510(k) Report

510(k) · Cleared

Cleared

SteriCUBE® Multiple Tray Sterilization Systems

K222328510(k) TraditionalCleared Sep 28, 202257-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stericube, LLC · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Sep 28, 2022 as substantially equivalent to K190541 (Pmbs, LLC, 2019) after a 57-day review. Its family tree reaches back 3 generations to 2012.

Decided
Sep 28, 2022
Review
57 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K110898 (2012). No later device cites it as a predicate yet.

  1. Root · 3 generations backK110898 S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC · 2012 · Recalled
  2. 2 generations backK152014 MTS300 Multiple Tray Sterilization SystemPmbs, LLC · 2015
  3. 1 generation backK190541 MTS225 and MTS300 Multiple Tray Sterilization SystemsPmbs, LLC · 2019
  4. This deviceK222328 SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K110898S.C.O.R.E.S. UnitAmmed Surgical Equipment, LLC20121
−2K152014MTS300 Multiple Tray Sterilization SystemPmbs, LLC2015—
−1K190541MTS225 and MTS300 Multiple Tray Sterilization SystemsPmbs, LLC2019—
0 (this device)K222328SteriCUBE® Multiple Tray Sterilization SystemsStericube, LLC2022—
FIG. 1 — Lineage of K222328, 2012–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K190541MTS225 and MTS300 Multiple Tray Sterilization Systems
Summary page 4
…ray Sterilization Systems Predicate MTS225 and MTS300 Multiple Tray Sterilization Systems (“MTS Systems”) Device cleared most recently November 5, 2019, via K190541 Device The SteriCUBE® Multiple Tray Sterilization Systems are also referred to as Description the “SteriCUBE Systems”…
Pmbs, LLC2019—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 3.2 years on median.
  • 1 of the 3 ancestors has a recall of any class.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The SteriCUBE Multiple Tray Sterilization System (SteriCUBE System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SteriCUBE System Transfer Cart, SteriCUBE System Filters and SteriCUBE System Integrity Locks. The SteriCUBE System is intended to be used in…

Read the full statement (191 words)

The SteriCUBE Multiple Tray Sterilization System (SteriCUBE System) is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit must be used with the SteriCUBE System Transfer Cart, SteriCUBE System Filters and SteriCUBE System Integrity Locks.

The SteriCUBE System is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F and exposure time of 4 minutes. Use no more than 3 trays per shelf or 25 Ibs. per tray.

The SteriCUBE System was tested and validated with rigid instruments containing lumens with an inner diameter of 0.7mm and an overall length of 500mm as well as lumens with an inner diameter of 1.0mm and an overall length of 850mm. Do not use with instruments containing lumens with an inner diameter smaller than 0.7mm and an overall length longer than 500mm or lumens with an inner diameter smaller than 1.0mm and an overall length longer than 850mm.

Use only uncovered, perforated or wire mesh general delivery trays within the SteriCUBE System.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K222328: SteriCUBE® Multiple Tray Sterilization Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K222328

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026