510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Maximus System

K192143510(k) TraditionalCleared Feb 14, 2020190-day reviewClass II21 CFR 868.5905Recalls 7 (7 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hill-Rom Services Pte, Ltd. · NHJ Device, Positive Pressure Breathing, Intermittent

Cleared on Feb 14, 2020 as substantially equivalent to K124032 (Hill-Rom Services Pte, Ltd., 2013) and 3 other predicates after a 190-day review. Its family tree reaches back 5 generations to 1995. 2 of the 16 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 7 times, including 7 Class I recalls.

Decided
Feb 14, 2020
Review
190 days
Class
Class II
Product code
NHJ
Recalls
7 (7 Class I)
Predicates
4

Its longest line of descent runs back 5 generations to K944900 (1995). 11 ancestors in all, the oldest K895485 (1989). 4 later devices descend from it, 1 citing it directly. 2 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants3 generations forward: 1 descendant K151689 · MetaNeb 4 System · Hill-Rom Services Pte, Ltd. · 2016K124032 · Metaneb · Hill-Rom Services Pte, Ltd. · 2013K051964 · Powerneb · Comedica, Inc. · 2005K991300 · Boeing, Positive Airway Pressure (Pap) Therapy Device · Dhd Healthcare Corp. · 1999K944900 · Therapep · Dhd Diemolding Healthcare Div. · 1995−50
  1. Root · 5 generations backK944900 TherapepDhd Diemolding Healthcare Div. · 1995
  2. 4 generations backK991300 Boeing, Positive Airway Pressure (Pap) Therapy DeviceDhd Healthcare Corp. · 1999
  3. 3 generations backK051964 PowernebComedica, Inc. · 2005
  4. 2 generations backK124032 MetanebHill-Rom Services Pte, Ltd. · 2013
  5. 1 generation backK151689 MetaNeb 4 SystemHill-Rom Services Pte, Ltd. · 2016
  6. This deviceK192143 Maximus SystemHill-Rom Services Pte, Ltd. · 2020
  7. Descendant · cites it directly · 3 of its ownK200988 Maximus SystemHill-Rom Services Pte, Ltd. · 2020 · Recalled (Class I)
  8. Descendant · 2 generations on · 1 of its ownK231728 BiWaze Clear SystemAbmrc, LLC · 2024
  9. Descendant · 3 generations on · 0 of its ownK260604 BiWaze Clear SystemAbmrc, LLC · 2026
  10. Descendant · 2 generations on · 0 of its ownK213564 BiWaze Clear SystemAbmrc, LLC · 2022

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K200988Maximus SystemHill-Rom Services Pte, Ltd. · 2020 · cites it directly3descendants
  • K231728BiWaze Clear SystemAbmrc, LLC · 2024 · 2 generations on1descendant
  • K260604BiWaze Clear SystemAbmrc, LLC · 2026 · 3 generations on0descendants
  • K213564BiWaze Clear SystemAbmrc, LLC · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K944900TherapepDhd Diemolding Healthcare Div.1995—
−4K991300Boeing, Positive Airway Pressure (Pap) Therapy DeviceDhd Healthcare Corp.1999—
−3K051964PowernebComedica, Inc.2005—
−2K124032MetanebHill-Rom Services Pte, Ltd.2013—
−1K151689MetaNeb 4 SystemHill-Rom Services Pte, Ltd.2016—
0 (this device)K192143Maximus SystemHill-Rom Services Pte, Ltd.2020—
+1K200988Maximus SystemHill-Rom Services Pte, Ltd.2020Class I
+2K231728BiWaze Clear SystemAbmrc, LLC2024—
+3K260604BiWaze Clear SystemAbmrc, LLC2026—
+2K213564BiWaze Clear SystemAbmrc, LLC2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199610
200020
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202011
202210
202410
202610
FIG. 1 — Lineage of K192143, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NHJ device

