510(k) Report

510(k) · Cleared

Cleared

DR 100s

K191884510(k) TraditionalCleared Aug 9, 201925-day reviewClass II21 CFR 892.1720

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Agfa N.V. · IZL System, X-Ray, Mobile

Cleared on Aug 9, 2019 as substantially equivalent to K103597 (Agfa Healthcare N.V., 2011) after a 25-day review. Its family tree reaches back 7 generations to 1998.

Decided
Aug 9, 2019
Review
25 days
Class
Class II
Product code
IZL
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K981556 (1998). 31 ancestors in all, the oldest K803130 (1981). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2009: 2 other ancestors2003: 1 other ancestor2001: 2 other ancestors2008: 1 other ancestor2000: 1 other ancestor1999: 2 other ancestors1998: 2 other ancestors2007: 1 other ancestor2002: 1 other ancestor1997: 2 other ancestors1995: 2 other ancestors1994: 2 other ancestors2006: 1 other ancestor1992: 1 other ancestor1986: 1 other ancestor1988: 1 other ancestor1981: 1 other ancestorK981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K043002 · Sedecal Mobile X-Ray Units with Digital Detector · Sedecal., Sa. · 2004K090322 · Easy Moving Plus, SM-XXHF-YY · Sedecal., Sa. · 2009 · recalledK103597 · DX-D 100 · Agfa Healthcare N.V. · 2011 · recalled19801990200020102020
  1. Root · 7 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 6 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 5 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 4 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 3 generations backK043002 Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa. · 2004
  6. 2 generations backK090322 Easy Moving Plus, SM-XXHF-YYSedecal., Sa. · 2009 · Recalled
  7. 1 generation backK103597 DX-D 100Agfa Healthcare N.V. · 2011 · Recalled
  8. This deviceK191884 DR 100sAgfa N.V. · 2019
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−6K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−5K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−4K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−3K043002Sedecal Mobile X-Ray Units with Digital DetectorSedecal., Sa.2004—
−2K090322Easy Moving Plus, SM-XXHF-YYSedecal., Sa.200931
−1K103597DX-D 100Agfa Healthcare N.V.20116
0 (this device)K191884DR 100sAgfa N.V.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198810
199210
199420
199520
199720
199820
199920
200010
200120
200210
200310
200610
200710
200810
200920
FIG. 1 — Lineage of K191884, 1998–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103597DX-D 100
Summary page 4
…892.1720 Product Code: IZL III. PREDICATE DEVICES This is a 510(k) for Agfa’s DR 100s, which mobile x-ray system. It is substantially equivalent to Agfa’s DX-D 100 (K103597). Name of Device: DX-D 100 Common Name: System, X-Ray, Mobile Classification Name: Mobile X-ray System Regulatory Classification…
Agfa Healthcare N.V.20116

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 1.8 years on median.
  • 16 of the 31 ancestors have a recall of any class.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

The DR 100s system is a mobile X-ray imaging system used in hospitals, clinics and medical practices by radiographers and radiologists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

Applications can be performed with the patient in the sitting, standing or lying position.

This device is not intended for mammography applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191884: DR 100s.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191884

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026