510(k) Report

510(k) · Cleared

Cleared

Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Balance Biosurface

K183490510(k) TraditionalCleared Apr 12, 2019116-day reviewClass II21 CFR 870.4350

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Apr 12, 2019 after a 116-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 3 later devices descend from it.

Decided
Apr 12, 2019
Review
116 days
Class
Class II
Product code
DTZ
Recalls
None
Predicates
0

3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2021: 1 descendant cleared2023: 1 descendant cleared2024: 1 descendant cleared201620182020202220242026
  1. This deviceK183490 Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium…Medtronic, Inc. · 2019
  2. Descendant · cites it directly · 2 of its ownK203111 Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation SystemMedtronic · 2021 · Recalled
  3. Descendant · 2 generations on · 1 of its ownK230640 Affinity Fusion™ Oxygenator SystemMedtronic, Inc. · 2023
  4. Descendant · 3 generations on · 0 of its ownK240666 Affinity Fusion Oxygenator with Integrated Arterial Filter and…Medtronic Perfusion Systems · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K203111Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation SystemMedtronic · 2021 · cites it directly2descendants
  • K230640Affinity Fusion™ Oxygenator SystemMedtronic, Inc. · 2023 · 2 generations on1descendant
  • K240666Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface…Medtronic Perfusion Systems · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K183490Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface…Medtronic, Inc.2019—
+1K203111Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation SystemMedtronic20216
+2K230640Affinity Fusion™ Oxygenator SystemMedtronic, Inc.2023—
+3K240666Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance…Medtronic Perfusion Systems2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202310
202410
FIG. 1 — Lineage of K183490, 2019–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203111Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation SystemMedtronic20212

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K203111 (Medtronic, 2021) has gone furthest, with 2 descendants of its own.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—
Cite this page

510(k) Report. “K183490: Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Balance Biosurface.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K183490

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026