510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Roadrunner PC Wire Guide, Olcott Torque Device

K182985510(k) TraditionalCleared Jul 22, 2019266-day reviewClass II21 CFR 870.1330Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook Incorporated · DQX Wire, Guide, Catheter

Cleared on Jul 22, 2019 as substantially equivalent to K920891 (Cook, Inc., 1992) after a 266-day review. Its family tree reaches back 1 generation to 1992. It has been recalled 8 times.

Decided
Jul 22, 2019
Review
266 days
Class
Class II
Product code
DQX
Recalls
8
Predicates
1

Its longest line of descent runs back 1 generation to K920891 (1992). No later device cites it as a predicate yet.

  1. Root · 1 generation backK920891 Roadrunner (TM) Wire Guide, ModificationCook, Inc. · 1992
  2. This deviceK182985 Roadrunner PC Wire Guide, Olcott Torque DeviceCook Incorporated · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K920891Roadrunner (TM) Wire Guide, ModificationCook, Inc.1992—
0 (this device)K182985Roadrunner PC Wire Guide, Olcott Torque DeviceCook Incorporated2019—
FIG. 1 — Lineage of K182985, 1992–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQX device

  1. #1 K213854 SavvyWirescore 15.7 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K920891Roadrunner (TM) Wire Guide, Modification
Summary page 4
…res, exclusive of the coronary arteries. Predicate Device: The predicate device, the Roadrunner Wire Guide, was cleared for commercial distribution under 510(k) number K920891, on December 21, 1992. Comparison to Predicate Device: It has been demonstrated that the subject device, Roadrunner® PC Wire Gu…
Cook, Inc.1992—
†K171997Mandril Wire GuideCook Incorporated20183
†K130766Roadrunner Uniglide Hydrophilic Wire GuideCook, Inc.2013Class I

Also cleared against K920891: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 16, 2022Class IIZ-1839-2022Under Investigation by firmCook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Aug 16, 2022Class IIZ-1840-2022Under Investigation by firmCook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Aug 16, 2022Class IIZ-1838-2022Under Investigation by firmCook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Aug 16, 2022Class IIZ-1841-2022Under Investigation by firmCook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Jul 19, 2022Class IIZ-1708-2022Under Investigation by firmProduct did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Jul 19, 2022Class IIZ-1706-2022Under Investigation by firmProduct did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Jul 19, 2022Class IIZ-1709-2022Under Investigation by firmProduct did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Jul 19, 2022Class IIZ-1707-2022Under Investigation by firmProduct did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 2 other devices also cite its predicate K920891.

How does its category compare?

All DQX devices

DQX Wire, Guide, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · DQX

10in 2025

2007: 142008: 192009: 182010: 212011: 182012: 202013: 182014: 182015: 152016: 142017: 192018: 192019: 132020: 142021: 62022: 152023: 102024: 112025: 102026 (partial year): 12
2007high 21 (2010) · low 6 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQX

177in 2025

2007: 85 days median2008: 87 days median2009: 87 days median2010: 63 days median2011: 43 days median2012: 62 days median2013: 119 days median2014: 103 days median2015: 112 days median2016: 179 days median2017: 85 days median2018: 114 days median2019: 173 days median2020: 106 days median2021: 161 days median2022: 212 days median2023: 136 days median2024: 134 days median2025: 177 days median2026 (partial year): 145 days median
2007high 212 (2022) · low 43 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQX

3in 2024

2007: 122008: 72009: 492010: 342011: 32012: 52013: 7, 4 Class I2015: 12016: 2, 2 Class I2017: 172018: 92019: 22020: 1, 1 Class I2024: 32026 (partial year): 4, 1 Class I
2007high 49 (2009) · low 0 (2014)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · DQX

2,178in 2025

2007: 7602008: 7152009: 6872010: 9872011: 1,3112012: 1,3372013: 1,4742014: 1,3232015: 1,3232016: 1,4662017: 1,5242018: 1,7372019: 1,6792020: 1,5362021: 1,7132022: 1,6472023: 2,0682024: 2,2102025: 2,178
2007high 2,210 (2024) · low 687 (2009)2025

MAUDE reports naming product code DQX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007148512760
200819877715
2009188749687
2010216334987
2011184331,311
2012206251,337
2013181197 (4)1,474
20141810301,323
20151511211,323
2016141792 (2)1,466
20171985171,524
20181911491,737
20191317321,679
2020141061 (1)1,536
2021616101,713
20221521201,647
20231013602,068
20241113432,210
20251017702,178
2026121454 (1)—

What is it indicated for?

The Roadrunner® PC Wire Guide is used for catheter positioning and exchange in diagnostic and interventional procedures, exclusive of the coronary arteries.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182985: Roadrunner PC Wire Guide, Olcott Torque Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K182985

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026