510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Salto Talaris Total Ankle Prosthesis, RHS

K130533510(k) SpecialCleared May 17, 201377-day reviewClass II21 CFR 888.3110Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tornier S.A.S. · HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer

Cleared on May 17, 2013 as substantially equivalent to K060544 (Tornier, 2006) and 8 other predicates after a 77-day review. Its family tree reaches back 5 generations to 1999. 16 later devices descend from it. It has been recalled 19 times.

Decided
May 17, 2013
Review
77 days
Class
Class II
Product code
HSN
Recalls
19
Predicates
9

Its longest line of descent runs back 5 generations to K984290 (1999). 16 ancestors in all, the oldest K920802 (1992). 16 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2006: 1 other ancestor2005: 1 other ancestor2002: 2 other ancestors2000: 2 other ancestors1992: 1 other ancestor1999: 1 other ancestor1998: 1 other ancestor1995: 1 other ancestor2016: 1 descendant cleared2017: 1 descendant cleared2020: 3 descendants cleared2018: 1 descendant cleared2023: 2 descendants cleared2024: 4 descendants cleared2021: 2 descendants cleared2025: 1 descendant cleared2022: 1 descendant clearedK984290 · Implex Radial Head Replacement System · Implex Corp. · 1999K012551 · Liverpool Radial Head Replacement Device · Biomet, Inc. · 2002K040611 · Modular Radial Head Replacement Device · Biomet, Inc. · 2004 · recalledK051385 · Explor Modular Radial Head Replacement Device · Biomet, Inc. · 2005 · recalledK060438 · RHS · Tornier · 20061990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 7 other ancestors2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 6 descendants4 generations forward: 2 descendants5 generations forward: 1 descendant K060438 · RHS · Tornier · 2006K051385 · Explor Modular Radial Head Replacement Device · Biomet, Inc. · 2005 · recalledK040611 · Modular Radial Head Replacement Device · Biomet, Inc. · 2004 · recalledK012551 · Liverpool Radial Head Replacement Device · Biomet, Inc. · 2002K984290 · Implex Radial Head Replacement System · Implex Corp. · 1999−50+5
  1. Root · 5 generations backK984290 Implex Radial Head Replacement SystemImplex Corp. · 1999
  2. 4 generations backK012551 Liverpool Radial Head Replacement DeviceBiomet, Inc. · 2002
  3. 3 generations backK040611 Modular Radial Head Replacement DeviceBiomet, Inc. · 2004 · Recalled
  4. 2 generations backK051385 Explor Modular Radial Head Replacement DeviceBiomet, Inc. · 2005 · Recalled
  5. 1 generation backK060438 RHSTornier · 2006
  6. This deviceK130533 Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S. · 2013
  7. Descendant · cites it directly · 15 of its ownK153452 Salto XT, Salto TalarisTornier S.A.S. · 2016 · Recalled
  8. Descendant · 2 generations on · 8 of its ownK192994 Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc. · 2020
  9. Descendant · 3 generations on · 2 of its ownK210390 Apex 3D Total Ankle Replacement SystemParagon 28, Inc. · 2021
  10. Descendant · cites it directly · 2 of its ownK171004 Hintermann Series H2 Total Ankle SystemDt Medtech, LLC · 2017
  11. Descendant · 4 generations on · 1 of its ownK231231 MR Safety and Compatibility Testing and Labeling for Paragon 28…Paragon 28, Inc. · 2023

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2016: 12017: 12018: 12020: 32021: 22022: 12023: 22024: 42025: 1
2016high 4 (2024) · low 0 (2019)2026

Most influential descendants

Showing 5 of 16 · See all
  • K153452Salto XT, Salto TalarisTornier S.A.S. · 2016 · cites it directly15descendants
  • K192994Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc. · 2020 · 2 generations on8descendants
  • K210390Apex 3D Total Ankle Replacement SystemParagon 28, Inc. · 2021 · 3 generations on2descendants
  • K171004Hintermann Series H2 Total Ankle SystemDt Medtech, LLC · 2017 · cites it directly2descendants
  • K231231MR Safety and Compatibility Testing and Labeling for Paragon 28 Orthopedic Fixation…Paragon 28, Inc. · 2023 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K984290Implex Radial Head Replacement SystemImplex Corp.1999—
−4K012551Liverpool Radial Head Replacement DeviceBiomet, Inc.2002—
−3K040611Modular Radial Head Replacement DeviceBiomet, Inc.20041
−2K051385Explor Modular Radial Head Replacement DeviceBiomet, Inc.20052
−1K060438RHSTornier2006—
0 (this device)K130533Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S.2013—
+1K153452Salto XT, Salto TalarisTornier S.A.S.20168
+2K192994Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc.2020—
+3K210390Apex 3D Total Ankle Replacement SystemParagon 28, Inc.2021—
+1K171004Hintermann Series H2 Total Ankle SystemDt Medtech, LLC2017—
+4K231231MR Safety and Compatibility Testing and Labeling for Paragon 28 Orthopedic…Paragon 28, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199510
199810
199910
200020
200220
200510
200610
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201710
201810
202030
202120
202210
202320
202440
202510
FIG. 1 — Lineage of K130533, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HSN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HSN device

