510(k) Report

510(k) · Cleared

Cleared

Saniclave 250

K180272510(k) TraditionalCleared Jul 26, 2018176-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Revolutionary Science · FLE Sterilizer, Steam

Cleared on Jul 26, 2018 as substantially equivalent to K112811 (Alternative Pioneering Research and Development, 2012) after a 176-day review. Its family tree reaches back 3 generations to 1984.

Decided
Jul 26, 2018
Review
176 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K833837 (1984). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK833837 American Gold Series AutoclaveTuttnauer Co., Ltd. · 1984
  2. 2 generations backK953938 SterilemaxBarnstead/Thermolyne Corp. · 1997
  3. 1 generation backK112811 Saniclave 200Alternative Pioneering Research and Development · 2012
  4. This deviceK180272 Saniclave 250Revolutionary Science · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K833837American Gold Series AutoclaveTuttnauer Co., Ltd.1984—
−2K953938SterilemaxBarnstead/Thermolyne Corp.1997—
−1K112811Saniclave 200Alternative Pioneering Research and Development2012—
0 (this device)K180272Saniclave 250Revolutionary Science2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
FIG. 1 — Lineage of K180272, 1984–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112811Saniclave 200
Summary page 4
…Predicate Device: The Revolutionary Science Saniclave 250 (RS-SC-250) is claimed to be substantially equivalent to the FDA cleared Saniclave 200 under the 510(k) number K112811. Technical Characteristics, Intended Use, and cycle parameters are all similar to the predicate device. Intended Use: The Revo…
Alternative Pioneering Research and Development2012—

Also cleared against K112811: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 34 years and passes through 4 companies; a new generation was cleared every 13.1 years on median.
  • None of the 4 ancestors has a recall.
  • 1 other device also cites its predicate K112811.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Revolutionary Science Saniclave 250 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize wrapped and unwrapped heat and moisture stable solid instruments. mated surfaces, knurled, hinged devices, including lumened devices and trocars (lumens greater than 3mm inner diameter and less than 70mm in length, excluding dental handpieces) that are compatible…

Read the full statement (232 words)

The Revolutionary Science Saniclave 250 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize wrapped and unwrapped heat and moisture stable solid instruments. mated surfaces, knurled, hinged devices, including lumened devices and trocars (lumens greater than 3mm inner diameter and less than 70mm in length, excluding dental handpieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes.

The Saniclave 250 has 3 factory-programmed cycles:

Cycle Sterilization Temperature | Sterilize Time | Dry Time | Recommended Load

Gravity, 121°C, small load, wrapped* | 30 minutes | 30 minutes | 1 tray, wrapped* instruments, maximum weight 4.5 Ibs; Heat and moisture stable solid instruments, mated surfaces, knurled, hinged devices, including lumened devices and trocars

Gravity, 121°C, small load, unwrapped* | 30 minutes | N/A | 1 tray, unwrapped* instruments, maximum weight 4.5 Ibs; Heat and moisture stable solid instruments, mated surfaces, knurled, hinged devices, including lumened devices and trocars

Gravity, 121°C, large load, unwrapped* | 30 minutes | N/A | 3 trays, unwrapped* loaded evenly, maximum weight 8.0 Ibs; Heat and moisture stable solid instruments, lumened devices and trocars only

*4 wrapped load is wrapped or pouched in autoclave packaging that is FDA cleared for 30 min at 121°C. An unwrapped load is not wrapped or pouched and is intended for immediate use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K180272: Saniclave 250.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K180272

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026