510(k) Report

510(k) · Cleared

Cleared

MTS225 and MTS300 Multiple Tray Sterilization Systems

K190541510(k) TraditionalCleared Nov 5, 2019246-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pmbs, LLC · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Nov 5, 2019 as substantially equivalent to K152014 (Pmbs, LLC, 2015) after a 246-day review. Its family tree reaches back 2 generations to 2012. 1 later device descends from it.

Decided
Nov 5, 2019
Review
246 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K110898 (2012). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2022: 1 descendant clearedK110898 · S.C.O.R.E.S. Unit · Ammed Surgical Equipment, LLC · 2012 · recalledK152014 · MTS300 Multiple Tray Sterilization System · Pmbs, LLC · 20152010201520202025
Descendants, by generation 1 generation forward: 1 descendant K152014 · MTS300 Multiple Tray Sterilization System · Pmbs, LLC · 2015K110898 · S.C.O.R.E.S. Unit · Ammed Surgical Equipment, LLC · 2012 · recalled0
  1. Root · 2 generations backK110898 S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC · 2012 · Recalled
  2. 1 generation backK152014 MTS300 Multiple Tray Sterilization SystemPmbs, LLC · 2015
  3. This deviceK190541 MTS225 and MTS300 Multiple Tray Sterilization SystemsPmbs, LLC · 2019
  4. Descendant · cites it directly · 0 of its ownK222328 SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC · 2022

Most influential descendants

Showing 1 of 1 · See all
  • K222328SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K110898S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC20121
−1K152014MTS300 Multiple Tray Sterilization SystemPmbs, LLC2015—
0 (this device)K190541MTS225 and MTS300 Multiple Tray Sterilization SystemsPmbs, LLC2019—
+1K222328SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K190541, 2012–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152014MTS300 Multiple Tray Sterilization System
Summary page 4
…Device MTS225 and MTS300 Multiple Tray Sterilization Systems Predicate MTS300 Multiple Tray Sterilization System (“MTS300 System”) Device K152014 Device The MTS System is comprised of the stainless steel sterilization cabinet and Description adjustable tra…
Pmbs, LLC2015—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K222328SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC20220

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 4.0 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K222328 (Stericube, LLC, 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The MTS System (Multiple Tray Sterilization System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days. The unit must be used with the MTS System Transfer Cart, MTS System Filters and MTS System Integrity Locks. The MTS System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle…

Read the full statement (189 words)

The MTS System (Multiple Tray Sterilization System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days. The unit must be used with the MTS System Transfer Cart, MTS System Filters and MTS System Integrity Locks.

The MTS System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270° F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 lbs. per tray.

The MTS System was tested and validated with rigid instruments containing lumens with an inner diameter of .7mm and an overall length of 500mm as well as lumens with an inner diameter of 1.0mm and an overall length of 850mm. Do not use with instruments containing lumens with an inner diameter smaller than .7mm and an overall length longer than 500mm or lumens with an inner diameter smaller than 1.0mm and an overall length longer than 850mm.

Use only uncovered, perforated or wire mesh general delivery trays within the MTS System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K190541: MTS225 and MTS300 Multiple Tray Sterilization Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K190541

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026