510(k) Report

510(k) · Cleared

Cleared

Panther Stereotactic

K193459510(k) TraditionalCleared Apr 27, 2020133-day reviewClass II21 CFR 892.5050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Prowess, Inc. · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Apr 27, 2020 as substantially equivalent to K173791 (Elekta Instrument AB, 2018) and 2 other predicates after a 133-day review. Its family tree reaches back 9 generations to 1992. 1 later device descends from it.

Decided
Apr 27, 2020
Review
133 days
Class
Class II
Product code
MUJ
Recalls
None
Predicates
3

Its longest line of descent runs back 9 generations to K914698 (1992). 61 ancestors in all, the oldest K862643 (1987). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2018: 1 other ancestor2010: 3 other ancestors2015: 1 other ancestor2008: 2 other ancestors2007: 3 other ancestors2009: 2 other ancestors2003: 4 other ancestors2002: 2 other ancestors2001: 4 other ancestors2000: 5 other ancestors2006: 1 other ancestor2005: 2 other ancestors2004: 2 other ancestors1998: 3 other ancestors1997: 2 other ancestors1995: 1 other ancestor1994: 3 other ancestors1992: 3 other ancestors1999: 1 other ancestor1996: 2 other ancestors1990: 1 other ancestor1987: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestor2023: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K032724 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2003 · recalledK041577 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2004 · recalledK071938 · Monaco RTP System · Computerized Medical Systems, Inc. · 2007 · recalledK091179 · Monaco RTP System, Vmat Option · Computerized Medical Systems, Inc. · 2009 · recalledK101076 · Prowess Panther Proarc, Model Version 5.0 · Prowess, Inc. · 20101990200020102020
  1. Root · 9 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 5 generations between
  3. 3 generations backK071938 Monaco RTP SystemComputerized Medical Systems, Inc. · 2007 · Recalled
  4. 2 generations backK091179 Monaco RTP System, Vmat OptionComputerized Medical Systems, Inc. · 2009 · Recalled
  5. 1 generation backK101076 Prowess Panther Proarc, Model Version 5.0Prowess, Inc. · 2010
  6. This deviceK193459 Panther StereotacticProwess, Inc. · 2020
  7. Descendant · cites it directly · 0 of its ownK231825 Panther TPSProwess, Inc. · 2023
Show all 9 generations
  1. Root · 9 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 8 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 7 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 6 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 5 generations backK032724 PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories · 2003 · Recalled
  6. 4 generations backK041577 PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories · 2004 · Recalled
  7. 3 generations backK071938 Monaco RTP SystemComputerized Medical Systems, Inc. · 2007 · Recalled
  8. 2 generations backK091179 Monaco RTP System, Vmat OptionComputerized Medical Systems, Inc. · 2009 · Recalled
  9. 1 generation backK101076 Prowess Panther Proarc, Model Version 5.0Prowess, Inc. · 2010
  10. This deviceK193459 Panther StereotacticProwess, Inc. · 2020

Most influential descendants

Showing 1 of 1 · See all
  • K231825Panther TPSProwess, Inc. · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−8K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−7K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−6K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−5K032724PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories20031
−4K041577PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories20047
−3K071938Monaco RTP SystemComputerized Medical Systems, Inc.20075
−2K091179Monaco RTP System, Vmat OptionComputerized Medical Systems, Inc.20095
−1K101076Prowess Panther Proarc, Model Version 5.0Prowess, Inc.2010—
0 (this device)K193459Panther StereotacticProwess, Inc.2020—
+1K231825Panther TPSProwess, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
198910
199010
199110
199230
199310
199430
199510
199620
199720
199830
199910
200050
200140
200220
200340
200420
200520
200610
200730
200820
200920
201030
201510
201810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K193459, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K173791Leksell GammaPlan
Summary page 4
…Radiation Therapy Treatment Planning System Predicate Devices: Primary Predicate: Elekta’s Leksell GammaPlan, K173791 Additional Predicates: American Radiosurgery, Inc.’s…
Elekta Instrument AB20182
*K101220Explorer 4D Treatment Planning System
Summary page 4
…Additional Predicates: American Radiosurgery, Inc.’s Explorer 4D Treatment Planning System, K101220 Prowess Inc.’s Panther ProArc, K101076 1844 Clayton Road, Concord, CA 94520 Tel: (…
American Radiosurgery, Inc.2010—
*K101076Prowess Panther Proarc, Model Version 5.0
Summary page 4
…surgery, Inc.’s Explorer 4D Treatment Planning System, K101220 Prowess Inc.’s Panther ProArc, K101076 1844 Clayton Road, Concord, CA 94520 Tel: (925) 356-0360 Fax: (925) 356-0363 www.prowess.com 3. Dev…
Prowess, Inc.2010—

Also cleared against K173791: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231825Panther TPSProwess, Inc.20230

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 6 companies; a new generation was cleared every 1.8 years on median.
  • 18 of the 61 ancestors have a recall of any class.
  • Of its descendants, K231825 (Prowess, Inc., 2023) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K173791.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

Panther Stereotactic is intended to support highly advanced precision-targeted radiation planning.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K193459: Panther Stereotactic.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K193459

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026