510(k) Report

510(k) · Cleared

Cleared

KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System

K193320510(k) TraditionalCleared Feb 4, 202064-day reviewClass II21 CFR 888.3080

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Kyocera Medical Technologies, Inc. · OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

Cleared on Feb 4, 2020 as substantially equivalent to K131122 (Renovis Surgical Technologies, LLC, 2013) and 2 other predicates after a 64-day review. Its family tree reaches back 16 generations to 1984. 3 later devices descend from it.

Decided
Feb 4, 2020
Review
64 days
Class
Class II
Product code
OVD
Recalls
None
Predicates
3

Its longest line of descent runs back 16 generations to K841189 (1984). 70 ancestors in all. 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2011: 1 other ancestor2009: 2 other ancestors2008: 10 other ancestors2007: 7 other ancestors2005: 5 other ancestors1999: 2 other ancestors1996: 2 other ancestors2006: 1 other ancestor2004: 7 other ancestors2003: 5 other ancestors2002: 5 other ancestors2001: 5 other ancestors2000: 1 other ancestor2022: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedReference device: K181655 · Renovis S180 Lateral Lumbar Interbody Fusion System · Renovis Surgical Technologies · 2018K841189 · Oec/Rezaian Spinal Fixator · Orthopedic Equipment Co., Inc. · 1984K990148 · Stackable Cage System · Depuy Motech Acromed · 1999K003043 · Surgical Titanium Mesh System · Depuyacromed · 2001×2: 2 generations cleared close together, K022793 (2003), K031757 (2003) ×2: 2 generations cleared close together, K041888 (2004), K042930 (2004) K062759 · Paramount VBR System · Innovative Spinal Technologies, Inc. · 2007K072120 · Paramount Intervertebral Body Fusion Device · Innovative Spinal Technologies, Inc. · 2007×2: 2 generations cleared close together, K080083 (2008), K073109 (2008) K083475 · Lucent Magnum · Spinal Elements, Inc. · 2009K092211 · Chesapeake Spinal System · K2m, Inc. · 2010K131122 · S128 Anterior Lumbar Interbody Fusion (Alif) System · Renovis Surgical Technologies, LLC · 2013K140106 · S128 Anterior Lumbar Interbody Fusion (Alif) System · Renovis Surgical Technologies, Inc. · 2014K180502 · S128 Anterior Lumbar Interbody Fusion (ALIF) System · Renovis Surgical Technologies, Inc. · 201819801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 6 other ancestors3 generations back: 10 other ancestors4 generations back: 19 other ancestors5 generations back: 11 other ancestors6 generations back: 4 other ancestors7 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants K180502 · S128 Anterior Lumbar Interbody Fusion (ALIF) System · Renovis Surgical Technologies, Inc. · 2018K140106 · S128 Anterior Lumbar Interbody Fusion (Alif) System · Renovis Surgical Technologies, Inc. · 2014K131122 · S128 Anterior Lumbar Interbody Fusion (Alif) System · Renovis Surgical Technologies, LLC · 2013K092211 · Chesapeake Spinal System · K2m, Inc. · 2010K083475 · Lucent Magnum · Spinal Elements, Inc. · 2009K073109 · Stalif TT Intervertebral Body Fusion System · Surgicraft, Ltd. · 2008K080083 · Intrepid Spinal System · Medtronic Sofamor Danek · 2008 · recalledK072120 · Paramount Intervertebral Body Fusion Device · Innovative Spinal Technologies, Inc. · 2007K062759 · Paramount VBR System · Innovative Spinal Technologies, Inc. · 2007K042930 · Spacevision Cage System · Spine Vision, Inc. · 2004K041888 · Curved Peek Tetris Spinal Implant · Signus Medizintechnik GmbH · 2004K031757 · Peek Tetris Spinal Implant · Signus Medical, LLC · 2003K022793 · Tetris Spinal Implant · Signus Medical, LLC · 2003K003043 · Surgical Titanium Mesh System · Depuyacromed · 2001K990148 · Stackable Cage System · Depuy Motech Acromed · 1999K841189 · Oec/Rezaian Spinal Fixator · Orthopedic Equipment Co., Inc. · 1984−15−10−50
  1. Root · 16 generations backK841189 Oec/Rezaian Spinal FixatorOrthopedic Equipment Co., Inc. · 1984
  2. 12 generations between
  3. 3 generations backK131122 S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, LLC · 2013
  4. 2 generations backK140106 S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, Inc. · 2014
  5. 1 generation backK180502 S128 Anterior Lumbar Interbody Fusion (ALIF) SystemRenovis Surgical Technologies, Inc. · 2018
  6. This deviceK193320 KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) SystemKyocera Medical Technologies, Inc. · 2020
  7. Descendant · cites it directly · 2 of its ownK212980 Tesera-k ALIF SystemKyocera Medical Technologies, Inc. · 2022
  8. Descendant · 2 generations on · 1 of its ownK242771 Tesera-k PL System and Tesera-k XL SystemKyocera Medical Technologies, Inc. · 2025
  9. Descendant · 2 generations on · 0 of its ownK261353 Tesera-K Anterior Lumbar Interbody Fusion (ALIF) System; Tesera-K XL…Kyocera Medical Technologies, Inc. · 2026
Show all 16 generations
  1. Root · 16 generations backK841189 Oec/Rezaian Spinal FixatorOrthopedic Equipment Co., Inc. · 1984
  2. 15 generations backK990148 Stackable Cage SystemDepuy Motech Acromed · 1999
  3. 14 generations backK003043 Surgical Titanium Mesh SystemDepuyacromed · 2001
  4. 13 generations backK022793 Tetris Spinal ImplantSignus Medical, LLC · 2003
  5. 12 generations backK031757 Peek Tetris Spinal ImplantSignus Medical, LLC · 2003
  6. 11 generations backK041888 Curved Peek Tetris Spinal ImplantSignus Medizintechnik GmbH · 2004
  7. 10 generations backK042930 Spacevision Cage SystemSpine Vision, Inc. · 2004
  8. 9 generations backK062759 Paramount VBR SystemInnovative Spinal Technologies, Inc. · 2007
  9. 8 generations backK072120 Paramount Intervertebral Body Fusion DeviceInnovative Spinal Technologies, Inc. · 2007
  10. 7 generations backK080083 Intrepid Spinal SystemMedtronic Sofamor Danek · 2008 · Recalled
  11. 6 generations backK073109 Stalif TT Intervertebral Body Fusion SystemSurgicraft, Ltd. · 2008
  12. 5 generations backK083475 Lucent MagnumSpinal Elements, Inc. · 2009
  13. 4 generations backK092211 Chesapeake Spinal SystemK2m, Inc. · 2010
  14. 3 generations backK131122 S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, LLC · 2013
  15. 2 generations backK140106 S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, Inc. · 2014
  16. 1 generation backK180502 S128 Anterior Lumbar Interbody Fusion (ALIF) SystemRenovis Surgical Technologies, Inc. · 2018
  17. This deviceK193320 KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) SystemKyocera Medical Technologies, Inc. · 2020

