Tetracxp System (Software and System Guide)

K030828510(k) AbbreviatedCleared May 21, 200368-day reviewClass II21 CFR 864.5220Recalls 10 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on May 21, 2003 as substantially equivalent to K990172 (Coulter Corp., 1999) after a 68-day review. Its family tree reaches back 3 generations to 1993. 4 of the 13 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 10 times, including a Class I recall.

Decided
May 21, 2003
Review
68 days
Class
Class II
Product code
GKZ
Recalls
10 (1 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K922745 (1993). 6 ancestors in all. 6 later devices descend from it, 3 citing it directly. 4 of the 13 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor1993: 2 other ancestors2019: 2 descendants cleared2007: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2017: 1 descendant cleared2023: 1 descendant clearedK922745 · Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1 · Coulter Immunology · 1993K963263 · Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome CD45-Fitc/CD4-Rdi/CD8-Ecd/CD3-PC5 · Coulter Corp. · 1997 · Class I recallK990172 · Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome CD45-Fitc/CD4-RD1/CD8-Ecd/CD3-PC5 Monocla · Coulter Corp. · 1999 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 1 other ancestor 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 2 descendants3 generations forward: 1 descendant K990172 · Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome CD45-Fitc/CD4-RD1/CD8-Ecd/CD3-PC5 Monocla · Coulter Corp. · 1999 · Class I recallK963263 · Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome CD45-Fitc/CD4-Rdi/CD8-Ecd/CD3-PC5 · Coulter Corp. · 1997 · Class I recallK922745 · Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1 · Coulter Immunology · 19930
  1. Root · 3 generations backK922745 Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1Coulter Immunology · 1993
  2. 2 generations backK963263 Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat…Coulter Corp. · 1997 · Recalled (Class I)
  3. 1 generation backK990172 Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat…Coulter Corp. · 1999 · Recalled (Class I)
  4. This deviceK030828 Tetracxp System (Software and System Guide)Beckman Coulter, Inc. · 2003
  5. Descendant · cites it directly · 3 of its ownK130373 Navios Flow Cytometer SystemBeckman Coulter, Inc. · 2013
  6. Descendant · 2 generations on · 2 of its ownK162897 Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer…Beckman Coulter · 2017
  7. Descendant · 2 generations on · 1 of its ownK183592 ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios…Beckman Coulter · 2019
  8. Descendant · cites it directly · 1 of its ownK071681 Cytomics FC 500 MPL Flow Cytometer, Model # 626554Beckman Coulter, Inc. · 2007 · Recalled (Class I)
  9. Descendant · 3 generations on · 0 of its ownK232600 DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc. · 2023 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2007: 12013: 12017: 12019: 22023: 1
2007high 2 (2019) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K130373Navios Flow Cytometer SystemBeckman Coulter, Inc. · 2013 · cites it directly3descendants
  • K162897Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer, 8 Color/2 Laser…Beckman Coulter · 2017 · 2 generations on2descendants
  • K183592ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow CytometerBeckman Coulter · 2019 · 2 generations on1descendant
  • K071681Cytomics FC 500 MPL Flow Cytometer, Model # 626554Beckman Coulter, Inc. · 2007 · cites it directly1descendant
  • K232600DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc. · 2023 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K922745Cyto-Stat/Coulter Clone CD3(IGG1)-Fitc/T4-RD1Coulter Immunology1993—
−2K963263Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome…Coulter Corp.1997Class I
−1K990172Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome…Coulter Corp.1999Class I
0 (this device)K030828Tetracxp System (Software and System Guide)Beckman Coulter, Inc.2003—
+1K130373Navios Flow Cytometer SystemBeckman Coulter, Inc.2013—
+2K162897Navios EX Flow Cytometer, 6 Color/2 Laser, Navios EX Flow Cytometer, 8 Color/2…Beckman Coulter2017—
+2K183592ClearLLab 10C Panels (B, T, M1, M2), Navios Flow Cytometer, Navios EX Flow…Beckman Coulter2019—
+1K071681Cytomics FC 500 MPL Flow Cytometer, Model # 626554Beckman Coulter, Inc.2007Class I
+3K232600DxFLEX Flow Cytometer; ClearLLab 10C PanelsBeckman Coulter, Inc.20231

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200711
201310
201710
201920
202310
FIG. 1 — Lineage of K030828, 1993–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990172Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome…
Summary page 1
…oclonal Antibody Reagent and CYTO-STAT tetraCHROME CD-45-FITC/CD56-RDl/CD19-ECD/CD3-PC5 Monoclonal Antibody Reagent. FDA 510(k)Number(s): K990172 3.0 Device Description tetraCXP SYSTEM for the Cytomics FC 500 with CXP Software consists of CYTO-ST…
Coulter Corp.1999Class I

