510(k) Report

510(k) · Cleared

Cleared

Cook Unimpregnated Central Venous Catheter

K182252510(k) TraditionalCleared May 17, 2019270-day reviewClass II21 CFR 880.5200

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook Incorporated · FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Cleared on May 17, 2019 as substantially equivalent to K081113 (Cook, Inc., 2008) after a 270-day review. Its family tree reaches back 5 generations to 1997. 1 later device descends from it.

Decided
May 17, 2019
Review
270 days
Class
Class II
Product code
FOZ
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K960764 (1997). 10 ancestors in all, the oldest K904404 (1991). 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK960764 Safe-CathScientific Device Manufacturer, LLC · 1997
  2. 4 generations backK984189 Argon Multi-Lumen Central Venous CatheterMaxxim Medical · 1999
  3. 3 generations backK032274 Cook Five Lumen Central Venous CatheterCook, Inc. · 2004
  4. 2 generations backK060174 Spectrum Five Lumen Central Venous Catheter Set, Model C-UQLM10.0-AbrmCook, Inc. · 2006
  5. 1 generation backK081113 Cook and Cook Spectrum Central Venous CathetersCook, Inc. · 2008 · Recalled
  6. This deviceK182252 Cook Unimpregnated Central Venous CatheterCook Incorporated · 2019
  7. Descendant · cites it directly · 0 of its ownK260165 Unimpregnated Central Venous CatheterSpectrum Vascular · 2026

Most influential descendants

Showing 1 of 1 · See all
  • K260165Unimpregnated Central Venous CatheterSpectrum Vascular · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K960764Safe-CathScientific Device Manufacturer, LLC1997—
−4K984189Argon Multi-Lumen Central Venous CatheterMaxxim Medical1999—
−3K032274Cook Five Lumen Central Venous CatheterCook, Inc.2004—
−2K060174Spectrum Five Lumen Central Venous Catheter Set, Model C-UQLM10.0-AbrmCook, Inc.2006—
−1K081113Cook and Cook Spectrum Central Venous CathetersCook, Inc.2008105
0 (this device)K182252Cook Unimpregnated Central Venous CatheterCook Incorporated2019—
+1K260165Unimpregnated Central Venous CatheterSpectrum Vascular2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199610
199710
200110
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K182252, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FOZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FOZ device

  1. #1 K243403 BD Nexiva™ Closed IV Catheter Systemscore 14.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081113Cook and Cook Spectrum Central Venous Catheters
Summary page 5
…ice: The Cook Unimpregnated Central Venous Catheter is substantially equivalent to the following devices: - Cook Central Venous Catheters with or without Heparin (K081113, Cook Incorporated) cleared on July 30, 2008 Cook Incorporated – Traditional 510(k)…
Cook, Inc.2008105
†K183036Dilator SetsCook Incorporated20183
†K072625Turbo-Ject Peripherally Inserted Central Venous Catheter (Picc)Cook, Inc.20079

Also cleared against K081113: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K260165Unimpregnated Central Venous CatheterSpectrum Vascular20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • Of its descendants, K260165 (Spectrum Vascular, 2026) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K081113.

How does its category compare?

All FOZ devices

FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · FOZ

11in 2025

2007: 62008: 92009: 92010: 32011: 62012: 132013: 42014: 92015: 102016: 92017: 92018: 12019: 82020: 102021: 42022: 82023: 92024: 62025: 112026 (partial year): 7
2007high 13 (2012) · low 1 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FOZ

91in 2025

2007: 159 days median2008: 103 days median2009: 30 days median2010: 28 days median2011: 132 days median2012: 122 days median2013: 104 days median2014: 99 days median2015: 107 days median2016: 63 days median2017: 51 days median2019: 140 days median2020: 191 days median2021: 242 days median2022: 183 days median2023: 129 days median2024: 190 days median2025: 91 days median2026 (partial year): 203 days median
2007high 242 (2021) · low 28 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FOZ

4in 2023

2007: 162009: 232010: 1152013: 22014: 42016: 3, 1 Class I2017: 222018: 42019: 32020: 32021: 12022: 4, 2 Class I2023: 4, 2 Class I2026 (partial year): 2
2007high 115 (2010) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · FOZ

5,110in 2025

2007: 3742008: 4222009: 5082010: 3702011: 7862012: 5602013: 6402014: 8152015: 7292016: 9252017: 1,9662018: 3,3052019: 4,3652020: 3,7972021: 4,6742022: 4,1332023: 4,4302024: 4,7722025: 5,110
2007high 5,110 (2025) · low 370 (2010)2025

MAUDE reports naming product code FOZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007615916374
200891030422
200993023508
2010328115370
201161320786
2012131220560
201341042640
20149994815
2015101070729
20169633 (1)925
2017951221,966
20181—43,305
2019814034,365
20201019133,797
2021424214,674
202281834 (2)4,133
202391294 (2)4,430
2024619004,772
2025119105,110
202672032—

What is it indicated for?

The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only The device is a short-term use catheter, intended for less than 30 days. The Cook dilator is used for dilating…

Read the full statement (261 words)

The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:

1. Continuous or intermittent drug infusions

2. Central venous blood pressure monitoring (CVP)

3. Acute hyperalimentation

4. Blood sampling

5. Delivery of whole blood or blood products

6. Simultaneous, separate infusion of drugs for multi-lumen catheters only

The device is a short-term use catheter, intended for less than 30 days.

The Cook dilator is used for dilating puncture sites or catheter tracts.

The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations.

The Cook Unimpregnated Central Venous Catheter with power injection is used for:

1. Continuous or intermittent drug infusions

2. Central venous blood pressure monitoring (CVP)

3. Acute hyperalimentation

4. Blood sampling

5. Delivery of whole blood or blood products

6. Simultaneous, separate infusion of drugs for multi-lumen catheters only

7. Power injection of contrast media*

*The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0, and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0, and 10.0 French catheters.

The device is a short-term use catheter, intended for less than 30 days.

The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter.

The Cook dilator is used for dilating puncture sites or catheter tracts.

The Cook Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182252: Cook Unimpregnated Central Venous Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K182252

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026