510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Mammomat Revelation

K173408510(k) TraditionalCleared Mar 21, 2018140-day reviewClass II21 CFR 892.1715Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medi Cal Solutions, Inc. · MUE Full Field Digital, System, X-Ray, Mammographic

Cleared on Mar 21, 2018 as substantially equivalent to K123520 (Siemens Medical Solutions USA, Inc., 2013) and 2 other predicates after a 140-day review. Its family tree reaches back 5 generations to 1995. 6 later devices descend from it. It has been recalled 2 times.

Decided
Mar 21, 2018
Review
140 days
Class
Class II
Product code
MUE
Recalls
2
Predicates
3

Its longest line of descent runs back 5 generations to K941191 (1995). 13 ancestors in all, the oldest K921962 (1992). 6 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 2 other ancestors2011: 1 other ancestor2004: 2 other ancestors2000: 1 other ancestor2003: 1 other ancestor1992: 1 other ancestor2019: 1 descendant cleared2020: 1 descendant cleared2024: 1 descendant cleared2021: 2 descendants cleared2026: 1 descendant clearedK941191 · Real Time Stereotix · GE Medical Systems · 1995 · recalledK040125 · Senographic Stereo · Ge Medical Systems, Inc. · 2004K071542 · Digital Stereoloc II · Lorad, A Hologic Co. · 2007K103512 · Affirm Breast Biopsy Guidance System · Hologic, Inc. · 2011K122836 · Affirm Breast Biopsy Guidance System · Lorad, A Hologic Co. · 20131990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors3 generations back: 1 other ancestor4 generations back: 1 other ancestor5 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K122836 · Affirm Breast Biopsy Guidance System · Lorad, A Hologic Co. · 2013K103512 · Affirm Breast Biopsy Guidance System · Hologic, Inc. · 2011K071542 · Digital Stereoloc II · Lorad, A Hologic Co. · 2007K040125 · Senographic Stereo · Ge Medical Systems, Inc. · 2004K941191 · Real Time Stereotix · GE Medical Systems · 1995 · recalled−50
  1. Root · 5 generations backK941191 Real Time StereotixGE Medical Systems · 1995 · Recalled
  2. 4 generations backK040125 Senographic StereoGe Medical Systems, Inc. · 2004
  3. 3 generations backK071542 Digital Stereoloc IILorad, A Hologic Co. · 2007
  4. 2 generations backK103512 Affirm Breast Biopsy Guidance SystemHologic, Inc. · 2011
  5. 1 generation backK122836 Affirm Breast Biopsy Guidance SystemLorad, A Hologic Co. · 2013
  6. This deviceK173408 Mammomat RevelationSiemens Medi Cal Solutions, Inc. · 2018
  7. Descendant · cites it directly · 4 of its ownK193166 MAMMOMAT RevelationSiemens Medical Solutions USA, Inc. · 2020 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK233539 MAMMOMAT B.brilliantSiemens Medical Solutions USA, Inc. · 2024
  9. Descendant · 3 generations on · 0 of its ownK253678 MAMMOMAT B.brilliant (11372950)Siemens Medical Solutions USA, Inc. · 2026
  10. Descendant · 2 generations on · 0 of its ownK210955 Kubtec Mozart Supra (XPERT 84) Radiography SystemKub Technologies, Inc. · 2021
  11. Descendant · 2 generations on · 0 of its ownK210957 Kubec Xpert 80 Radiography SystemKub Technologies, Inc. · 2021

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2019: 12020: 12021: 22024: 12026: 1
2019high 2 (2021) · low 0 (2022)2026