Most influential descendants

Showing 3 of 3 · See all
  • K212980Tesera-k ALIF SystemKyocera Medical Technologies, Inc. · 2022 · cites it directly2descendants
  • K242771Tesera-k PL System and Tesera-k XL SystemKyocera Medical Technologies, Inc. · 2025 · 2 generations on1descendant
  • K261353Tesera-K Anterior Lumbar Interbody Fusion (ALIF) System; Tesera-K XL System; Lumbar…Kyocera Medical Technologies, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−16K841189Oec/Rezaian Spinal FixatorOrthopedic Equipment Co., Inc.1984—
−15K990148Stackable Cage SystemDepuy Motech Acromed1999—
−14K003043Surgical Titanium Mesh SystemDepuyacromed2001—
−13K022793Tetris Spinal ImplantSignus Medical, LLC2003—
−12K031757Peek Tetris Spinal ImplantSignus Medical, LLC2003—
−11K041888Curved Peek Tetris Spinal ImplantSignus Medizintechnik GmbH2004—
−10K042930Spacevision Cage SystemSpine Vision, Inc.2004—
−9K062759Paramount VBR SystemInnovative Spinal Technologies, Inc.2007—
−8K072120Paramount Intervertebral Body Fusion DeviceInnovative Spinal Technologies, Inc.2007—
−7K080083Intrepid Spinal SystemMedtronic Sofamor Danek20088
−6K073109Stalif TT Intervertebral Body Fusion SystemSurgicraft, Ltd.2008—
−5K083475Lucent MagnumSpinal Elements, Inc.2009—
−4K092211Chesapeake Spinal SystemK2m, Inc.2010—
−3K131122S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, LLC2013—
−2K140106S128 Anterior Lumbar Interbody Fusion (Alif) SystemRenovis Surgical Technologies, Inc.2014—
−1K180502S128 Anterior Lumbar Interbody Fusion (ALIF) SystemRenovis Surgical Technologies, Inc.2018—
0 (this device)K193320KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) SystemKyocera Medical Technologies, Inc.2020—
+1K212980Tesera-k ALIF SystemKyocera Medical Technologies, Inc.2022—
+2K242771Tesera-k PL System and Tesera-k XL SystemKyocera Medical Technologies, Inc.2025—
+2K261353Tesera-K Anterior Lumbar Interbody Fusion (ALIF) System; Tesera-K XL System…Kyocera Medical Technologies, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199620
199920
200010
200150
200250
200350
200470
200550
200610
200770
2008100
200920
201110
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202510
202610
FIG. 1 — Lineage of K193320, 1984–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OVD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OVD device

