510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Elekta Unity

K182076510(k) TraditionalCleared Dec 4, 2018125-day reviewClass II21 CFR 892.5050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta Limited · IYE Accelerator, Linear, Medical

Cleared on Dec 4, 2018 as substantially equivalent to K170751 (Viewray, Incorporated, 2017) after a 125-day review. Its family tree reaches back 10 generations to 2001. 3 later devices descend from it. It has been recalled 4 times.

Decided
Dec 4, 2018
Review
125 days
Class
Class II
Product code
IYE
Recalls
4
Predicates
1

Its longest line of descent runs back 10 generations to K010975 (2001). 41 ancestors in all, the oldest K864886 (1987). 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2009: 1 other ancestor2008: 1 other ancestor2007: 3 other ancestors2006: 2 other ancestors2005: 2 other ancestors2004: 3 other ancestors2001: 1 other ancestor1999: 2 other ancestors2003: 2 other ancestors2000: 5 other ancestors1997: 2 other ancestors1987: 1 other ancestor2002: 1 other ancestor1996: 2 other ancestors1995: 1 other ancestor1991: 1 other ancestor2019: 1 descendant cleared2021: 1 descendant cleared2023: 1 descendant clearedK010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalledK030981 · Eclipse · Varian Medical Systems, Inc. · 2003K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005×2: 2 generations cleared close together, K071873 (2007), K073020 (2007) K091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK102915 · Viewray Treatment Planning and Delivery System · Viewray, Incorporated · 2011 · recalled×2: 2 generations cleared close together, K162393 (2017), K170751 (2017) 1990200020102020
  1. Root · 10 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 6 generations between
  3. 3 generations backK102915 Viewray Treatment Planning and Delivery SystemViewray, Incorporated · 2011 · Recalled
  4. 2 generations backK162393 The ViewRay (MRIdian) Linac SystemViewray, Incorporated · 2017
  5. 1 generation backK170751 MRIdian Linac System with 138-leaf CollimatorViewray, Incorporated · 2017 · Recalled
  6. This deviceK182076 Elekta UnityElekta Limited · 2018
  7. Descendant · cites it directly · 2 of its ownK192482 Elekta UnityElekta Limited · 2019 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK212114 Elekta UnityElekta Solutions AB · 2021 · Recalled
  9. Descendant · 3 generations on · 0 of its ownK223209 Elekta UnityElekta Solutions AB · 2023
Show all 10 generations
  1. Root · 10 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 9 generations backK030981 EclipseVarian Medical Systems, Inc. · 2003
  3. 8 generations backK041403 Eclipse with AaaVarian Medical Systems · 2004
  4. 7 generations backK050296 Eclipse Treatment Planning SystemVarian Medical Systems · 2005
  5. 6 generations backK071873 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  6. 5 generations backK073020 Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  7. 4 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
  8. 3 generations backK102915 Viewray Treatment Planning and Delivery SystemViewray, Incorporated · 2011 · Recalled
  9. 2 generations backK162393 The ViewRay (MRIdian) Linac SystemViewray, Incorporated · 2017
  10. 1 generation backK170751 MRIdian Linac System with 138-leaf CollimatorViewray, Incorporated · 2017 · Recalled
  11. This deviceK182076 Elekta UnityElekta Limited · 2018

Most influential descendants

Showing 3 of 3 · See all
  • K192482Elekta UnityElekta Limited · 2019 · cites it directly2descendants
  • K212114Elekta UnityElekta Solutions AB · 2021 · 2 generations on1descendant
  • K223209Elekta UnityElekta Solutions AB · 2023 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K010975Eclipse 6.5Varian Medical Systems, Inc.20013
−9K030981EclipseVarian Medical Systems, Inc.2003—
−8K041403Eclipse with AaaVarian Medical Systems2004—
−7K050296Eclipse Treatment Planning SystemVarian Medical Systems2005—
−6K071873Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20075
−5K073020Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20076
−4K091492Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20092
−3K102915Viewray Treatment Planning and Delivery SystemViewray, Incorporated20111
−2K162393The ViewRay (MRIdian) Linac SystemViewray, Incorporated2017—
−1K170751MRIdian Linac System with 138-leaf CollimatorViewray, Incorporated20174
0 (this device)K182076Elekta UnityElekta Limited2018—
+1K192482Elekta UnityElekta Limited20193
+2K212114Elekta UnityElekta Solutions AB20211
+3K223209Elekta UnityElekta Solutions AB2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199110
199510
199620
199720
199920
200050
200110
200210
200320
200430
200520
200620
200730
200810
200910
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202110
202310
FIG. 1 — Lineage of K182076, 2001–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170751MRIdian Linac System with 138-leaf Collimator
Summary page 4
…t Code: IYE, LNH Device classification name: Accelerator, Linear, Medical Predicate Device: ViewRay (MRIdian) Linac System, K170751 Product Description: Elekta Unity is a multifunctional digital linear accelerator designed to assist licensed medical practitio…
Viewray, Incorporated20174

Also cleared against K170751: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192482Elekta UnityElekta Limited20192

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 24, 2019Class IIZ-1980-2019Software designThe QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.
Apr 3, 2019Class IIZ-1297-2019Labeling designThere is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.
Mar 12, 2019Class IIZ-1109-2019Software designUsers need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images…
Jan 4, 2019Class IIZ-0706-2019Under Investigation by firmIn rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 27 of the 41 ancestors have a recall of any class.
  • Of its descendants, K192482 (Elekta Limited, 2019) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K170751.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with the compatible Treatment Planning and Oncology Information Systems. The Elekta Unity 1.5T MR scanner is a magnetic resonance imaging system…

Read the full statement (105 words)

Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Elekta Unity is intended for use with the compatible Treatment Planning and Oncology Information Systems.

The Elekta Unity 1.5T MR scanner is a magnetic resonance imaging system that produces cross-sectional images in any orientation of the internal structure of the whole body before, during, and after the radiotherapy treatment. Images provide information that may assist therapy planning, patient positioning and treatment delivery related to radiation oncology.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K182076: Elekta Unity.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K182076

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026