Maximus System

K200988510(k) SpecialCleared May 26, 202041-day reviewClass II21 CFR 868.5905Recalls 10 (9 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hill-Rom Services Pte, Ltd. · NHJ Device, Positive Pressure Breathing, Intermittent

Cleared on May 26, 2020 as substantially equivalent to K192143 (Hill-Rom Services Pte, Ltd., 2020) after a 41-day review. Its family tree reaches back 6 generations to 1995. 2 of the 16 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 10 times, including 9 Class I recalls.

Decided
May 26, 2020
Review
41 days
Class
Class II
Product code
NHJ
Recalls
10 (9 Class I)
Predicates
1

Its longest line of descent runs back 6 generations to K944900 (1995). 12 ancestors in all, the oldest K895485 (1989). 3 later devices descend from it, 2 citing it directly. 2 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2014: 1 other ancestor2012: 1 other ancestor2000: 2 other ancestors1989: 1 other ancestor1996: 1 other ancestor2024: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant clearedK944900 · Therapep · Dhd Diemolding Healthcare Div. · 1995K991300 · Boeing, Positive Airway Pressure (Pap) Therapy Device · Dhd Healthcare Corp. · 1999K051964 · Powerneb · Comedica, Inc. · 2005K124032 · Metaneb · Hill-Rom Services Pte, Ltd. · 2013K151689 · MetaNeb 4 System · Hill-Rom Services Pte, Ltd. · 2016K192143 · Maximus System · Hill-Rom Services Pte, Ltd. · 2020 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 3 other ancestors5 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K192143 · Maximus System · Hill-Rom Services Pte, Ltd. · 2020 · Class I recallK151689 · MetaNeb 4 System · Hill-Rom Services Pte, Ltd. · 2016K124032 · Metaneb · Hill-Rom Services Pte, Ltd. · 2013K051964 · Powerneb · Comedica, Inc. · 2005K991300 · Boeing, Positive Airway Pressure (Pap) Therapy Device · Dhd Healthcare Corp. · 1999K944900 · Therapep · Dhd Diemolding Healthcare Div. · 1995−50
  1. Root · 6 generations backK944900 TherapepDhd Diemolding Healthcare Div. · 1995
  2. 5 generations backK991300 Boeing, Positive Airway Pressure (Pap) Therapy DeviceDhd Healthcare Corp. · 1999
  3. 4 generations backK051964 PowernebComedica, Inc. · 2005
  4. 3 generations backK124032 MetanebHill-Rom Services Pte, Ltd. · 2013
  5. 2 generations backK151689 MetaNeb 4 SystemHill-Rom Services Pte, Ltd. · 2016
  6. 1 generation backK192143 Maximus SystemHill-Rom Services Pte, Ltd. · 2020 · Recalled (Class I)
  7. This deviceK200988 Maximus SystemHill-Rom Services Pte, Ltd. · 2020
  8. Descendant · cites it directly · 1 of its ownK231728 BiWaze Clear SystemAbmrc, LLC · 2024
  9. Descendant · 2 generations on · 0 of its ownK260604 BiWaze Clear SystemAbmrc, LLC · 2026
  10. Descendant · cites it directly · 0 of its ownK213564 BiWaze Clear SystemAbmrc, LLC · 2022

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K231728BiWaze Clear SystemAbmrc, LLC · 2024 · cites it directly1descendant
  • K260604BiWaze Clear SystemAbmrc, LLC · 2026 · 2 generations on0descendants
  • K213564BiWaze Clear SystemAbmrc, LLC · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K944900TherapepDhd Diemolding Healthcare Div.1995—
−5K991300Boeing, Positive Airway Pressure (Pap) Therapy DeviceDhd Healthcare Corp.1999—
−4K051964PowernebComedica, Inc.2005—
−3K124032MetanebHill-Rom Services Pte, Ltd.2013—
−2K151689MetaNeb 4 SystemHill-Rom Services Pte, Ltd.2016—
−1K192143Maximus SystemHill-Rom Services Pte, Ltd.2020Class I
0 (this device)K200988Maximus SystemHill-Rom Services Pte, Ltd.2020—
+1K231728BiWaze Clear SystemAbmrc, LLC2024—
+2K260604BiWaze Clear SystemAbmrc, LLC2026—
+1K213564BiWaze Clear SystemAbmrc, LLC2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199610
200020
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202410
202610
FIG. 1 — Lineage of K200988, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NHJ device

