510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

POWERSEAL Curved Jaw Sealer and Divider, Double Action

K212643510(k) SpecialCleared Sep 27, 202138-day reviewClass II21 CFR 878.4400Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Surgical Technologies America · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Sep 27, 2021 as substantially equivalent to K203682 (Olympus Winter & Ibe GmbH, 2021) after a 38-day review. Its family tree reaches back 8 generations to 1994. 1 later device descends from it. It has been recalled 7 times.

Decided
Sep 27, 2021
Review
38 days
Class
Class II
Product code
GEI
Recalls
7
Predicates
1

Its longest line of descent runs back 8 generations to K932293 (1994). 37 ancestors in all, the oldest K911904 (1991). 1 later device descends from it, 1 citing it directly.

  1. Root · 8 generations backK932293 Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp. · 1994 · Recalled
  2. 7 generations backK981916 Ligasure Vessel Sealing SystemValleylab, Inc. · 1998
  3. 6 generations backK031011 Ligasure 5MM Laparoscopic Sealer-Divider, Model LS1500Valleylab, Inc. · 2003 · Recalled
  4. 5 generations backK102470 Lugasure Curved, Small Jaw, Open Sealer/DividerCovidien, Formerly Valleylab, · 2011
  5. 4 generations backK113572 Ligasure Curved, Small Jaw, Open Sealer/DividerCovidien, Formerly Valleylab, A Division of Tyco H · 2012
  6. 3 generations backK133338 Ligasure 5 MM Maryland Jaw Sealer/Divider One-Step Sealing (LF 17XX)Covidien, Formerly Valleylab, A Division of Tyco H · 2013
  7. 2 generations backK141153 Ligasure; Curved Small Jaw Sealer/Divider, 5MM Blunt Tip…Covidien · 2014
  8. 1 generation backK203682 Electrosurgical Generator ESG-400, Foot Switches, POWERSEAL Curved…Olympus Winter & Ibe GmbH · 2021
  9. This deviceK212643 POWERSEAL Curved Jaw Sealer and Divider, Double ActionOlympus Surgical Technologies America · 2021
  10. Descendant · cites it directly · 0 of its ownK231327 POWERSEAL Sealer and DividerGyrus Acmi, Inc. · 2023

Most influential descendants

Showing 1 of 1 · See all
  • K231327POWERSEAL Sealer and DividerGyrus Acmi, Inc. · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K932293Cabot Medical Bipolar Cutting ForcepsCabot Medical Corp.19943
−7K981916Ligasure Vessel Sealing SystemValleylab, Inc.1998—
−6K031011Ligasure 5MM Laparoscopic Sealer-Divider, Model LS1500Valleylab, Inc.20032
−5K102470Lugasure Curved, Small Jaw, Open Sealer/DividerCovidien, Formerly Valleylab,2011—
−4K113572Ligasure Curved, Small Jaw, Open Sealer/DividerCovidien, Formerly Valleylab, A Division of Tyco H2012—
−3K133338Ligasure 5 MM Maryland Jaw Sealer/Divider One-Step Sealing (LF 17XX)Covidien, Formerly Valleylab, A Division of Tyco H2013—
−2K141153Ligasure; Curved Small Jaw Sealer/Divider, 5MM Blunt Tip Sealer/Divider…Covidien2014—
−1K203682Electrosurgical Generator ESG-400, Foot Switches, POWERSEAL Curved Jaw Sealer…Olympus Winter & Ibe GmbH2021—
0 (this device)K212643POWERSEAL Curved Jaw Sealer and Divider, Double ActionOlympus Surgical Technologies America2021—
+1K231327POWERSEAL Sealer and DividerGyrus Acmi, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
199540
199620
199720
199930
200120
200220
200310
200510
200610
200730
200910
201120
201320
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K212643, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K203682Electrosurgical Generator ESG-400, Foot Switches, POWERSEAL Curved Jaw Sealer and…
Summary page 4
…device and Accessories Predicate Devices The POWERSEAL Curved Jaw Sealer and Divider, Double Action (PS-0523CJDA, PS-0537CJDA, and PS-0544CJDA) were cleared under K203682. This submission is 1 === Page 5…
Olympus Winter & Ibe GmbH2021—

Also cleared against K203682: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231327POWERSEAL Sealer and DividerGyrus Acmi, Inc.20230

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2026Class IIZ-3035-2026Labeling Change ControlPotential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
Mar 7, 2024Class IIZ-1699-2024Process controlUpon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting…
Mar 7, 2024Class IIZ-1700-2024Process controlUpon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting…
Mar 7, 2024Class IIZ-1701-2024Process controlUpon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting…
Aug 12, 2022Class IIZ-1813-2022Under Investigation by firmAn out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
Aug 12, 2022Class IIZ-1815-2022Under Investigation by firmAn out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
Aug 12, 2022Class IIZ-1814-2022Under Investigation by firmAn out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 7 companies; a new generation was cleared every 4.1 years on median.
  • 8 of the 37 ancestors have a recall of any class.
  • Of its descendants, K231327 (Gyrus Acmi, Inc., 2023) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K203682.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and…

Read the full statement (125 words)

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy.

The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K212643: POWERSEAL Curved Jaw Sealer and Divider, Double Action.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K212643

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026