510(k) Report

510(k) · Cleared

Cleared

Faxitron VisionCT

K173309510(k) TraditionalCleared May 9, 2018203-day reviewClass II21 CFR 892.1680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Faxitron Bioptics, LLC · MWP Cabinet, X-Ray System

Cleared on May 9, 2018 as substantially equivalent to K170786 (Faxitron Bioptics, LLC, 2017) and 2 other predicates after a 203-day review. Its family tree reaches back 7 generations to 2002. 1 later device descends from it.

Decided
May 9, 2018
Review
203 days
Class
Class II
Product code
MWP
Recalls
None
Predicates
3

Its longest line of descent runs back 7 generations to K021113 (2002). 13 ancestors in all, the oldest K852634 (1985). 1 later device descends from it, 1 citing it directly.

  1. Root · 7 generations backK021113 MammopathFischer Imaging Corp. · 2002
  2. 6 generations backK052433 Pixarray 100 Digital Specimen Radiography (DSR) SystemBioptics, Inc. · 2005
  3. 5 generations backK071233 Xpert 40 Specimen Radiography SystemKub Technologies, Inc. · 2007
  4. 4 generations backK083510 Kubtec Xpert 20Kub Technologies, Inc. · 2009
  5. 3 generations backK111508 Trident Specimen Radiography System, Model : RC Tri-00001Hologic, Inc. · 2011
  6. 2 generations backK122428 Faxitron Specien Radiography SystemFaxitron Bioptics, LLC · 2012
  7. 1 generation backK170786 Faxitron Bioptics Speciemen Radiography SystemFaxitron Bioptics, LLC · 2017
  8. This deviceK173309 Faxitron VisionCTFaxitron Bioptics, LLC · 2018
  9. Descendant · cites it directly · 0 of its ownK192939 Volumetric Specimen ImagerClarix Imaging · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K192939Volumetric Specimen ImagerClarix Imaging · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K021113MammopathFischer Imaging Corp.2002—
−6K052433Pixarray 100 Digital Specimen Radiography (DSR) SystemBioptics, Inc.2005—
−5K071233Xpert 40 Specimen Radiography SystemKub Technologies, Inc.2007—
−4K083510Kubtec Xpert 20Kub Technologies, Inc.2009—
−3K111508Trident Specimen Radiography System, Model : RC Tri-00001Hologic, Inc.2011—
−2K122428Faxitron Specien Radiography SystemFaxitron Bioptics, LLC2012—
−1K170786Faxitron Bioptics Speciemen Radiography SystemFaxitron Bioptics, LLC2017—
0 (this device)K173309Faxitron VisionCTFaxitron Bioptics, LLC2018—
+1K192939Volumetric Specimen ImagerClarix Imaging2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
199510
200610
200710
200910
201610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K173309, 2002–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MWP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MWP device

  1. #1 K202713 TrueView 100 Proscore 7.9 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170786Faxitron Bioptics Speciemen Radiography System
Summary page 5
…required by 21 CFR 807.92(a)(3) The Faxitron VisionCT is substantially equivalent to following predicate devices: A. Faxitron VersaVision Specimen Radiography System (K170786) B. Faxitron BioVision Plus Specimen Radiography System (K153583) C. XPERT 40 Specimen Radiography System (K071233)…
Faxitron Bioptics, LLC2017—
*K153583BioVision Plus
Summary page 5
…ly equivalent to following predicate devices: A. Faxitron VersaVision Specimen Radiography System (K170786) B. Faxitron BioVision Plus Specimen Radiography System (K153583) C. XPERT 40 Specimen Radiography System (K071233) Page 1 of 5 DEVI…
Faxitron Bioptics, LLC2016—
*K071233Xpert 40 Specimen Radiography System
Summary page 5
…axitron VersaVision Specimen Radiography System (K170786) B. Faxitron BioVision Plus Specimen Radiography System (K153583) C. XPERT 40 Specimen Radiography System (K071233) Page 1 of 5 DEVICE DESCRIPTION - as required by 21 CFR 807.92(a)(4) T…
Kub Technologies, Inc.2007—
†K160140DENTRIa seriesHdx Will Corp.2016—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192939Volumetric Specimen ImagerClarix Imaging20190

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 1.7 years on median.
  • None of the 13 ancestors has a recall.
  • Of its descendants, K192939 (Clarix Imaging, 2019) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MWP devices

MWP Cabinet, X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · MWP

2in 2023

2007: 22008: 12009: 12010: 12011: 12012: 12016: 12017: 12018: 12019: 52020: 22021: 42023: 2
2007high 5 (2019) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MWP

154in 2021

2019: 104 days median2021: 154 days median
2007high 154 (2021) · low 104 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MWP

No recall names MWP since 2007.

Adverse event reports per year · MWP

1in 2025

2018: 12021: 32022: 12023: 32024: 12025: 1
2007high 3 (2021) · low 0 (2007)2025

MAUDE reports naming product code MWP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20081—00
20091—00
20101—00
20111—00
20121—00
20130—00
20140—00
20150—00
20161—00
20171—00
20181—01
2019510400
20202—00
2021415403
20220—01
20232—03
20240—01
20250—01
20260—0—

What is it indicated for?

The Faxitron VisionCT is a Cabinet x-ray system that is used to provide two and three dimensional digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173309: Faxitron VisionCT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173309

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026