  1. #1 K260604 BiWaze Clear Systemscore 14.0 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K192143Maximus System
Summary page 4
…PB) Classification Code/Name: NHJ – non-continuous ventilator (IPPB) 21 CFR 868.5905, Class II Predicate Devices: K192143 – Hill-Rom Maximus™ System Device Description and Modification: The modification is to add the ability to provide aerosol from t…
Hill-Rom Services Pte, Ltd.2020Class I

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231728BiWaze Clear SystemAbmrc, LLC20241
K213564BiWaze Clear SystemAbmrc, LLC20220

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 3, 2026Class IIZ-3257-2026Device DesignContamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
May 21, 2026Class IZ-2549-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2551-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2548-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2550-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
Feb 18, 2026Class IZ-1637-2026Process controlUrgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue…
May 30, 2024Class IZ-2321-2024Device DesignPotential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
May 30, 2024Class IZ-2322-2024Device DesignPotential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
Apr 26, 2022Class IZ-1139-2022Device DesignThere is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.
Apr 26, 2022Class IZ-1138-2022Device DesignThere is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K192143Maximus SystemHill-Rom Services Pte, Ltd. · 2020 · Direct predicate · ◆ on the line of descent in FIG. 17 Class I recalls

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 4.5 years on median.
  • 2 of the 12 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K231728 (Abmrc, LLC, 2024) has gone furthest, with 1 descendant of its own.

How does its category compare?

All NHJ devices

NHJ Device, Positive Pressure Breathing, Intermittent: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · NHJ

1in 2025

2008: 12012: 22013: 22014: 32016: 12020: 32022: 12024: 12025: 12026 (partial year): 2
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NHJ

190in 2020

2014: 207 days median2020: 190 days median
2007high 207 (2014) · low 190 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NHJ

11so far in 2026, the most since 2007

2024: 1, 1 Class I2026 (partial year): 11, 4 Class I
2007high 11 (2026) · low 0 (2007)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · NHJ

35in 2025

2007: 162008: 72009: 12010: 22012: 22013: 22014: 12015: 22016: 22017: 62018: 22019: 12020: 32021: 52022: 62023: 172024: 402025: 35
2007high 40 (2024) · low 0 (2011)2025

MAUDE reports naming product code NHJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—016
20081—07
20090—01
20100—02
20110—00
20122—02
20132—02
2014320701
20150—02
20161—02
20170—06
20180—02
20190—01
2020319003
20210—05
20221—06
20230—017
20241—1 (1)40
20251—035
20262—11 (4)—

What is it indicated for?

The Maximus™ System provides features of both the Synclara™ System and the Volara™ System. The Maximus™ System, when used as a Synclara™ Cough System is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness. The Maximus™ System, when used as a Volara™ System is intended for the mobilization of secretions, lung expansion therapy, the treatment and…

Read the full statement (177 words)

The Maximus™ System provides features of both the Synclara™ System and the Volara™ System.

The Maximus™ System, when used as a Synclara™ Cough System is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness.

The Maximus™ System, when used as a Volara™ System is intended for the mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and has the ability to provide supplemental oxygen when used with oxygen supply.

Patient Population

The Maximus™ System, when used as a Synclara™ Cough System, is intended to deliver therapy to the population of pediatric to adult patients in both acute and home care settings.

The Maximus™ System, when used as a Volara™ System, is intended to deliver therapy to adults and children over the age of 2 in the acute care setting.

The Maximus™ System, when used as a Volara™ System, is intended to deliver therapy to adults and children over the age of 5 in the home care setting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200988: Maximus System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K200988

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026