  1. #1 K192994 Paragon 28 APEX 3D Total Ankle Replacement…score 16.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (9)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060544Salto Talaris Total Ankle Prosthesis
Summary page 2
…code: JOB & KWI 6) Equivalent / Predicate device For the Salto Talaris Total Ankle Prosthesis: Salto Talaris Total Ankle Prosthesis, Tornier. K060544, K090076 Alvine Total Ankle Prosthesis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthop…
Tornier200618
*K090076Salto Talaris Total Ankle Prosthesis
Summary page 2
…JOB & KWI 6) Equivalent / Predicate device For the Salto Talaris Total Ankle Prosthesis: Salto Talaris Total Ankle Prosthesis, Tornier. K060544, K090076 Alvine Total Ankle Prosthesis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, In…
Tornier200918
*K920802Alvine Total Ankle
Summary page 2
…For the Salto Talaris Total Ankle Prosthesis: Salto Talaris Total Ankle Prosthesis, Tornier. K060544, K090076 Alvine Total Ankle Prosthesis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis,…
Depuy, Inc.1992—
*K020541Agility Ankle Revision Prosthesis
Summary page 2
…alto Talaris Total Ankle Prosthesis: Salto Talaris Total Ankle Prosthesis, Tornier. K060544, K090076 Alvine Total Ankle Prosthesis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier,…
Depuy, Inc.2002—
*K051023Topez Total Ankle Replacement
Summary page 2
…rosthesis, Tornier. K060544, K090076 Alvine Total Ankle Prosthesis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier, K994041, K060438 Radial Head, Avanta, K023604 Explor, Biom…
Topez Orthopedics, Inc.20054
*K994041Tornier Bipolar Radial Head Prosthesis
Summary page 2
…esis (Agility), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier, K994041, K060438 Radial Head, Avanta, K023604 Explor, Biomet. K051385 7) Device description For the Salto Talar…
Tornier, Inc.2000—
*K060438RHS
Summary page 2
…lity), DePuy. K920802, K020541 Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier, K994041, K060438 Radial Head, Avanta, K023604 Explor, Biomet. K051385 7) Device description For the Salto Talaris Total…
Tornier2006—
*K023604Modification to Radial Head Implant
Summary page 2
…Topez Total Ankle Replacement, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier, K994041, K060438 Radial Head, Avanta, K023604 Explor, Biomet. K051385 7) Device description For the Salto Talaris Total Ankle Prosthesis: The goal of…
Avanta Orthopaedics, Inc.200218
*K051385Explor Modular Radial Head Replacement Device
Summary page 2
…t, Topez Orthopedics, Inc., K05 1023 For the RHS: Radial Head Prosthesis, Tornier, K994041, K060438 Radial Head, Avanta, K023604 Explor, Biomet. K051385 7) Device description For the Salto Talaris Total Ankle Prosthesis: The goal of total ankle replacement is to…
Biomet, Inc.20052

Also cleared against K060544: 4 other devices →

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153452Salto XT, Salto TalarisTornier S.A.S.201615
K171004Hintermann Series H2 Total Ankle SystemDt Medtech, LLC20172

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 8, 2019Class IIZ-1163-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1152-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1160-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1165-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1153-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1162-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1166-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1159-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1156-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1167-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1164-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1151-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1161-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1150-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1158-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1157-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1155-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Mar 8, 2019Class IIZ-1154-2019OtherThe Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Aug 6, 2015Class IIZ-2465-2015Nonconforming Material/ComponentTornier has initiated a voluntary recall for the Salto Talaris Tibial Guide (part number MJU333) due to a complaint that occurred during physician training, that the guide cannot be assembled to the Salto Talaris Talar Pin Setting guide.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 9 of the 16 ancestors have a recall of any class.
  • Of its descendants, K153452 (Tornier S.A.S., 2016) has gone furthest, with 15 descendants of its own.
  • 4 other devices also cite its predicate K060544.

How does its category compare?

All HSN devices

HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · HSN

3in 2025

2009: 12010: 22011: 12012: 22013: 42014: 22015: 32016: 32017: 42018: 42019: 22020: 62021: 42022: 12023: 62024: 72025: 32026 (partial year): 2
2007high 7 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HSN

58in 2025

2013: 89 days median2015: 233 days median2016: 216 days median2017: 148 days median2018: 89 days median2020: 234 days median2021: 109 days median2023: 112 days median2024: 115 days median2025: 58 days median
2007high 234 (2020) · low 58 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HSN

1in 2023

2013: 32016: 12017: 22019: 12021: 32022: 22023: 1
2007high 3 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HSN

1,645in 2025

2007: 502008: 662009: 662010: 1202011: 1392012: 1002013: 1062014: 652015: 732016: 1112017: 962018: 1872019: 2772020: 2812021: 2372022: 4532023: 7962024: 1,0952025: 1,645
2007high 1,645 (2025) · low 50 (2007)2025

MAUDE reports naming product code HSN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—050
20080—066
20091—066
20102—0120
20111—0139
20122—0100
20134893106
20142—065
20153233073
201632161111
20174148296
20184890187
20192—1277
202062340281
202141093237
20221—2453
202361121796
2024711501,095
202535801,645
20262—0—

What is it indicated for?

Salto Talaris Total Ankle Prosthesis: The Salto Talaris Total Ankle Prosthesis is indicated as a total ankle replacement in primary or revision surgery for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. Components are intended for cemented use only. RHS: The RHS is intended for: 1) Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation and decreased…

Read the full statement (134 words)

Salto Talaris Total Ankle Prosthesis:

The Salto Talaris Total Ankle Prosthesis is indicated as a total ankle replacement in primary or revision surgery for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis.

Components are intended for cemented use only.

RHS:

The RHS is intended for:

1) Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with:

a. Joint destruction and/or subluxation visible on x-ray

b. Resistance to conservative treatment

2) Primary replacement after fracture of the radial head

3) Symptomatic sequelae after radial head resection

4) Revision following failed radial head arthroplasty

The long stem is for single cemented use only. The short stem coated with titanium plasma-spray is for single use with or without cement.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130533: Salto Talaris Total Ankle Prosthesis, RHS.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130533

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026