510(k) · Cleared
ClearedQuantum® Total Ankle Prosthesis
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
In2bones Sas · HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer
Cleared on Jan 29, 2020 as substantially equivalent to K090076 (Tornier, 2009) and 3 other predicates after a 251-day review. Its family tree reaches back 7 generations to 1992.
- Decided
- Jan 29, 2020
- Review
- 251 days
- Class
- Class II
- Product code
- HSN
- Recalls
- None
- Predicates
- 4
Its longest line of descent runs back 7 generations to K920802 (1992). 25 ancestors in all. No later device cites it as a predicate yet.
- Root · 7 generations backK920802 Alvine Total AnkleDepuy, Inc. · 1992
- 6 generations backK020541 Agility Ankle Revision ProsthesisDepuy, Inc. · 2002
- 5 generations backK051023 Topez Total Ankle ReplacementTopez Orthopedics, Inc. · 2005 · Recalled
- 4 generations backK100886 Inbone II Total Ankle ReplacementWrightmedicaltechnologyinc · 2010 · Recalled
- 3 generations backK123954 Infinity Total Ankle SystemWrightmedicaltechnologyinc · 2013 · Recalled
- 2 generations backK140749 Infinity Total Ankle SystemWrightmedicaltechnologyinc · 2014 · Recalled
- 1 generation backK152217 Exactech Vantage Total Ankle SystemExactech, Inc. · 2016 · Recalled
- This deviceK191380 Quantum® Total Ankle ProsthesisIn2bones Sas · 2020
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K920802 | Alvine Total Ankle | Depuy, Inc. | 1992 | — |
| −6 | K020541 | Agility Ankle Revision Prosthesis | Depuy, Inc. | 2002 | — |
| −5 | K051023 | Topez Total Ankle Replacement | Topez Orthopedics, Inc. | 2005 | 4 |
| −4 | K100886 | Inbone II Total Ankle Replacement | Wrightmedicaltechnologyinc | 2010 | 2 |
| −3 | K123954 | Infinity Total Ankle System | Wrightmedicaltechnologyinc | 2013 | 2 |
| −2 | K140749 | Infinity Total Ankle System | Wrightmedicaltechnologyinc | 2014 | 1 |
| −1 | K152217 | Exactech Vantage Total Ankle System | Exactech, Inc. | 2016 | 6 |
| 0 (this device) | K191380 | Quantum® Total Ankle Prosthesis | In2bones Sas | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1995 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2000 | 2 | 0 |
| 2002 | 2 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2006 | 3 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HSN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HSN device
- #1 K192994 Paragon 28 APEX 3D Total Ankle Replacement…score 16.7 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K090076 | Salto Talaris Total Ankle ProsthesisSummary page 5…te: In the United States, the ankle prosthesis is intended for cement use only. Predicate Primary predicate: Devices - Salto Talaris (K090076 & K153452), Tornier Additional predicates: - Cadence / Integra Total Ankle Replacement System… | 2009 |
| * | K153452 | Salto XT, Salto TalarisSummary page 5…United States, the ankle prosthesis is intended for cement use only. Predicate Primary predicate: Devices - Salto Talaris (K090076 & K153452), Tornier Additional predicates: - Cadence / Integra Total Ankle Replacement System (K151459)… | 2016 |
| * | K151459 | Integra Total Ankle Replacement SystemSummary page 5…ces - Salto Talaris (K090076 & K153452), Tornier Additional predicates: - Cadence / Integra Total Ankle Replacement System (K151459), Ascension Orthopedics (acquired by Integra LifeSciences) - Vantage (K152217), Exactech Comp… | 2015 |
| * | K152217 | Exactech Vantage Total Ankle SystemSummary page 5…- Cadence / Integra Total Ankle Replacement System (K151459), Ascension Orthopedics (acquired by Integra LifeSciences) - Vantage (K152217), Exactech Comparison of The In2Bones QUANTUM® total ankle prosthesis is similar to the Indications and predicate devices S… | 2016 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 28 years and passes through 5 companies; a new generation was cleared every 3.5 years on median.
- 16 of the 25 ancestors have a recall of any class.
- 4 other devices also cite its predicate K090076.
How does its category compare?
All HSN devicesHSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.
FDA decisions per year · HSN
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HSN
58in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HSN
1in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HSN
1,645in 2025
MAUDE reports naming product code HSN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 50 |
| 2008 | 0 | — | 0 | 66 |
| 2009 | 1 | — | 0 | 66 |
| 2010 | 2 | — | 0 | 120 |
| 2011 | 1 | — | 0 | 139 |
| 2012 | 2 | — | 0 | 100 |
| 2013 | 4 | 89 | 3 | 106 |
| 2014 | 2 | — | 0 | 65 |
| 2015 | 3 | 233 | 0 | 73 |
| 2016 | 3 | 216 | 1 | 111 |
| 2017 | 4 | 148 | 2 | 96 |
| 2018 | 4 | 89 | 0 | 187 |
| 2019 | 2 | — | 1 | 277 |
| 2020 | 6 | 234 | 0 | 281 |
| 2021 | 4 | 109 | 3 | 237 |
| 2022 | 1 | — | 2 | 453 |
| 2023 | 6 | 112 | 1 | 796 |
| 2024 | 7 | 115 | 0 | 1,095 |
| 2025 | 3 | 58 | 0 | 1,645 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
The Quantum® total ankle prosthesis is indicated as a total ankle replacement in primary or revision surgery for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis.
Note: In the United States, the ankle prosthesis is intended for cement use only.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K191380 · printed Nov 6, 2026
Cite this page
510(k) Report. “K191380: Quantum® Total Ankle Prosthesis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191380
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026