Most influential descendants

Showing 5 of 6 · See all
  • K193166MAMMOMAT RevelationSiemens Medical Solutions USA, Inc. · 2020 · cites it directly4descendants
  • K233539MAMMOMAT B.brilliantSiemens Medical Solutions USA, Inc. · 2024 · 2 generations on1descendant
  • K253678MAMMOMAT B.brilliant (11372950)Siemens Medical Solutions USA, Inc. · 2026 · 3 generations on0descendants
  • K210955Kubtec Mozart Supra (XPERT 84) Radiography SystemKub Technologies, Inc. · 2021 · 2 generations on0descendants
  • K210957Kubec Xpert 80 Radiography SystemKub Technologies, Inc. · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K941191Real Time StereotixGE Medical Systems19952
−4K040125Senographic StereoGe Medical Systems, Inc.2004—
−3K071542Digital Stereoloc IILorad, A Hologic Co.2007—
−2K103512Affirm Breast Biopsy Guidance SystemHologic, Inc.2011—
−1K122836Affirm Breast Biopsy Guidance SystemLorad, A Hologic Co.2013—
0 (this device)K173408Mammomat RevelationSiemens Medi Cal Solutions, Inc.2018—
+1K193166MAMMOMAT RevelationSiemens Medical Solutions USA, Inc.20201
+2K233539MAMMOMAT B.brilliantSiemens Medical Solutions USA, Inc.2024—
+3K253678MAMMOMAT B.brilliant (11372950)Siemens Medical Solutions USA, Inc.2026—
+2K210955Kubtec Mozart Supra (XPERT 84) Radiography SystemKub Technologies, Inc.2021—
+2K210957Kubec Xpert 80 Radiography SystemKub Technologies, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
200010
200310
200420
201110
201320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
202120
202410
202610
FIG. 1 — Lineage of K173408, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUE device

  1. #1 K253678 MAMMOMAT B.brilliant (11372950)score 13.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K123520Mammomat Inspiration Prime
Summary page 5
…4. Legally Marketed Predicate Devices Primary predicate: Trade Name: MAMMOMAT Inspiration Prime 510(k) #: K123520 Classification Name: Full Field Digital, System, X-Ray Mammographic Classification Panel: Radiolog…
Siemens Medical Solutions USA, Inc.20131
*K122836Affirm Breast Biopsy Guidance System
Summary page 5
…t Code: MUE Secondary predicates: Trade Name: AFFIRM BREAST BIOPSY GUIDANCE SYSTEM 510(k) #: K122836 Classification Name: System, X-ray, Mammographic Classification Panel: Radiology Classifica…
Lorad, A Hologic Co.2013—
*K123873Contrast Enhanced Digital Mammography
Summary page 5
…Class II Product Code: IZH Trade Name: CONTRAST ENHANCED DIGITAL MAMMOGRAPHY 510(k) #: K123873 Classification Name: Full Field Digital, System, X-Ray Mammographic Classification Panel: Radiolog…
Hologic, Inc.20131

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193166MAMMOMAT RevelationSiemens Medical Solutions USA, Inc.20204
K190809Sample Imaging for Senographe PristinaGE Healthcare20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2020Class IIZ-2388-2020Device DesignIf an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to…
Jul 23, 2019Class IIZ-0160-2020Software designAn error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 3.5 years on median.
  • 5 of the 13 ancestors have a recall of any class.
  • Of its descendants, K193166 (Siemens Medical Solutions USA, Inc., 2020) has gone furthest, with 4 descendants of its own.

How does its category compare?

All MUE devices

MUE Full Field Digital, System, X-Ray, Mammographic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · MUE

3in 2025

2007: 62008: 52009: 82010: 102011: 102012: 42013: 72014: 12015: 12016: 12017: 82018: 22019: 22020: 42021: 42022: 32023: 12024: 12025: 32026 (partial year): 4
2007high 10 (2010) · low 1 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUE

255in 2025

2011: 192 days median2012: 123 days median2013: 207 days median2017: 95 days median2020: 188 days median2021: 164 days median2022: 404 days median2025: 255 days median2026 (partial year): 185 days median
2007high 404 (2022) · low 95 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUE

4in 2025

2008: 12009: 12010: 22011: 52014: 22017: 62018: 12019: 12023: 12025: 42026 (partial year): 1
2007high 6 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUE

40in 2025

2008: 42009: 32010: 22011: 102012: 62013: 112014: 72015: 62016: 192017: 1152018: 1452019: 802020: 492021: 252022: 362023: 292024: 332025: 40
2007high 145 (2018) · low 0 (2007)2025

MAUDE reports naming product code MUE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076—00
20085—14
20098—13
201010—22
201110192510
2012412306
20137207011
20141—27
20151—06
20161—019
20178956115
20182—1145
20192—180
20204188049
20214164025
20223404036
20231—129
20241—033
20253255440
202641851—

What is it indicated for?

The MAMMOMAT Revelation is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals. The Mammography images can be interpreted by either hard copy film or soft copy workstation.

With Biopsy Option:

The InSpect feature for MAMMOMAT Revelation with HD Biopsy options is intended to provide digital X-ray images of core biopsy specimens in order to allow rapid verification that the correct tissue has been excised with the biopsy procedure.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173408: Mammomat Revelation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173408

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026