510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™Laser Fibers, and Accessories)

K183647510(k) TraditionalCleared Aug 9, 2019226-day reviewClass II21 CFR 878.4810Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Surgical Technologies America · GEX Powered Laser Surgical Instrument

Cleared on Aug 9, 2019 as substantially equivalent to K170121 (Lumenis, Ltd., 2017) and 1 other predicate after a 226-day review. Its family tree reaches back 8 generations to 1994. 12 later devices descend from it. It has been recalled 4 times.

Decided
Aug 9, 2019
Review
226 days
Class
Class II
Product code
GEX
Recalls
4
Predicates
2

Its longest line of descent runs back 8 generations to K936109 (1994). 34 ancestors in all, the oldest K853620 (1986). 12 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor2007: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor1995: 4 other ancestors1998: 3 other ancestors1996: 2 other ancestors1994: 3 other ancestors1993: 1 other ancestor1992: 2 other ancestors1991: 3 other ancestors1990: 1 other ancestor1989: 1 other ancestor1997: 1 other ancestor1986: 1 other ancestor2025: 4 descendants cleared2021: 2 descendants cleared2023: 1 descendant cleared2026: 3 descendants cleared2024: 2 descendants clearedK936109 · Sharplan Model 1120 CO2 Surgical Laser · Sharplan Lasers, Inc. · 1994K960521 · Sharplan Lasers, Inc. Silktouch Flash Scanner · Sharplan Lasers, Inc. · 1996K963339 · Coherent Ultrapulse CO2 Surgical Lasers · Coherent Medical Group · 1997K974789 · Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and L · Lumenis, Inc. · 1998K990947 · Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO: Yag/ND: Yag Surgical Lasers and Delivery · Lumenis, Inc. · 1999K011703 · Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength (HO:Yag/ND:Yag) Surgical Lasers and Delivery Devices · Lumenis · 2001 · recalledK140388 · Lumenis Pulse 120H · Lumenis, Ltd. · 2014K170121 · Lumenis Family of Holmium Surgical Lasers and Delivery Devices and Accessories · Lumenis, Ltd. · 2017 · recalled1990200020102020
  1. Root · 8 generations backK936109 Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc. · 1994
  2. 7 generations backK960521 Sharplan Lasers, Inc. Silktouch Flash ScannerSharplan Lasers, Inc. · 1996
  3. 6 generations backK963339 Coherent Ultrapulse CO2 Surgical LasersCoherent Medical Group · 1997
  4. 5 generations backK974789 Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and LLumenis, Inc. · 1998
  5. 4 generations backK990947 Modified Versapulse Select Single Wavelength, HO: Yag and Dual…Lumenis, Inc. · 1999
  6. 3 generations backK011703 Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength…Lumenis · 2001 · Recalled
  7. 2 generations backK140388 Lumenis Pulse 120HLumenis, Ltd. · 2014
  8. 1 generation backK170121 Lumenis Family of Holmium Surgical Lasers and Delivery Devices and…Lumenis, Ltd. · 2017 · Recalled
  9. This deviceK183647 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™…Olympus Surgical Technologies America · 2019
  10. Descendant · cites it directly · 7 of its ownK211401 SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium…Gyrus Acmi, Inc. · 2021 · Recalled
  11. Descendant · cites it directly · 4 of its ownK221306 SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser…Gyrus Acmi, Inc. · 2023 · Recalled
  12. Descendant · 2 generations on · 3 of its ownK242191 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™…Olympus Surgical Technologies America · 2024 · Recalled
  13. Descendant · 3 generations on · 1 of its ownK261221 SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™…Olympus Surgical Technologies America · 2026
  14. Descendant · 2 generations on · 1 of its ownK232568 IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and…Ipg Medical Corporation · 2024 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2021: 22023: 12024: 22025: 42026: 3
2021high 4 (2025) · low 0 (2022)2026

Most influential descendants

Showing 5 of 12 · See all
  • K211401SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser…Gyrus Acmi, Inc. · 2021 · cites it directly7descendants
  • K221306SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE SuperPulsed…Gyrus Acmi, Inc. · 2023 · cites it directly4descendants
  • K242191SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed…Olympus Surgical Technologies America · 2024 · 2 generations on3descendants
  • K261221SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed…Olympus Surgical Technologies America · 2026 · 3 generations on1descendant
  • K232568IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and AccessoriesIpg Medical Corporation · 2024 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K936109Sharplan Model 1120 CO2 Surgical LaserSharplan Lasers, Inc.1994—
−7K960521Sharplan Lasers, Inc. Silktouch Flash ScannerSharplan Lasers, Inc.1996—
−6K963339Coherent Ultrapulse CO2 Surgical LasersCoherent Medical Group1997—
−5K974789Coherent Ultrapulse S Series CO2 Surgical Lasers Model C and LLumenis, Inc.1998—
−4K990947Modified Versapulse Select Single Wavelength, HO: Yag and Dual Wavelength, HO…Lumenis, Inc.1999—
−3K011703Lumenis Versapulse Powersuite Holmium (HO:Yag) and Dual Wavelength…Lumenis20012
−2K140388Lumenis Pulse 120HLumenis, Ltd.2014—
−1K170121Lumenis Family of Holmium Surgical Lasers and Delivery Devices and AccessoriesLumenis, Ltd.20171
0 (this device)K183647SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium…Olympus Surgical Technologies America2019—
+1K211401SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium…Gyrus Acmi, Inc.20211
+1K221306SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE…Gyrus Acmi, Inc.20238
+2K242191SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium…Olympus Surgical Technologies America20247
+3K261221SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium…Olympus Surgical Technologies America2026—
+2K232568IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and AccessoriesIpg Medical Corporation20241

