Recalled · Class II · 2025Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medi Cal Solutions, Inc. · OWB Interventional Fluoroscopic X-Ray System
Cleared on Apr 2, 2018 as substantially equivalent to K111292 (Siemens Medical Solutions USA, Inc., 2011) and 3 other predicates after a 129-day review. Its family tree reaches back 9 generations to 1980. 1 later device descends from it. It has been recalled 23 times.
- Decided
- Apr 2, 2018
- Review
- 129 days
- Class
- Class II
- Product code
- OWB
- Recalls
- 23
- Predicates
- 4
Its longest line of descent runs back 9 generations to K801623 (1980). 61 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 5 generations between
- 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
- 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
- 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
- This deviceK173639 Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc. · 2018
- Descendant · cites it directly · 0 of its ownK250660 LUMINOS Q.namix T; LUMINOS Q.namix RSiemens Medical Solutions · 2025 · Recalled
Show all 9 generations
- Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 8 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
- 7 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
- 6 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
- 5 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
- 4 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
- 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
- 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
- 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
- This deviceK173639 Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom RaxSiemens Medi Cal Solutions, Inc. · 2018
Most influential descendants
Showing 1 of 1 · See all- K250660LUMINOS Q.namix T; LUMINOS Q.namix RSiemens Medical Solutions · 2025 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K801623 | Cavitron Ultrasonic Surg. Aspir. Cusa | Cavitron Corp. | 1980 | — |
| −8 | K962952 | Olympus Ultrasonic Surgical System (Usu) | Olympus Optical, Co. | 1996 | — |
| −7 | K972114 | Olympus Sonosurg System | The Olympus Optical Co. | 1998 | — |
| −6 | K981993 | Olympus Integrated Endosurgery System Endoalpha | The Olympus Optical Co. | 1998 | — |
| −5 | K994231 | Siemens Integrated Operating System (Sios) | Siemens Medical Corp. | 2000 | — |
| −4 | K052202 | Axiom Artis - Modular Angiographic System | Siemens Medical Solutions USA, Inc. | 2006 | 29 |
| −3 | K061054 | Axiom Aristos FX Plus | Siemens Medical Solutions USA, Inc. | 2006 | 3 |
| −2 | K062623 | Axiom Luminos DRF | Siemens Medical Solutions USA, Inc. | 2007 | 20 |
| −1 | K101491 | Uroskop Omnia | Siemens Medical Solutions USA, Inc. | 2010 | 12 |
| 0 (this device) | K173639 | Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax | Siemens Medi Cal Solutions, Inc. | 2018 | — |
| +1 | K250660 | LUMINOS Q.namix T; LUMINOS Q.namix R | Siemens Medical Solutions | 2025 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1986 | 1 | 0 |
| 1988 | 2 | 0 |
| 1989 | 1 | 0 |
| 1991 | 4 | 0 |
| 1992 | 2 | 0 |
| 1993 | 3 | 0 |
| 1994 | 1 | 0 |
| 1995 | 12 | 0 |
| 1996 | 4 | 0 |
| 1997 | 4 | 0 |
| 1999 | 3 | 0 |
| 2000 | 3 | 0 |
| 2001 | 2 | 0 |
| 2002 | 4 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
| 2011 | 1 | 0 |
| 2015 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWB device
- #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K111292 | Luminos AgileSummary page 8…Trade Name Luminos AXIOM Uroskop Multitom Agile Luminos dRF Omnia Rax 510(k) K111292 K062623 K101491 K152928 Number Device System, x- interventional System, x-ray,… | 2011 |
| * | K062623 | Axiom Luminos DRFSummary page 8…Luminos AXIOM Uroskop Multitom Agile Luminos dRF Omnia Rax 510(k) K111292 K062623 K101491 K152928 Number Device System, x- interventional System, x-ray, intervent… | 2007 |
| * | K101491 | Uroskop OmniaSummary page 8…XIOM Uroskop Multitom Agile Luminos dRF Omnia Rax 510(k) K111292 K062623 K101491 K152928 Number Device System, x- interventional System, x-ray, interventional Class… | 2010 |
| * | K152928 | Multitom RaxSummary page 8…p Multitom Agile Luminos dRF Omnia Rax 510(k) K111292 K062623 K101491 K152928 Number Device System, x- interventional System, x-ray, interventional Classification ray,… | 2015 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K250660 | LUMINOS Q.namix T; LUMINOS Q.namix R | 2025 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 5, 2025 | Class II | Z-0999-2026 | Radiation Control for Health and Safety Act | A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the… |
| Jun 4, 2025 | Class II | Z-1982-2025 | Device Design | A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire. |
| Jun 4, 2025 | Class II | Z-1981-2025 | Device Design | A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire. |
| Jul 2, 2024 | Class II | Z-2526-2024 | Under Investigation by firm | The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays |
| Jul 2, 2024 | Class II | Z-2527-2024 | Under Investigation by firm | The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays |
| Jul 2, 2024 | Class II | Z-2525-2024 | Under Investigation by firm | The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays |
| Dec 21, 2022 | Class II | Z-1168-2023 | Software design | Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values |
