Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD

K210500510(k) TraditionalCleared Jun 11, 2021109-day reviewClass II21 CFR 892.5050Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta Limited · IYE Accelerator, Linear, Medical

Cleared on Jun 11, 2021 as substantially equivalent to K192242 (Elekta Limited, 2019) after a 109-day review. Its family tree reaches back 10 generations to 1995. 1 later device descends from it. It has been recalled 10 times.

Decided
Jun 11, 2021
Review
109 days
Class
Class II
Product code
IYE
Recalls
10
Predicates
1

Its longest line of descent runs back 10 generations to K926449 (1995). 33 ancestors in all, the oldest K864886 (1987). 1 later device descends from it, 1 citing it directly.

  1. Root · 10 generations backK926449 Varian Multileaf Collimator, Release 4.0Varian Assoc., Inc. · 1995
  2. 6 generations between
  3. 3 generations backK123808 AgilityElekta, Ltd. · 2013 · Recalled
  4. 2 generations backK182138 Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited · 2018
  5. 1 generation backK192242 Synergy Platform, Synergy, Infinity, VersaHDElekta Limited · 2019
  6. This deviceK210500 Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HDElekta Limited · 2021
  7. Descendant · cites it directly · 0 of its ownK252188 EMLA (Elekta Evo); EMLA (VersaHD); EMLA (Elekta Harmony Pro); EMLA…Elekta Solutions AB · 2026
Show all 10 generations
  1. Root · 10 generations backK926449 Varian Multileaf Collimator, Release 4.0Varian Assoc., Inc. · 1995
  2. 9 generations backK973889 Varian Multileaf Collimator with Dynamic Arc Therapy FeatureVarian Medical Systems, Inc. · 1997
  3. 8 generations backK990085 Millennium MLC -120Varian Assoc., Inc. · 1999 · Recalled
  4. 7 generations backK050442 Millennium Multi-Leaf CollimatorVarian Medical Systems · 2005 · Recalled
  5. 6 generations backK071992 High-Definition 120 Multileaf Collimator (HD120 MLC)Varian Medical Systems · 2007 · Recalled
  6. 5 generations backK092871 Trilogy MXVarian Medical Systems, Inc. · 2009 · Recalled
  7. 4 generations backK111106 TruebeamVarian Medical Systems, Inc. · 2011 · Recalled
  8. 3 generations backK123808 AgilityElekta, Ltd. · 2013 · Recalled
  9. 2 generations backK182138 Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited · 2018
  10. 1 generation backK192242 Synergy Platform, Synergy, Infinity, VersaHDElekta Limited · 2019
  11. This deviceK210500 Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HDElekta Limited · 2021

Most influential descendants

Showing 1 of 1 · See all
  • K252188EMLA (Elekta Evo); EMLA (VersaHD); EMLA (Elekta Harmony Pro); EMLA (Elekta Infintiy)…Elekta Solutions AB · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K926449Varian Multileaf Collimator, Release 4.0Varian Assoc., Inc.1995—
−9K973889Varian Multileaf Collimator with Dynamic Arc Therapy FeatureVarian Medical Systems, Inc.1997—
−8K990085Millennium MLC -120Varian Assoc., Inc.19992
−7K050442Millennium Multi-Leaf CollimatorVarian Medical Systems20052
−6K071992High-Definition 120 Multileaf Collimator (HD120 MLC)Varian Medical Systems20074
−5K092871Trilogy MXVarian Medical Systems, Inc.20095
−4K111106TruebeamVarian Medical Systems, Inc.20115
−3K123808AgilityElekta, Ltd.20132
−2K182138Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited2018—
−1K192242Synergy Platform, Synergy, Infinity, VersaHDElekta Limited2019—
0 (this device)K210500Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HDElekta Limited2021—
+1K252188EMLA (Elekta Evo); EMLA (VersaHD); EMLA (Elekta Harmony Pro); EMLA (Elekta…Elekta Solutions AB2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199110
199620
199710
199910
200020
200110
200320
200430
200520
200620
200720
200820
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K210500, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K192242Synergy Platform, Synergy, Infinity, VersaHD
Summary page 4
…Accelerator, Linear, Medical, 21CFR 892.5050 Product Code: IYE III. PREDICATE DEVICE K192242 IV. DEVICE DESCRIPTION The Elekta Medical Linear Accelerator (EMLA) system is an image guided Radiation Therap…
Elekta Limited2019—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K252188EMLA (Elekta Evo); EMLA (VersaHD); EMLA (Elekta Harmony Pro); EMLA (Elekta Infintiy)…Elekta Solutions AB20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 14, 2025Class IIZ-1164-2026Process controlDue to a manufacturing issue and electrical grounding of systems.
Oct 25, 2023Class IIZ-0443-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
Oct 25, 2023Class IIZ-0444-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
Oct 25, 2023Class IIZ-0447-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
Oct 25, 2023Class IIZ-0442-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
Oct 25, 2023Class IIZ-0445-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
Oct 25, 2023Class IIZ-0446-2024Under Investigation by firmElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of…
May 5, 2023Class IIZ-1665-2023Under Investigation by firmA component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601
Sep 28, 2021Class IIZ-0115-2022Nonconforming Material/ComponentThere is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
Sep 17, 2021Class IIZ-0116-2022Under Investigation by firmThere is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 6 companies; a new generation was cleared every 2.3 years on median.
  • 24 of the 33 ancestors have a recall of any class.
  • Of its descendants, K252188 (Elekta Solutions AB, 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The Elekta Medical Linear Accelerator (EMLA) is intended to be used for external beam radiation therapy (EBRT) treatments as determined by a licensed medical practitioner. It is intended to assist a licensed medical practitioner in the delivery of EBRT to defined target volumes, while sparing surrounding normal tissue and critical organs from excess radiation. −Elekta Synergy and Elekta Harmony are the default entry-level configurations. They are…

Read the full statement (238 words)

The Elekta Medical Linear Accelerator (EMLA) is intended to be used for external beam radiation therapy (EBRT) treatments as determined by a licensed medical practitioner.

It is intended to assist a licensed medical practitioner in the delivery of EBRT to defined target volumes, while sparing surrounding normal tissue and critical organs from excess radiation.

−Elekta Synergy and Elekta Harmony are the default entry-level configurations. They are intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, and hypofractionation in all areas of the body where such treatment is indicated.

−Elekta Infinity is the default mid-level configuration. It is intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR) in all areas of the body where such treatment is indicated.

−Versa HD is the default high-level configuration. It is intended to be used for single or multiple fractions using standard fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR; stereotactic radio surgery - SRS) in all areas of the body where such treatment is indicated and for the treatment of functional disorders, such as trigeminal neuralgia.

The EMLA is indicated for the delivery of curative and palliative intent EBRT to Adult and Pediatric patients with primary benign and malignant tumor and metastasis (or secondaries) anywhere in the body.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K210500: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K210500

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026