510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO

K183394510(k) SpecialCleared May 1, 2019145-day reviewClass II21 CFR 868.1400Recalls 4 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare Finland Oy · CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Cleared on May 1, 2019 as substantially equivalent to K171028 (GE Healthcare Finland Oy, 2017) after a 145-day review. Its family tree reaches back 5 generations to 2000. 4 of the 7 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 4 times, including 4 Class I recalls.

Decided
May 1, 2019
Review
145 days
Class
Class II
Product code
CCK
Recalls
4 (4 Class I)
Predicates
1

Its longest line of descent runs back 5 generations to K001814 (2000). 1 later device descends from it, 1 citing it directly. 4 of the 7 devices in this family, counting this one, have a Class I recall.

  1. Root · 5 generations backK001814 Datex-Ohmeda Compact Airway Module M-Caiovx and M-Covx (Version 01)…Datex-Ohmeda · 2000
  2. 4 generations backK051092 Datex-Ohmeda S/5 Compact Airway Module (Model Family E-Caiovx)…GE Healthcare · 2005 · Recalled
  3. 3 generations backK123195 Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and AccessoriesGE Healthcare Finland Oy · 2013 · Recalled (Class I)
  4. 2 generations backK150245 CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO…GE Healthcare Finland Oy · 2015 · Recalled (Class I)
  5. 1 generation backK171028 CARESCAPE Respiratory MOdules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO…GE Healthcare Finland Oy · 2017 · Recalled (Class I)
  6. This deviceK183394 CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiOGE Healthcare Finland Oy · 2019
  7. Descendant · cites it directly · 0 of its ownK211171 CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCOVX, E-sCAiO…GE Healthcare Finland Oy · 2021

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K211171CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX…GE Healthcare Finland Oy · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K001814Datex-Ohmeda Compact Airway Module M-Caiovx and M-Covx (Version 01) and…Datex-Ohmeda2000—
−4K051092Datex-Ohmeda S/5 Compact Airway Module (Model Family E-Caiovx) E-Caiovx…GE Healthcare20052
−3K123195Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and AccessoriesGE Healthcare Finland Oy2013Class I
−2K150245CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV…GE Healthcare Finland Oy2015Class I
−1K171028CARESCAPE Respiratory MOdules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV…GE Healthcare Finland Oy2017Class I
0 (this device)K183394CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiOGE Healthcare Finland Oy2019—
+1K211171CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV…GE Healthcare Finland Oy2021—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K183394, 2000–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CCK device

  1. #1 K253030 Capnostream™35 Portable Respiratory Monitor…score 15.8 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K171028CARESCAPE Respiratory MOdules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and…
Summary page 6
…(Anesthetic Concentration) Additional Product CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Codes: Predicate Device(s): K171028: CARESCAPE Respiratory Modules, E-sCO, E- sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and…
GE Healthcare Finland Oy2017Class I

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211171CARESCAPE Respiratory Modules E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX…GE Healthcare Finland Oy20210

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 27, 2019Class IZ-0755-2020Process controlDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Sep 27, 2019Class IZ-0757-2020Process controlDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Sep 27, 2019Class IZ-0754-2020Process controlDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Sep 27, 2019Class IZ-0756-2020Process controlDisplayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K171028CARESCAPE Respiratory MOdules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX and accessoriesGE Healthcare Finland Oy · 2017 · Direct predicate · ◆ on the line of descent in FIG. 14 Class I recalls
  • K150245CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVXGE Healthcare Finland Oy · 2015 · 2 generations back · ◆ on the line of descent in FIG. 14 Class I recalls
  • K123195Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and AccessoriesGE Healthcare Finland Oy · 2013 · 3 generations back · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 4 of the 5 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K211171 (GE Healthcare Finland Oy, 2021) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CCK devices

CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · CCK

3in 2025

2007: 52008: 32009: 32010: 32011: 22012: 72013: 42014: 92015: 52016: 22017: 52018: 62019: 72020: 42021: 42022: 32023: 42024: 32025: 32026 (partial year): 3
2007high 9 (2014) · low 2 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CCK

169in 2025

2007: 56 days median2008: 248 days median2009: 101 days median2010: 125 days median2012: 150 days median2013: 132 days median2014: 166 days median2015: 105 days median2017: 134 days median2018: 296 days median2019: 261 days median2020: 188 days median2021: 227 days median2022: 226 days median2023: 555 days median2024: 592 days median2025: 169 days median2026 (partial year): 224 days median
2007high 592 (2024) · low 56 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CCK

1in 2022

2008: 12009: 192010: 42011: 142012: 32013: 2, 1 Class I2015: 12018: 52019: 7, 7 Class I2020: 12022: 1
2007high 19 (2009) · low 0 (2007)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · CCK

1,259in 2025

2007: 72008: 202009: 212010: 122011: 182012: 132013: 282014: 512015: 1142016: 882017: 582018: 632019: 1412020: 2,3022021: 2,2722022: 2,4752023: 1,6372024: 1,2162025: 1,259
2007high 2,475 (2022) · low 7 (2007)2025

MAUDE reports naming product code CCK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200755607
20083248120
200931011921
20103125412
20112—1418
20127150313
201341322 (1)28
20149166051
201551051114
20162—088
20175134058
20186296563
201972617 (7)141
2020418812,302
2021422702,272
2022322612,475
2023455501,637
2024359201,216
2025316901,259
202632240—

What is it indicated for?

The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, 02, N20, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2 VO32) of adult and pediatric patients. When monitoring neonatal…

Read the full statement (103 words)

The CARESCAPE Respiratory Modules (E-sCO, E-sCOV, E-sCOVX, E-sCAiO, E-sCAiOV, E-sCAiOVX) are indicated for use with a host device for monitoring respiratory parameters (CO2, 02, N20, anesthetic agents, anesthetic agent identification and respiratory rate) and ventilatory parameters (airway pressure, flow and volume) of adult, pediatric and neonatal patients and gas exchange parameters (VCO2 VO32) of adult and pediatric patients.

When monitoring neonatal or other patients that have high respiration rate or low tidal volume these modules shall be used within the limits of respiration rates and tidal volumes to ensure specified measurement accuracy.

These modules are intended for use by qualified medical personnel only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K183394: CARESCAPE Respiratory Modules, E-sCO, E-sCOV, E-sCOVX, E-sCAiO.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K183394

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026