510(k) Report

510(k) · Cleared

Recalled · 2011

Monaco RTP System, Vmat Option

K091179510(k) TraditionalCleared Jul 16, 200984-day reviewClass II21 CFR 892.5050Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Computerized Medical Systems, Inc. · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Jul 16, 2009 as substantially equivalent to K071938 (Computerized Medical Systems, Inc., 2007) and 2 other predicates after an 84-day review. Its family tree reaches back 7 generations to 1992. 12 later devices descend from it. It has been recalled 5 times.

Decided
Jul 16, 2009
Review
84 days
Class
Class II
Product code
MUJ
Recalls
5
Predicates
3

Its longest line of descent runs back 7 generations to K914698 (1992). 43 ancestors in all, the oldest K862643 (1987). 12 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2007: 2 other ancestors2003: 4 other ancestors2002: 2 other ancestors2001: 4 other ancestors2000: 5 other ancestors2005: 1 other ancestor1997: 2 other ancestors1995: 1 other ancestor1994: 3 other ancestors1992: 2 other ancestors2004: 1 other ancestor1998: 2 other ancestors1996: 2 other ancestors1990: 1 other ancestor1987: 1 other ancestor1993: 1 other ancestor1989: 1 other ancestor2010: 1 descendant cleared2011: 1 descendant cleared2020: 1 descendant cleared2013: 1 descendant cleared2015: 1 descendant cleared2023: 2 descendants cleared2018: 1 descendant cleared2019: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K032724 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2003 · recalledK041577 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2004 · recalledK071938 · Monaco RTP System · Computerized Medical Systems, Inc. · 2007 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 8 other ancestors3 generations back: 7 other ancestors4 generations back: 13 other ancestors5 generations back: 5 other ancestors6 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 1 descendant9 generations forward: 1 descendant K071938 · Monaco RTP System · Computerized Medical Systems, Inc. · 2007 · recalledK041577 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2004 · recalledK032724 · PINNACLE3 Radiation Therapy Planning System · Adac Laboratories · 2003 · recalledK993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−50+5
  1. Root · 7 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 6 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 5 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 4 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 3 generations backK032724 PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories · 2003 · Recalled
  6. 2 generations backK041577 PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories · 2004 · Recalled
  7. 1 generation backK071938 Monaco RTP SystemComputerized Medical Systems, Inc. · 2007 · Recalled
  8. This deviceK091179 Monaco RTP System, Vmat OptionComputerized Medical Systems, Inc. · 2009
  9. Descendant · cites it directly · 8 of its ownK110730 Monaco RTP SystemComputerized Medical Systems, Inc. · 2011 · Recalled
  10. Descendant · 2 generations on · 7 of its ownK132971 MonacoImpac Medical Systems, Inc. · 2013 · Recalled
  11. Descendant · 3 generations on · 6 of its ownK151233 MonacoImpac Medical Systems, Inc. · 2015 · Recalled
  12. Descendant · 4 generations on · 5 of its ownK183037 Monaco RTP SystemElekta, Inc. · 2018 · Recalled
  13. Descendant · 5 generations on · 4 of its ownK190178 Monaco RTP SystemElekta, Inc. · 2019 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12013: 12015: 12018: 12019: 12020: 12021: 12022: 12023: 22026: 1
2010high 2 (2023) · low 0 (2012)2026

Most influential descendants

Showing 5 of 12 · See all
  • K110730Monaco RTP SystemComputerized Medical Systems, Inc. · 2011 · cites it directly8descendants
  • K132971MonacoImpac Medical Systems, Inc. · 2013 · 2 generations on7descendants
  • K151233MonacoImpac Medical Systems, Inc. · 2015 · 3 generations on6descendants
  • K183037Monaco RTP SystemElekta, Inc. · 2018 · 4 generations on5descendants
  • K190178Monaco RTP SystemElekta, Inc. · 2019 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−6K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−5K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−4K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−3K032724PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories20031
−2K041577PINNACLE3 Radiation Therapy Planning SystemAdac Laboratories20047
−1K071938Monaco RTP SystemComputerized Medical Systems, Inc.20075
0 (this device)K091179Monaco RTP System, Vmat OptionComputerized Medical Systems, Inc.2009—
+1K110730Monaco RTP SystemComputerized Medical Systems, Inc.20116
+2K132971MonacoImpac Medical Systems, Inc.20134
+3K151233MonacoImpac Medical Systems, Inc.20159
+4K183037Monaco RTP SystemElekta, Inc.20186
+5K190178Monaco RTP SystemElekta, Inc.20195

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
198910
199010
199220
199310
199430
199510
199620
199720
199820
200050
200140
200220
200340
200410
200510
200720
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201310
201510
201810
201910
202010
202110
202210
202320
202610
FIG. 1 — Lineage of K091179, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071938Monaco RTP System
Summary page 2
…is substantially equivalent to these product(s) in intended use and safety and effectiveness. 1. Monaco RTP System Computerized Medical Systems, Inc. K071938 2. ERGO++ 3D Line Medical Systems K080601 3. Eclipse Treatment Planning System Varian Medical Syst…
Computerized Medical Systems, Inc.20075
*K080601Ergo++ Version 1.7
Summary page 2
…use and safety and effectiveness. 1. Monaco RTP System Computerized Medical Systems, Inc. K071938 2. ERGO++ 3D Line Medical Systems K080601 3. Eclipse Treatment Planning System Varian Medical Systems K073020 MONACO INTENDED USE The Monaco system is…
3D Line Medical Systems S.R.L.20082
*K073020Eclipse Treatment Planning System
Summary page 2
…Systems, Inc. K071938 2. ERGO++ 3D Line Medical Systems K080601 3. Eclipse Treatment Planning System Varian Medical Systems K073020 MONACO INTENDED USE The Monaco system is used to create treatment plans for any cancer patient for whom external beam intensity…
Varian Medical Systems, Inc.20076

What descends from it? (12)

See all 12

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110730Monaco RTP SystemComputerized Medical Systems, Inc.20118
K101076Prowess Panther Proarc, Model Version 5.0Prowess, Inc.20102

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 13, 2011Not classifiedZ-2598-2011Software designMonaco: When the isocenter location is edited in the IMRT Activity screen, the shift coordinates are not updated in the IMRT Plan report.
Jan 13, 2011Not classifiedZ-2561-2011Software designMonaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.
Oct 27, 2010Not classifiedZ-2569-2011Software designMonaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.
May 31, 2010Not classifiedZ-2564-2011Software designMonaco: due to a software defect, the Radiological Depth (cm) and SSD to DRP (cm) parameters are reported incorrectly on the Beam Summary Report. This occurs for all Monaco IMRT plans.
Jun 30, 2008Not classifiedZ-2560-2011Software designMonaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may only use the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 1.8 years on median.
  • 11 of the 43 ancestors have a recall of any class.
  • Of its descendants, K110730 (Computerized Medical Systems, Inc., 2011) has gone furthest, with 8 descendants of its own.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

The Monaco system is used to create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.

The Monaco product line is intended for use in radiation treatment planning using generally accepted methods for contouring, image manipulation, simulation, image fusion, plan optimization and QA and plan review.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091179: Monaco RTP System, Vmat Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K091179

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026