Also cleared against K990172: 4 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130373Navios Flow Cytometer SystemBeckman Coulter, Inc.20133
K071681Cytomics FC 500 MPL Flow Cytometer, Model # 626554 Class IBeckman Coulter, Inc.20071
K182886Cytomics FC 500 Series (MPL or MCL) Flow CytometerBeckman Coulter20190

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 12, 2018Class IZ-0471-2019Nonconforming Material/ComponentA manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
May 22, 2012Class IIZ-0035-2013Process controlThe recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stability claims stated in the product labeling for tetraCXP software and CYTO-STAT reagents (CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and…
Oct 11, 2007Not classifiedZ-0129-2009Software designSoftware error: This field action was initiated after Beckman Coulter determined that the CXP software does not refresh the dataset automatically under certain circumstances. The User documentation does not clearly state that the user…
Aug 20, 2007Not classifiedZ-1529-2008Software designLabeling provided by two integrated software systems are not clear enough to avoid potential demographic and sample type mismatches under certain conditions.
Aug 1, 2007Not classifiedZ-0455-2009Software designSoftware error: The recall was initiated after Beckman Coulter identified that the CXP User documentation does not adequately characterize the use of the Live Gate feature. The primary use of a Live Gate is for the international removal…
Jun 29, 2007Not classifiedZ-0959-2008Software designData generation error: If the cytosettings is not refreshed or restarted during data importation, the output will include old and new data.
Aug 29, 2006Not classifiedZ-0418-2007OtherTwo workflow scenarios associated with renaming regions may produce incorrect results.
Aug 29, 2006Not classifiedZ-0419-2007OtherTwo workflow scenarios associated with renaming regions may produce incorrect results.
Aug 29, 2006Not classifiedZ-0420-2007OtherTwo workflow scenarios associated with renaming regions may produce incorrect results.
Jul 26, 2006Not classifiedZ-0197-2007OtherMis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will generate results and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K990172Tetraone System for Epics XL Flow Cytometry Systems with Cyto-Stat Tetrachrome…Coulter Corp. · 1999 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K963263Tetraone System for Epics XL Flow Cytometry Systems and Cyto-Stat Tertrachrome…Coulter Corp. · 1997 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K071681Cytomics FC 500 MPL Flow Cytometer, Model # 626554Beckman Coulter, Inc. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 4.0 years on median.
  • 2 of the 6 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K130373 (Beckman Coulter, Inc., 2013) has gone furthest, with 3 descendants of its own.
  • 4 other devices also cite its predicate K990172.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The tetraCXP Software for Cytomics FC 500 flow cytometry systems and CYTO-STAT tetraCHROMETMCD45-FITC/CD4-RDl/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROMETM CD45-FlTC/CD56-RDl/CD19-ECD/CD3-PC5 MonoclonalAntibody Reagents combine fourcolor fluorescent monoclonal antibody reagents, quality control reagents, an optional absolute count reagent, and software for automated analysis of lymphocyte populations in whole blood using Cytomics FC 500 flow cytometry…

Read the full statement (154 words)

The tetraCXP Software for Cytomics FC 500 flow cytometry systems and CYTO-STAT tetraCHROMETMCD45-FITC/CD4-RDl/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROMETM CD45-FlTC/CD56-RDl/CD19-ECD/CD3-PC5 MonoclonalAntibody Reagents combine fourcolor fluorescent monoclonal antibody reagents, quality control reagents, an optional absolute count reagent, and software for automated analysis of lymphocyte populations in whole blood using Cytomics FC 500 flow cytometry systems with CXP Software.

The system with CMO-STAT tetraCHROME CD45-FITC/CD4-RDllCD8-ECD/CD3-PC5 is intended “For In Vitro Diagnostic Use” and allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+ T lymphocyte population percentages and absolute counts.

The system with CMO-STAT tetraCHROME CD45-FITC/CD56-RDl/CD19-ECD/CD3-PC5 is intended “For In Vitro Diagnostic Use” and allows simultaneous identification and enumeration of total CD3+ (T), CD19+ (B), and CD3-/CD56+ (NK) lymphocyte population percentages and absolute counts. This reagent reflects the distribution of the three major subsets comprising the lymphocyte population upon which other lymphocyte enumeration studies are based and provides the total lymphocyte percentage.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030828: Tetracxp System (Software and System Guide).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K030828

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026