510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Abiomed 23 Fr Sheath

K201116510(k) TraditionalCleared Jun 15, 202049-day reviewClass II21 CFR 870.1340Recalls 13 (13 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abiomed, Inc. · DYB Introducer, Catheter

Cleared on Jun 15, 2020 as substantially equivalent to K122084 (Oscor, Inc., 2012) after a 49-day review. Its family tree reaches back 2 generations to 1994. 2 of the 3 devices in its family, counting itself, have a Class I recall. It has been recalled 13 times, including 13 Class I recalls.

Decided
Jun 15, 2020
Review
49 days
Class
Class II
Product code
DYB
Recalls
13 (13 Class I)
Predicates
1

Its longest line of descent runs back 2 generations to K934901 (1994). No later device cites it as a predicate yet. 2 of the 3 devices in this family, counting this one, have a Class I recall.

  1. Root · 2 generations backK934901 Tear-Away Sheath Introducer Set with Integral Hemostasis ValveThomas Medical Products, Inc. · 1994 · Recalled
  2. 1 generation backK122084 Introducer Set, Model Adelante-S SeriesOscor, Inc. · 2012 · Recalled (Class I)
  3. This deviceK201116 Abiomed 23 Fr SheathAbiomed, Inc. · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K934901Tear-Away Sheath Introducer Set with Integral Hemostasis ValveThomas Medical Products, Inc.19947
−1K122084Introducer Set, Model Adelante-S SeriesOscor, Inc.2012Class I
0 (this device)K201116Abiomed 23 Fr SheathAbiomed, Inc.2020—
FIG. 1 — Lineage of K201116, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122084Introducer Set, Model Adelante-S Series
Summary page 4
…e Standard: (i.e. none established under section 514 of the F D & C Act.) C. Predicate Device: Oscor Adelante-S Introducer Series, K122084 D. Device Description: The Abiomed 23 Fr Sheath is packaged as a sterile kit, which includes an introducer sheath (consisting o…
Oscor, Inc.2012Class I

Also cleared against K122084: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 11, 2026Class IZ-2685-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2689-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2688-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2681-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2690-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2682-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2691-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2687-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2684-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2693-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2683-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2686-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…
Jun 11, 2026Class IZ-2692-2026Under Investigation by firmA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122084Introducer Set, Model Adelante-S SeriesOscor, Inc. · 2012 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 26 years and passes through 3 companies; a new generation was cleared every 18.6 years on median.
  • 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
  • 1 other device also cites its predicate K122084.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

The Abiomed 23 Fr Sheath is intended for introduction of pacing leads or catheters.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K201116: Abiomed 23 Fr Sheath.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K201116

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026