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198910
199010
199130
199220
199310
199430
199540
199620
199710
199830
199910
200010
200710
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202120
202310
202420
202540
202630
FIG. 1 — Lineage of K183647, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEX device

  1. #1 K222064 The Alma Soprano Titaniumscore 38.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K170121Lumenis Family of Holmium Surgical Lasers and Delivery Devices and Accessories
Summary page 5
…Laser Fibers, and Accessories) Generic/Common Name: Laser Instrument for Use in General Surgery Predicate Devices Lumenis P120 K170121 StarMedTec, vela XL K103654…
Lumenis, Ltd.20171
*K103654Vela XL
Summary page 5
…on Name: Laser Instrument for Use in General Surgery Predicate Devices Lumenis P120 K170121 StarMedTec, vela XL K103654 1…
Starmedtec GmbH2011—

Also cleared against K170121: 4 other devices →

What descends from it? (12)

See all 12

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211401SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser…Gyrus Acmi, Inc.20217
K221306SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE SuperPulsed…Gyrus Acmi, Inc.20234
K252971Elyra™ Thulium Fiber Laser System and Elyra™ Plus Thulium Fiber Laser System (Laser…C.R. Bard, Inc.20260
K251303Tamburo Thulium Fiber Laser SystemMenovex Medical Technology (Shenzhen) Co., Ltd.20250
K251824Patriot DuoPatriot Laser, LLC20250
K211517Optica XT Thulium Fiber Laser and Accessories, also known as Optica XT, Optica XT…Convergent Laser Technologies20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 4, 2023Class IIZ-0716-2024Under Investigation by firmThe strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a…
Dec 4, 2023Class IIZ-0717-2024Under Investigation by firmThe strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a…
Dec 1, 2023Class IIZ-0748-2024Labeling Change ControlOlympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use…
May 26, 2022Class IIZ-1296-2022Labeling mix-upsDevice model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 2.2 years on median.
  • 2 of the 34 ancestors have a recall of any class.
  • Of its descendants, K211401 (Gyrus Acmi, Inc., 2021) has gone furthest, with 7 descendants of its own.
  • 4 other devices also cite its predicate K170121.

How does its category compare?

All GEX devices

GEX Powered Laser Surgical Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · GEX

83in 2025

2007: 732008: 1072009: 692010: 502011: 412012: 452013: 462014: 612015: 472016: 512017: 672018: 512019: 412020: 522021: 532022: 602023: 522024: 822025: 832026 (partial year): 57
2007high 107 (2008) · low 41 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEX

90in 2025

2007: 79 days median2008: 86 days median2009: 100 days median2010: 96 days median2011: 83 days median2012: 118 days median2013: 144 days median2014: 147 days median2015: 112 days median2016: 114 days median2017: 90 days median2018: 78 days median2019: 88 days median2020: 90 days median2021: 123 days median2022: 152 days median2023: 87 days median2024: 90 days median2025: 90 days median2026 (partial year): 90 days median
2007high 152 (2022) · low 78 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEX

3in 2024

2007: 92008: 22010: 42011: 182012: 52013: 32014: 62016: 12017: 5, 1 Class I2018: 52019: 62020: 12023: 62024: 32026 (partial year): 2
2007high 18 (2011) · low 0 (2009)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · GEX

2,720in 2025

2007: 3422008: 2322009: 2092010: 1,7112011: 3,0242012: 2,0002013: 2,2362014: 2,0362015: 2,0992016: 2,5852017: 2,9772018: 3,5252019: 2,7092020: 2,1832021: 3,4652022: 2,2882023: 2,3372024: 2,1872025: 2,720
2007high 3,525 (2018) · low 209 (2009)2025

MAUDE reports naming product code GEX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200773799342
2008107862232
2009691000209
2010509641,711
20114183183,024
20124511852,000
20134614432,236
20146114762,036
20154711202,099
20165111412,585
201767905 (1)2,977
2018517853,525
2019418862,709
2020529012,183
20215312303,465
20226015202,288
2023528762,337
2024829032,187
2025839002,720
202657902—

What is it indicated for?

The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. Urology Ablation of Benign Prostatic Hypertrophy (BPH) • Laser…

Read the full statement (264 words)

The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

Urology

Ablation of Benign Prostatic Hypertrophy (BPH) • Laser Resection of the Prostrate (LRP) • Laser Enucleation of the Prostate (LEP) • Laser Ablation of the Prostate (LAP) • Transurethral Incision of the Prostate (TUIP) • Condylomas • Urethral strictures • Lesions of external genitalia • Bladder neck incisions (BNI)

Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors • Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi • Treatment of distal impacted fragments remaining in the ureters following lithotripsy.

Lithotripsy and Percutaneous Urinary Lithotripsy Indications

Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones • Endoscopic fragmentation of renal calculi • Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.

Gastroenterology

Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:

Appendectomy • Angiodysplasia

Polyps • Colorectal cancer

Biopsy • Telangiectasias

Gall Bladder calculi • Telangiectasias of the Osler-Weber-Renu disease

Biliary/Bile duct calculi • Vascular Malformation

Ulcers • Gastritis

Gastric ulcers • Esophagitis

Duodenal ulcers • Esophageal ulcers

Non Bleeding Ulcers • Varices

Pancreatitis • Colitis

Haemorrhoids • Mallory-Weiss tear

Cholecystectomy • Gastric Erosions

Benign and Malignant Neoplasm

Gynecology

Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K183647: SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™Laser Fibers, and Accessories).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K183647

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026