicate Device A: Stryker ICONIX with Needles, K151201 This predicate has not been subject to a design-related recall. Predicate Device B: NanoTack Suture Anchor 1…
Stryker2015—
*K131769Cinchlock Knotless Suture Anchor, Nanotack Suture Anchor 1.4MM
Summary page 5
…yker ICONIX with Needles, K151201 This predicate has not been subject to a design-related recall. Predicate Device B: NanoTack Suture Anchor 1.4 mm, K131769 This predicate has not been subject to a design-related recall. ICONIX XBraid TT with Needles an…
Pivot Medical, Inc.20132
†K162310XBraid TT Suture TapeStryker2016—

Also cleared against K151201: 2 other devices →

What descends from it? (5)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231078All-Suture AnchorRiverpoint Medical, LLC20233
K240407ICONIX All-Suture Anchor; ICONIX TT All-Suture Anchor; ICONIX with Needles All-Suture…Stryker Endoscopy20240

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 6 of the 30 ancestors have a recall of any class.
  • Of its descendants, K231078 (Riverpoint Medical, LLC, 2023) has gone furthest, with 3 descendants of its own.
  • 2 other devices also cite its predicate K151201.

How does its category compare?

All MBI devices

MBI Fastener, Fixation, Nondegradable, Soft Tissue: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · MBI

27in 2025

2007: 162008: 102009: 192010: 172011: 302012: 242013: 312014: 232015: 232016: 172017: 172018: 162019: 312020: 352021: 182022: 192023: 522024: 222025: 272026 (partial year): 30
2007high 52 (2023) · low 10 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBI

58in 2025

2007: 80 days median2008: 69 days median2009: 82 days median2010: 76 days median2011: 75 days median2012: 94 days median2013: 100 days median2014: 101 days median2015: 85 days median2016: 94 days median2017: 70 days median2018: 91 days median2019: 90 days median2020: 83 days median2021: 107 days median2022: 64 days median2023: 57 days median2024: 30 days median2025: 58 days median2026 (partial year): 63 days median
2007high 107 (2021) · low 30 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBI

1in 2025

2007: 242008: 62009: 22010: 32011: 232012: 62013: 22017: 42018: 22020: 12023: 12024: 12025: 1
2007high 24 (2007) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBI

3,826in 2025

2007: 2402008: 3382009: 3552010: 2912011: 3282012: 3712013: 4422014: 5042015: 5072016: 7952017: 6752018: 1,0722019: 1,3032020: 1,4542021: 1,8902022: 1,5912023: 2,5032024: 3,7792025: 3,826
2007high 3,826 (2025) · low 240 (2007)2025

MAUDE reports naming product code MBI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007168024240
200810696338
200919822355
201017763291
2011307523328
201224946371
2013311002442
2014231010504
201523850507
201617940795
201717704675
2018169121,072
2019319001,303
2020358311,454
20211810701,890
2022196401,591
2023525712,503
2024223013,779
2025275813,826
202630630—

What is it indicated for?

The Stryker ICONIX XBraid TT with Needles devices are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, wrist, elbow, and shoulder. See indications below. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair…

Read the full statement (261 words)

The Stryker ICONIX XBraid TT with Needles devices are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, wrist, elbow, and shoulder. See indications below.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

Hand/Wrist: Scaphulolunate Ligament Reconstruction, Carpal Ligament Reconstruction,

Repair/Reconstruction of Collateral Ligaments, Repair of Flexor and Extenser Tendons at the PIP, DIP, and MCP Joint for All Digits, Digital Tendon Repair

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-Foot Reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, [liotibial Band Tenodesis

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair

The NanoTack TT Suture Anchor is intended for the fixation of soft tissue to bone in the hip, shoulder, foot/ankle, hand/wrist, elbow, and knee in the following procedures:

Hip: Hip capsule repair, acetabular labrum reattachment

Shoulder: Capsular stabilization (Bankart repair, anterior shoulder instability, SLAP lesion repair, capsular shifts or capsulolabral reconstructions), acromioclavicular separation repair, deitoid repair, rotator cuff tear repair, biceps tenodesis

Foot and Ankle: Hallux valgus repair, Medial or lateral instability repair/reconstruction, Achilles tendon repair/reconstruction, midfoot reconstruction, Metatarsal ligament/tendon repair/reconstructions, bunionectomy

Elbow, Wrist, and Hand: Biceps tendon reattachment, ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair

Knee: Extra-capsular repair {(medial collateral ligament, lateral collateral ligament, posterior oblique ligament), patellar realignment and tendon repairs (vastus medialis obliquous advancement), lliotibial band tenodesis

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173074: ICONIX XBraid TT with Needles, NanoTack TT Suture Anchor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173074

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026