510(k) · Cleared
Recalled · 2010Tear-Away Sheath Introducer Set with Integral Hemostasis Valve
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Thomas Medical Products, Inc. · DYB Introducer, Catheter
Cleared on Jan 21, 1994 after a 99-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 4 of the 7 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 7 times.
- Decided
- Jan 21, 1994
- Review
- 99 days
- Class
- Class II
- Product code
- DYB
- Recalls
- 7
- Predicates
- 0
6 later devices descend from it, 3 citing it directly. 4 of the 7 devices in this family, counting this one, have a Class I recall.
- This deviceK934901 Tear-Away Sheath Introducer Set with Integral Hemostasis ValveThomas Medical Products, Inc. · 1994
- Descendant · cites it directly · 2 of its ownK122084 Introducer Set, Model Adelante-S SeriesOscor, Inc. · 2012 · Recalled (Class I)
- Descendant · cites it directly · 1 of its ownK143255 Prelude SNAP Splittable Sheath IntroducerMerit Medical Systems, Inc. · 2015 · Recalled
- Descendant · 2 generations on · 0 of its ownK201116 Abiomed 23 Fr SheathAbiomed, Inc. · 2020 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK192769 Impella CP IntroducerAbiomed, Inc. · 2019 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK152381 Prelude SNAP Splittable Hydrophilic Sheath IntroducerMerit Medical Systems, Inc. · 2015 · Recalled
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 6 · See all- K122084Introducer Set, Model Adelante-S SeriesOscor, Inc. · 2012 · cites it directly2descendants
- K143255Prelude SNAP Splittable Sheath IntroducerMerit Medical Systems, Inc. · 2015 · cites it directly1descendant
- K201116Abiomed 23 Fr SheathAbiomed, Inc. · 2020 · 2 generations on0descendants
- K192769Impella CP IntroducerAbiomed, Inc. · 2019 · 2 generations on0descendants
- K152381Prelude SNAP Splittable Hydrophilic Sheath IntroducerMerit Medical Systems, Inc. · 2015 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K934901 | Tear-Away Sheath Introducer Set with Integral Hemostasis Valve | Thomas Medical Products, Inc. | 1994 | — |
| +1 | K122084 | Introducer Set, Model Adelante-S Series | Oscor, Inc. | 2012 | Class I |
| +1 | K143255 | Prelude SNAP Splittable Sheath Introducer | Merit Medical Systems, Inc. | 2015 | 4 |
| +2 | K201116 | Abiomed 23 Fr Sheath | Abiomed, Inc. | 2020 | Class I |
| +2 | K192769 | Impella CP Introducer | Abiomed, Inc. | 2019 | Class I |
| +2 | K152381 | Prelude SNAP Splittable Hydrophilic Sheath Introducer | Merit Medical Systems, Inc. | 2015 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 1 |
| 2012 | 1 | 1 |
| 2015 | 2 | 0 |
| 2019 | 1 | 1 |
| 2020 | 1 | 1 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DYB device
- #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
What descends from it? (6)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 16, 2010 | Not classified | Z-1226-2011 | Packaging process control | Compromised sterility/weak or open seals. |
| Dec 16, 2010 | Not classified | Z-1225-2011 | Packaging process control | Compromised sterility/weak or open seals. |
| Dec 16, 2010 | Not classified | Z-1228-2011 | Packaging process control | Compromised sterility/weak or open seals. |
| Oct 1, 2008 | Not classified | Z-0280-2009 | Process control | Failure to insert the guidewire through the introducer needle. |
| Sep 12, 2008 | Not classified | Z-0131-2009 | Process control | Difficulty inserting guidewire through introducer needle. Needle hub is not tapered down to the needle cannula. |
| Nov 7, 2005 | Not classified | Z-0297-06 | Other | mislabeled as 9.5F actually contains 10.5F |
| Oct 14, 2004 | Not classified | Z-0056-05 | Other | Sheaths may fracture if exposed to excess fluorescent light |
The 5 latest of 7 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K122084Introducer Set, Model Adelante-S SeriesOscor, Inc. · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls
- K043438Large Bore Splittable Introducer KitThomas Medical Products, Inc. · 2005 · Cites it directly · ◆ in the 2005 descendants bar of FIG. 11 Class I recall
- K192769Impella CP IntroducerAbiomed, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls
- K201116Abiomed 23 Fr SheathAbiomed, Inc. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 113 Class I recalls
What does this lineage say?
- Of its descendants, K122084 (Oscor, Inc., 2012) has gone furthest, with 2 descendants of its own.
How does its category compare?
All DYB devicesDYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DYB
16in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DYB
63in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DYB
3in 2025
Class I (48 since 2007) Class II/III not classified
Adverse event reports per year · DYB
4,539in 2025
MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 95 | 15 | 316 |
| 2008 | 17 | 37 | 9 | 423 |
| 2009 | 14 | 52 | 25 | 399 |
| 2010 | 18 | 123 | 18 | 445 |
| 2011 | 17 | 100 | 4 | 628 |
| 2012 | 22 | 128 | 3 | 898 |
| 2013 | 26 | 118 | 5 (1) | 743 |
| 2014 | 30 | 122 | 2 | 780 |
| 2015 | 17 | 129 | 2 (1) | 913 |
| 2016 | 19 | 95 | 45 (3) | 1,234 |
| 2017 | 24 | 70 | 22 (1) | 1,672 |
| 2018 | 19 | 156 | 1 | 1,903 |
| 2019 | 18 | 47 | 2 | 2,450 |
| 2020 | 9 | 49 | 21 (21) | 2,465 |
| 2021 | 12 | 78 | 6 | 2,835 |
| 2022 | 13 | 143 | 2 | 4,084 |
| 2023 | 15 | 55 | 10 | 3,775 |
| 2024 | 18 | 56 | 2 | 3,931 |
| 2025 | 16 | 63 | 3 | 4,539 |
| 2026 | 17 | 59 | 23 (21) | — |
https://510kreport.com/device/K934901 · printed Nov 6, 2026
Cite this page
510(k) Report. “K934901: Tear-Away Sheath Introducer Set with Integral Hemostasis Valve.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K934901
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026