510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

DR 800 with MUSICA Dynamic

K180589510(k) TraditionalCleared Apr 5, 201830-day reviewClass II21 CFR 892.1650Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Agfa Healthcare N.V. · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Apr 5, 2018 as substantially equivalent to K140380 (General Medical Merate S.P.A, 2014) after a 30-day review. Its family tree reaches back 1 generation to 2014. 9 later devices descend from it. It has been recalled 3 times.

Decided
Apr 5, 2018
Review
30 days
Class
Class II
Product code
JAA
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K140380 (2014). 9 later devices descend from it, 1 citing it directly.

Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K140380 · Opera Swing · General Medical Merate S.P.A · 20140+5
  1. Root · 1 generation backK140380 Opera SwingGeneral Medical Merate S.P.A · 2014
  2. This deviceK180589 DR 800 with MUSICA DynamicAgfa Healthcare N.V. · 2018
  3. Descendant · cites it directly · 8 of its ownK183275 DR 800 with TomosynthesisAgfa N.V. · 2019 · Recalled
  4. Descendant · 2 generations on · 4 of its ownK193262 DR 600 with TomosynthesisAgfa N.V. · 2020
  5. Descendant · 3 generations on · 3 of its ownK222934 Nanox.ARCNano-X Imaging, Ltd. · 2023
  6. Descendant · 4 generations on · 2 of its ownK242395 Nanox.ARCNano-X Imaging, Ltd. · 2024
  7. Descendant · 5 generations on · 1 of its ownK250850 Nanox.ARC XNano-X Imaging, Ltd. · 2025

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2019: 12020: 12021: 12023: 22024: 12025: 22026: 1
2019high 2 (2023) · low 0 (2022)2026

Most influential descendants

Showing 5 of 9 · See all
  • K183275DR 800 with TomosynthesisAgfa N.V. · 2019 · cites it directly8descendants
  • K193262DR 600 with TomosynthesisAgfa N.V. · 2020 · 2 generations on4descendants
  • K222934Nanox.ARCNano-X Imaging, Ltd. · 2023 · 3 generations on3descendants
  • K242395Nanox.ARCNano-X Imaging, Ltd. · 2024 · 4 generations on2descendants
  • K250850Nanox.ARC XNano-X Imaging, Ltd. · 2025 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K140380Opera SwingGeneral Medical Merate S.P.A2014—
0 (this device)K180589DR 800 with MUSICA DynamicAgfa Healthcare N.V.2018—
+1K183275DR 800 with TomosynthesisAgfa N.V.20193
+2K193262DR 600 with TomosynthesisAgfa N.V.2020—
+3K222934Nanox.ARCNano-X Imaging, Ltd.2023—
+4K242395Nanox.ARCNano-X Imaging, Ltd.2024—
+5K250850Nanox.ARC XNano-X Imaging, Ltd.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
202110
202320
202410
202520
202610
FIG. 1 — Lineage of K180589, 2014–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140380Opera Swing
Summary page 4
…gfa’s DR 800 with MUSICA Dynamic, which is an image-intensified fluoroscopic x-ray system. It is substantially equivalent to General Medical Merate S.P.A.’s OPERA Swing (K140380). Name of Device: OPERA Swing Common Name: System, X-Ray, Fluoroscopic, Image-Intensified Classification Name: Image-Intensified…
General Medical Merate S.P.A2014—
†K152639DR 600Agfa Healthcare N.V.2015—
†K141192DR 400Agfa Healthcare N.V.2014—
†K161368DX-D Imaging Package (DR 10s), DX-D Imaging Package (DR 14s), DX-D Imaging Package (DR…Agfa Healthcare N.V.2016—
†K103597DX-D 100Agfa Healthcare N.V.20116
†K103050DX-D 300Agfa Healthcare N.V.2010—
†K112670DX-D 600Agfa Healthcare N.V.20114
†K051901NIO 3MP-21Barcoview2005—
†K081963Agfa Computed Radiography Systems with NX2008 WorkstationsAgfa Healthcare Corporation2008—

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183275DR 800 with TomosynthesisAgfa N.V.20198

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 29, 2024Class IIZ-2738-2024Device DesignPotential for the front lever chain of the DR 800 table to fail.
Apr 6, 2020Class IIZ-2202-2020Software designUnder specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Jun 27, 2019Class IIZ-2056-2019Device DesignThe exposed area of fluo exams possibly does not match the active area of the detector within specified limits.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K183275 (Agfa N.V., 2019) has gone furthest, with 8 descendants of its own.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:

* Positioning fluoroscopy procedures

* Gastro-intestinal examinations

* Urogenital tract examinations

* Angiography

It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts.

The DR 800 is not intended for mammography applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K180589: DR 800 with MUSICA Dynamic.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K180589

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026