  1. #1 K260604 BiWaze Clear Systemscore 14.0 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K124032Metaneb
Summary page 4
…(IPPB) Classification Code/Name: NHJ – non-continuous ventilator (IPPB) 21 CFR 868.5905, Class II Predicate Devices: K124032 – Hill-Rom MetaNeb® K151689 – Hill-Rom MetaNeb® K132988 – Vital C…
Hill-Rom Services Pte, Ltd.2013—
*K151689MetaNeb 4 System
Summary page 4
…ventilator (IPPB) 21 CFR 868.5905, Class II Predicate Devices: K124032 – Hill-Rom MetaNeb® K151689 – Hill-Rom MetaNeb® K132988 – Vital Cough Reference Devices: K121955 - Philips Resp…
Hill-Rom Services Pte, Ltd.2016—
*K132988Vital Cough
Summary page 4
…8.5905, Class II Predicate Devices: K124032 – Hill-Rom MetaNeb® K151689 – Hill-Rom MetaNeb® K132988 – Vital Cough Reference Devices: K121955 - Philips Respironics Cough Assist T70 K89…
Lung Assist, Inc.2014—
*K120277Vital Cough
Summary page 11
…Flutter Vital Cough 510(k) Number - K132988 K120277 Regulation Number: 868.5905 CFR classificatio…
Lung Assist, Inc.2012—
†K121955Coughassist T70 DeviceRespironics, Inc.2012—
†K895485Bird (Ipv) Noncontinuous VentilatorsPercussionaire Corp.1989—
†K142888Fingertip Pulse OximeterBeijing Choice Electronic Technololgy Co., Ltd.2015—

Also cleared against K124032: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200988Maximus System Class IHill-Rom Services Pte, Ltd.20203

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 21, 2026Class IZ-2549-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2551-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2548-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
May 21, 2026Class IZ-2550-2026Nonconforming Material/ComponentThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the…
Feb 18, 2026Class IZ-1637-2026Process controlUrgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue…
May 30, 2024Class IZ-2321-2024Device DesignPotential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
May 30, 2024Class IZ-2322-2024Device DesignPotential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K200988Maximus SystemHill-Rom Services Pte, Ltd. · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 19 Class I recalls

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 4.5 years on median.
  • 1 of the 11 ancestors has a recall of any class.
  • Of its descendants, K200988 (Hill-Rom Services Pte, Ltd., 2020) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K124032.

How does its category compare?

All NHJ devices

NHJ Device, Positive Pressure Breathing, Intermittent: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · NHJ

1in 2025

2008: 12012: 22013: 22014: 32016: 12020: 32022: 12024: 12025: 12026 (partial year): 2
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NHJ

190in 2020

2014: 207 days median2020: 190 days median
2007high 207 (2014) · low 190 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NHJ

11so far in 2026, the most since 2007

2024: 1, 1 Class I2026 (partial year): 11, 4 Class I
2007high 11 (2026) · low 0 (2007)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · NHJ

35in 2025

2007: 162008: 72009: 12010: 22012: 22013: 22014: 12015: 22016: 22017: 62018: 22019: 12020: 32021: 52022: 62023: 172024: 402025: 35
2007high 40 (2024) · low 0 (2011)2025

MAUDE reports naming product code NHJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—016
20081—07
20090—01
20100—02
20110—00
20122—02
20132—02
2014320701
20150—02
20161—02
20170—06
20180—02
20190—01
2020319003
20210—05
20221—06
20230—017
20241—1 (1)40
20251—035
20262—11 (4)—

What is it indicated for?

The Maximus™ System provides features of both the Synclara™ System and the Volara™ System. The Maximus™ System, when used as a Synclara™ Cough System is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness. The Maximus™ System, when used as a Volara™ System is intended for the mobilization of secretions, lung expansion therapy, the treatment and…

Read the full statement (177 words)

The Maximus™ System provides features of both the Synclara™ System and the Volara™ System.

The Maximus™ System, when used as a Synclara™ Cough System is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness.

The Maximus™ System, when used as a Volara™ System is intended for the mobilization of secretions, lung expansion therapy, the treatment and prevention of pulmonary atelectasis, and has the ability to provide supplemental oxygen when used with oxygen supply.

Patient Population

The Maximus™ System, when used as a Synclara™ Cough System, is intended to deliver therapy to the population of pediatric to adult patients in both acute and home care settings.

The Maximus™ System, when used as a Volara™ System, is intended to deliver therapy to adults and children over the age of 2 in the acute care setting.

The Maximus™ System, when used as a Volara™ System, is intended to deliver therapy to adults and children over the age of 5 in the home care setting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K192143: Maximus System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192143

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026