| Dec 21, 2022 | Class II | Z-1165-2023 | Software design | Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values |
| Dec 21, 2022 | Class II | Z-1167-2023 | Software design | Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values |
| Dec 9, 2022 | Class II | Z-1002-2023 | Process control | There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical applications ranging from… |
| Nov 23, 2022 | Class II | Z-1481-2023 | Radiation Control for Health and Safety Act | Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC… |
| Nov 23, 2022 | Class II | Z-1480-2023 | Radiation Control for Health and Safety Act | Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC… |
| Nov 23, 2022 | Class II | Z-1479-2023 | Radiation Control for Health and Safety Act | Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC… |
| Sep 9, 2020 | Class II | Z-0806-2021 | Device Design | After the March 2019 corrective action on the monitor support arm, there is a residual risk of the support arm becoming dislodged in cases when the monitor support arm is used intensely over a long period of time. |
| Jan 8, 2020 | Class II | Z-1337-2020 | Software design | A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis. |
| Jan 8, 2020 | Class II | Z-1333-2020 | Use error | Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function. |
| Jan 8, 2020 | Class II | Z-1332-2020 | Use error | Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function. |
| Jan 8, 2020 | Class II | Z-2756-2020 | Device Design | Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or… |
| Jan 8, 2020 | Class II | Z-2757-2020 | Device Design | Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or… |
| Jan 8, 2020 | Class II | Z-1331-2020 | Use error | Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function. |
| Jan 8, 2020 | Class II | Z-1338-2020 | Software design | A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis. |
| Jan 8, 2020 | Class II | Z-2755-2020 | Device Design | Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or… |
| Feb 8, 2019 | Class II | Z-0985-2019 | Device Design | The swivel arm holding the monitors may become dislodged from the carrying arm. |
The 5 latest of 23 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 38 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
- 23 of the 61 ancestors have a recall of any class.
- Of its descendants, K250660 (Siemens Medical Solutions, 2025) has gone furthest, with 0 descendants of its own.
How does its category compare?
All OWB devicesOWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.
FDA decisions per year · OWB
22in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWB
148in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWB
42so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWB
14,008in 2025
MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 52 | 0 | 1 |
| 2008 | 4 | 135 | 1 | 0 |
| 2009 | 9 | 99 | 0 | 0 |
| 2010 | 7 | 66 | 0 | 0 |
| 2011 | 6 | 79 | 0 | 1 |
| 2012 | 6 | 37 | 0 | 1 |
| 2013 | 11 | 119 | 10 | 27 |
| 2014 | 15 | 158 | 9 | 103 |
| 2015 | 11 | 66 | 35 | 2,134 |
| 2016 | 16 | 103 | 13 | 4,074 |
| 2017 | 20 | 76 | 13 | 1,604 |
| 2018 | 23 | 54 | 6 | 1,449 |
| 2019 | 9 | 174 | 7 | 1,558 |
| 2020 | 16 | 53 | 8 | 1,581 |
| 2021 | 24 | 60 | 31 | 1,981 |
| 2022 | 21 | 90 | 24 | 2,932 |
| 2023 | 12 | 152 | 22 | 7,604 |
| 2024 | 13 | 126 | 11 | 7,942 |
| 2025 | 22 | 148 | 30 | 14,008 |
| 2026 | 15 | 221 | 42 | — |
What is it indicated for?
Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital…
Read the full statement (393 words)
Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography and digital subtraction angiography (DSA). The system may be used on pediatric, adult and bariatric patients.
Luminos Agile Max is not for mammography examinations.
Luminos dRF Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography and digital subtraction angiography (DSA). The system may be used on pediatric, adult and bariatric patients.
Luminos dRF Max is not for mammography examinations.
Multitom Rax is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography and digital subtraction angiography (DSA). The system may be used on pediatric, adult and bariatric patients.
Multitom Rax is not for mammography examinations.
Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system is designed primarily for urological diagnosis and the support of urological therapeutic applications such as examinations and small interventions of the urogenital tract. The table supports endourological and minimal invasive surgery in urology as there are transurethral interventions (e.g. ureterorenoscopy (URS), double stent placement, cystoscopy, transurethral resection of bladder tumors (TURB), transurethral resection of the prostate (TURP)), percutaneous urological procedures (e.g. percutaneous nephrostomy (PCN), percutaneous nephrolitholapaxy (PCNL)), urological X-ray diagnosis (e.g. survey imaging of the kidney, ureter, and bladder (KUB), intravenous pyelogram (IVP), retrograde pyelography), micturition cystourethrogram (MCU), videourodynamics, laparoscopic procedures and minor open urological interventions. The system may be used on pediatric, adult and bariatric patients.
Uroskop Omnia Max is not for mammography examinations.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K173639 · printed Nov 6, 2026
Cite this page
510(k) Report. “K173639: Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173639