510(k) Report

510(k) · Cleared

Cleared

Persyst 14 EEG Review and Analysis Software

K182181510(k) TraditionalCleared Nov 29, 2018108-day reviewClass II21 CFR 882.1400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Persyst Development Corporation · OMB Automatic Event Detection Software For Full-Montage Electroencephalograph

Cleared on Nov 29, 2018 after a 108-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 6 later devices descend from it.

Decided
Nov 29, 2018
Review
108 days
Class
Class II
Product code
OMB
Recalls
None
Predicates
0

6 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2024: 3 descendants cleared2019: 1 descendant cleared2020: 1 descendant cleared2026: 1 descendant cleared201620182020202220242026
  1. This deviceK182181 Persyst 14 EEG Review and Analysis SoftwarePersyst Development Corporation · 2018
  2. Descendant · cites it directly · 1 of its ownK231779 REMI AI Discrete Detection ModuleEpitel, Inc. · 2024
  3. Descendant · cites it directly · 1 of its ownK201910 EZTrackNeurologic, LLC · 2020
  4. Descendant · 2 generations on · 0 of its ownK260563 CN-Suite™FIND Surgical Sciences Inc. D/B/A FIND Neuro · 2026
  5. Descendant · cites it directly · 0 of its ownK241390 NeuroMatchLVIS Corporation · 2024
  6. Descendant · 2 generations on · 0 of its ownK240408 REMI-AI Rapid Detection Module (REMI-AI RDM)Epitel, Inc. · 2024

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2019: 12020: 12024: 32026: 1
2019high 3 (2024) · low 0 (2021)2026

Most influential descendants

Showing 5 of 6 · See all
  • K231779REMI AI Discrete Detection ModuleEpitel, Inc. · 2024 · cites it directly1descendant
  • K201910EZTrackNeurologic, LLC · 2020 · cites it directly1descendant
  • K260563CN-Suite™FIND Surgical Sciences Inc. D/B/A FIND Neuro · 2026 · 2 generations on0descendants
  • K241390NeuroMatchLVIS Corporation · 2024 · cites it directly0descendants
  • K240408REMI-AI Rapid Detection Module (REMI-AI RDM)Epitel, Inc. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K182181Persyst 14 EEG Review and Analysis SoftwarePersyst Development Corporation2018—
+1K231779REMI AI Discrete Detection ModuleEpitel, Inc.2024—
+1K201910EZTrackNeurologic, LLC2020—
+2K260563CN-Suite™FIND Surgical Sciences Inc. D/B/A FIND Neuro2026—
+1K241390NeuroMatchLVIS Corporation2024—
+2K240408REMI-AI Rapid Detection Module (REMI-AI RDM)Epitel, Inc.2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202010
202430
202610
FIG. 1 — Lineage of K182181, 2018–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OMB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OMB device

  1. #1 K240993 encevis (2.1)score 15.0 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (6)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231779REMI AI Discrete Detection ModuleEpitel, Inc.20241
K201910EZTrackNeurologic, LLC20201
K241390NeuroMatchLVIS Corporation20240
K192572CNS EnvisionMoberg Research, Inc.20190

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K231779 (Epitel, Inc., 2024) has gone furthest, with 1 descendant of its own.

How does its category compare?

All OMB devices

OMB Automatic Event Detection Software For Full-Montage Electroencephalograph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · OMB

5in 2025

2009: 12010: 12012: 12013: 12014: 22015: 32017: 12018: 32019: 12020: 12021: 12022: 12023: 22024: 52025: 52026 (partial year): 1
2007high 5 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OMB

143in 2025

2015: 233 days median2018: 108 days median2024: 201 days median2025: 143 days median
2007high 233 (2015) · low 108 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OMB

1in 2021

2021: 1
2007high 1 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OMB

6in 2025

2011: 12014: 22022: 12023: 42024: 212025: 6
2007high 21 (2024) · low 0 (2007)2025

MAUDE reports naming product code OMB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20101—00
20110—01
20121—00
20131—00
20142—02
2015323300
20160—00
20171—00
2018310800
20191—00
20201—00
20211—10
20221—01
20232—04
20245201021
2025514306
20261—0—

What is it indicated for?

1. Persyst 14 EEG Review and Analysis Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. The device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information. 2. The Seizure Detection component of Persyst 14 is intended to mark previously acquired…

Read the full statement (503 words)

1. Persyst 14 EEG Review and Analysis Software is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. The device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information.

2. The Seizure Detection component of Persyst 14 is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20 system.

3. The Spike Detection component of Persyst 14 is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in patients at least one month old. Persyst 14 Spike Detection performance has not been assessed for intracranial recordings.

4. Persyst 14 includes the calculation and display of a set of qualitative measures intended to monitor and analyze the EEG waveform. These include FFT, Rhythmicity, Peak Envelope, Artifact Intensity, Amplitude, Relative Symmetry and Suppression Ratio. Automatic event marking is not applicable to the quantitative measures. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

5. Persyst 14 displays physiological signals, including the calculation and display of a heart rate measurement based on the ECG channel in the EEG recording, which are intended to aid in the analysis of an EEG. Heart rate measurement of Persyst 14 is not applicable to patients with pacemaker and/or active implantable devices.

6. The aEEG functionality included in Persyst 14 is intended to monitor the state of the brain. The automated event marking function of Persyst 14 is not applicable to aEEG.

7. Persyst 14 provides notifications for seizure detection, quantitative EEG and aEEG that can be used when processing a record during acquisition. These include an on screen display and the optional sending of an email message. Delays of up to several minutes can occur between the beginning of a seizure and when the Persyst 14 notifications will be shown to a user. Persyst 14 notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

8. Persyst AR (Artifact Reduction) is intended to reduce EMG, eye movement, and electrode artifacts in a standard 10-20 EEG recording. AR does not remove the entire artifact signal, and is not effective for other types of artifacts. AR may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.

9. This device does not provide any diagnostic conclusion about the patient's condition to the user.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K182181: Persyst 14 EEG Review and Analysis Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182181

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026