510(k) Report

510(k) · Cleared

Cleared

ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow Cytometer

K183592510(k) TraditionalCleared Mar 21, 201990-day reviewClass II21 CFR 864.7010

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter · PWD Flow Cytometric Test System For Hematopoietic Neoplasms

Cleared on Mar 21, 2019 as substantially equivalent to DEN160047 (Beckman Coulter, 2017) and 2 other predicates after a 90-day review. Its family tree reaches back 6 generations to 1993. 3 of the 13 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Mar 21, 2019
Review
90 days
Class
Class II
Product code
PWD
Recalls
None
Predicates
3

Its longest line of descent runs back 6 generations to K922745 (1993). 11 ancestors in all. 1 later device descends from it, 1 citing it directly. 3 of the 13 devices in this family, counting this one, have a Class I recall.

  1. Root · 6 generations backK922745 Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1Coulter Immunology · 1993
  2. 5 generations backK963263 Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat…Coulter Corp. · 1997 · Recalled (Class I)
  3. 4 generations backK990172 Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat…Coulter Corp. · 1999 · Recalled (Class I)
  4. 3 generations backK030828 Tetracxp System (Software and System Guide)Beckman Coulter, Inc. · 2003 · Recalled (Class I)
  5. 2 generations backK130373 Navios Flow Cytometer SystemBeckman Coulter, Inc. · 2013
  6. 1 generation backK162897 Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer…Beckman Coulter · 2017
  7. This deviceK183592 ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios…Beckman Coulter · 2019
  8. Descendant · cites it directly · 0 of its ownK232600 DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc. · 2023 · Recalled

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K232600DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc. · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K922745Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1Coulter Immunology1993—
−5K963263Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome…Coulter Corp.1997Class I
−4K990172Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome…Coulter Corp.1999Class I
−3K030828Tetracxp System (Software and System Guide)Beckman Coulter, Inc.2003Class I
−2K130373Navios Flow Cytometer SystemBeckman Coulter, Inc.2013—
−1K162897Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer, 8 Color/2…Beckman Coulter2017—
0 (this device)K183592ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow…Beckman Coulter2019—
+1K232600DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc.20231

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
199410
199810
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K183592, 1993–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*DEN160047ClearLLab T1, ClearLLab T2, ClearLLab B1, ClearLLab B2, ClearLLab M
Summary page 8
…Classification 21 CFR Number Cleared Code ClearLLab Reagents DEN160047 06/29/2017 Class II 864.7010 PWD Navios Flow Cytometer K130373 9/18/2013 Class II…
Beckman Coulter2017—
*K130373Navios Flow Cytometer System
Summary page 8
…Code ClearLLab Reagents DEN160047 06/29/2017 Class II 864.7010 PWD Navios Flow Cytometer K130373 9/18/2013 Class II 864.5220 OYE Navios EX Flow Cytometer K162897 6/23/2017 Class II…
Beckman Coulter, Inc.2013—
*K162897Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer, 8 Color/2 Laser…
Summary page 8
…II 864.7010 PWD Navios Flow Cytometer K130373 9/18/2013 Class II 864.5220 OYE Navios EX Flow Cytometer K162897 6/23/2017 Class II 864.5220 OYE Device Description The new ClearLLab 10C reagent system is comprised of various…
Beckman Coulter2017—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K232600DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc.20230

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K030828Tetracxp System (Software and System Guide)Beckman Coulter, Inc. · 2003 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K990172Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome…Coulter Corp. · 1999 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K963263Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome…Coulter Corp. · 1997 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 4.0 years on median.
  • 3 of the 11 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K232600 (Beckman Coulter, Inc., 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All PWD devices

PWD Flow Cytometric Test System For Hematopoietic Neoplasms: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · PWD

1in 2019

2017: 12019: 1
2007high 1 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PWD

Too few PWD 510(k)s in any year since 2007 to give a median.

Recalls per year · PWD

No recall names PWD since 2007.

Adverse event reports per year · PWD

1in 2021

2020: 22021: 1
2007high 2 (2020) · low 0 (2007)2025

MAUDE reports naming product code PWD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20171—00
20180—00
20191—00
20200—02
20210—01
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The ClearLLab 10C Panels are intended for in vitro diagnostic use for qualitative identification of cell populations by multiparameter immunophenotyping on the Navios and Navios EX flow cytometers. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having, the following hematopoietic neoplasms: chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic…

Read the full statement (314 words)

The ClearLLab 10C Panels are intended for in vitro diagnostic use for qualitative identification of cell populations by multiparameter immunophenotyping on the Navios and Navios EX flow cytometers. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having, the following hematopoietic neoplasms: chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood (collected in K2EDTA, Acid Citrate Dextrose (ACD) or Heparin), bone marrow (collected in K2EDTA, ACD or Heparin) and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings.

These reagents provide multiparameter, qualitative results for the surface antigens listed below:

• ClearLLab 10C B Cell Tube: Kappa, Lambda, CD10, CD5, CD200, CD34, CD38, CD20, CD19, CD45

• ClearLLab 10C T Cell Tube: TCRȖį, CD4, CD2, CD56, CD5, CD34, CD3, CD8, CD7, CD45

• ClearLLab 10C M1 Cell Tube: CD16, CD7, CD10, CD13, CD64, CD34, CD14, HLA-DR, CD11b, CD45

• ClearLLab 10C M2 Cell Tube: CD15, CD123, CD117, CD13, CD33, CD34, CD38, HLA-DR, CD19, CD45

The Navios Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

The Navios EX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping using up to ten fluorescent detection channels using three lasers (488 nm, 638 nm, and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K183592: ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow Cytometer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K183592

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026