510(k) Report

510(k) · Cleared

Cleared

Biopsy Needle

K191472510(k) TraditionalCleared Jan 29, 2020240-day reviewClass II21 CFR 876.1075

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ushare Medical, Inc. · FCG Biopsy Needle

Cleared on Jan 29, 2020 as substantially equivalent to K141552 (Care Fusion, 2014) after a 240-day review. Its family tree reaches back 3 generations to 1993.

Decided
Jan 29, 2020
Review
240 days
Class
Class II
Product code
FCG
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K922939 (1993). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK922939 Bard Monopty Disposable Core Biopsy Isnt. W/ Cent.C.R. Bard, Inc. · 1993
  2. 2 generations backK133948 Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R)…Bard Peripheral Vascular, Inc. · 2014 · Recalled
  3. 1 generation backK141552 Achieve Programmable Auomated Biopsy SystemsCare Fusion · 2014
  4. This deviceK191472 Biopsy NeedleUshare Medical, Inc. · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K922939Bard Monopty Disposable Core Biopsy Isnt. W/ Cent.C.R. Bard, Inc.1993—
−2K133948Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R)…Bard Peripheral Vascular, Inc.20147
−1K141552Achieve Programmable Auomated Biopsy SystemsCare Fusion2014—
0 (this device)K191472Biopsy NeedleUshare Medical, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
FIG. 1 — Lineage of K191472, 1993–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FCG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FCG device

  1. #1 K250994 ClearTip FNA and FNB Typesscore 13.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141552Achieve Programmable Auomated Biopsy Systems
Summary page 4
…sification regulation: 21 CFR 876.1075 Classification: Class II. Regulation Medical Specialty: Gastroenterology/Urology Product code: KNW 3. Predicate device: K number: K141552 Company: CareFusion Address: 75 North Fairway Drive, Vernon Hills, IL 60061 USA Predicate Device: Achieve Programmable Automatic…
Care Fusion2014—

Also cleared against K141552: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 4 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • 4 other devices also cite its predicate K141552.

How does its category compare?

All FCG devices

FCG Biopsy Needle: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · FCG

2in 2025

2007: 42008: 12009: 42010: 32011: 12013: 12014: 52015: 42016: 32017: 32018: 62020: 12021: 12022: 22023: 42024: 12025: 2
2007high 6 (2018) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FCG

82in 2023

2007: 32 days median2009: 87 days median2010: 18 days median2014: 88 days median2015: 91 days median2016: 63 days median2017: 166 days median2018: 140 days median2023: 82 days median
2007high 166 (2017) · low 18 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FCG

5in 2021

2008: 192009: 52010: 42011: 22021: 5
2007high 19 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FCG

632in 2025

2007: 112008: 1152009: 2702010: 732011: 712012: 1072013: 1302014: 1542015: 1242016: 1912017: 2732018: 2822019: 2292020: 2402021: 2902022: 3022023: 4282024: 4462025: 632
2007high 632 (2025) · low 11 (2007)2025

MAUDE reports naming product code FCG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007432011
20081—19115
20094875270
2010318473
20111—271
20120—0107
20131—0130
20145880154
20154910124
20163630191
201731660273
201861400282
20190—0229
20201—0240
20211—5290
20222—0302
20234820428
20241—0446
20252—0632
20260—0—

What is it indicated for?

The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone. The Biopsy Needle is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The…

Read the full statement (133 words)

The Biopsy Needle is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue masses. Not intended for use in bone.

The Biopsy Needle is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K191472: Biopsy Needle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191472

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026