  1. #1 K222009 aprevo® anterior lumbar interbody fusion…score 26.0 · 23 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131122S128 Anterior Lumbar Interbody Fusion (Alif) System
Summary page 4
…: Class II Regulations 21 CFR 888.3080 and Product Code: OVD Classification panel: Orthopedic C) Predicates: Primary: K131122 Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System Secondary: K140…
Renovis Surgical Technologies, LLC2013—
*K140106S128 Anterior Lumbar Interbody Fusion (Alif) System
Summary page 4
…C) Predicates: Primary: K131122 Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System Secondary: K140106 Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System Secondary: K180…
Renovis Surgical Technologies, Inc.2014—
*K180502S128 Anterior Lumbar Interbody Fusion (ALIF) System
Summary page 4
…Secondary: K140106 Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System Secondary: K180502 Renovis S128 Anterior Lumbar Interbody Fusion (ALIF) System Kyocera Tesera SA ALIF System…
Renovis Surgical Technologies, Inc.2018—
†K181655Renovis S180 Lateral Lumbar Interbody Fusion SystemRenovis Surgical Technologies2018—

Also cleared against K131122: 5 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K212980Tesera-k ALIF SystemKyocera Medical Technologies, Inc.20222

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 36 years and passes through 14 companies; a new generation was cleared every 1.1 years on median.
  • 11 of the 70 ancestors have a recall of any class.
  • Of its descendants, K212980 (Kyocera Medical Technologies, Inc., 2022) has gone furthest, with 2 descendants of its own.
  • 5 other devices also cite its predicate K131122.

How does its category compare?

All OVD devices

OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · OVD

10in 2025

2007: 22008: 82009: 32010: 82011: 92012: 102013: 112014: 142015: 122016: 132017: 122018: 142019: 122020: 142021: 72022: 142023: 122024: 122025: 102026 (partial year): 8
2007high 14 (2014) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OVD

100in 2025

2008: 84 days median2009: 152 days median2010: 117 days median2011: 167 days median2012: 78 days median2013: 137 days median2014: 139 days median2015: 166 days median2016: 127 days median2017: 127 days median2018: 126 days median2019: 91 days median2020: 91 days median2021: 108 days median2022: 133 days median2023: 92 days median2024: 110 days median2025: 100 days median2026 (partial year): 93 days median
2007high 167 (2011) · low 78 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OVD

1in 2025

2013: 42016: 12021: 12022: 12025: 12026 (partial year): 1
2007high 4 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OVD

50in 2025

2009: 22010: 22011: 62012: 162013: 172014: 352015: 432016: 552017: 722018: 1122019: 1712020: 2032021: 2212022: 742023: 672024: 922025: 50
2007high 221 (2021) · low 0 (2007)2025

MAUDE reports naming product code OVD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
200888400
2009315202
2010811702
2011916706
20121078016
201311137417
201414139035
201512166043
201613127155
201712127072
2018141260112
201912910171
202014910203
202171081221
202214133174
20231292067
202412110092
202510100150
20268931—

What is it indicated for?

The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or…

Read the full statement (194 words)

The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI Tesera SA ALIF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI Tesera SA ALIF System is a standalone device and is intended to be used with the cover plate and screws provided and requires no additional supplementary fixation. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. Should the physician choose to use fewer than the four screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants ≥20°.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K193320: KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K193320

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)