Flat Panel Detector

K201591510(k) TraditionalCleared Sep 3, 202083-day reviewClass II21 CFR 892.1680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ka Imaging, Inc. · MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Cleared on Sep 3, 2020 as substantially equivalent to K191939 (Innocare Optoelectronics Corp., 2019) after an 83-day review. Its family tree reaches back 12 generations to 1998.

Decided
Sep 3, 2020
Review
83 days
Class
Class II
Product code
MQB
Recalls
None
Predicates
1

Its longest line of descent runs back 12 generations to K981556 (1998). 19 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2010: 1 other ancestor2009: 1 other ancestor2004: 1 other ancestor2002: 2 other ancestors2008: 1 other ancestor2003: 1 other ancestorReference device: K150766 · Carestream DRX-1 System · Carestream Health, Inc. · 2015Reference device: K122454 · Fujifilm Dual Energy Subtraction (Des) Software Option · Fujifilm Medical System U.S.A., Inc. · 2012Reference device: K013481 · Dual Energy and Tissue Equalization Software Option · GE Medical Systems · 2001K981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K060433 · Digital Radiography Cxdi-50C · Canon, Inc. · 2006K091545 · Digital Radiography Cxdi-60C · Canon, Inc. · 2009K102012 · Digital Radiography · Canon Inc. -Medical Equipment Group · 2010 · recalledK103596 · Wireless/Wired FDR D-Evo Flat Panel Detector · Fujifilm Medical Systems U.S.A, Inc. · 2011K111548 · Wireless/Wired FDR D-Evo Flat Pancel Detector System Model DR-ID600 W/DR-ID611SE · Fujifilm Medical System U.S.A., Inc. · 2011K132509 · FDR D-Evo Flat Panel Detector System · Fujifilm Medical Systems U.S.A, Inc. · 2013K142003 · FDR D-Evo II Flat Panel Detector System · Fujifilm Medical System U.S.A., Inc. · 2014K191939 · Yushan X-Ray Flat Panel Detector · Innocare Optoelectronics Corp. · 2019199520002005201020152020
  1. Root · 12 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 8 generations between
  3. 3 generations backK132509 FDR D-Evo Flat Panel Detector SystemFujifilm Medical Systems U.S.A, Inc. · 2013
  4. 2 generations backK142003 FDR D-Evo II Flat Panel Detector SystemFujifilm Medical System U.S.A., Inc. · 2014
  5. 1 generation backK191939 Yushan X-Ray Flat Panel DetectorInnocare Optoelectronics Corp. · 2019
  6. This deviceK201591 Flat Panel DetectorKa Imaging, Inc. · 2020
Show all 12 generations
  1. Root · 12 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 11 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 10 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 9 generations backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. 8 generations backK060433 Digital Radiography Cxdi-50CCanon, Inc. · 2006
  6. 7 generations backK091545 Digital Radiography Cxdi-60CCanon, Inc. · 2009
  7. 6 generations backK102012 Digital RadiographyCanon Inc. -Medical Equipment Group · 2010 · Recalled
  8. 5 generations backK103596 Wireless/Wired FDR D-Evo Flat Panel DetectorFujifilm Medical Systems U.S.A, Inc. · 2011
  9. 4 generations backK111548 Wireless/Wired FDR D-Evo Flat Pancel Detector System Model DR-ID600…Fujifilm Medical System U.S.A., Inc. · 2011
  10. 3 generations backK132509 FDR D-Evo Flat Panel Detector SystemFujifilm Medical Systems U.S.A, Inc. · 2013
  11. 2 generations backK142003 FDR D-Evo II Flat Panel Detector SystemFujifilm Medical System U.S.A., Inc. · 2014
  12. 1 generation backK191939 Yushan X-Ray Flat Panel DetectorInnocare Optoelectronics Corp. · 2019
  13. This deviceK201591 Flat Panel DetectorKa Imaging, Inc. · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−11K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−10K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−9K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
−8K060433Digital Radiography Cxdi-50CCanon, Inc.2006—
−7K091545Digital Radiography Cxdi-60CCanon, Inc.2009—
−6K102012Digital RadiographyCanon Inc. -Medical Equipment Group20101
−5K103596Wireless/Wired FDR D-Evo Flat Panel DetectorFujifilm Medical Systems U.S.A, Inc.2011—
−4K111548Wireless/Wired FDR D-Evo Flat Pancel Detector System Model DR-ID600 W/DR-ID611SEFujifilm Medical System U.S.A., Inc.2011—
−3K132509FDR D-Evo Flat Panel Detector SystemFujifilm Medical Systems U.S.A, Inc.2013—
−2K142003FDR D-Evo II Flat Panel Detector SystemFujifilm Medical System U.S.A., Inc.2014—
−1K191939Yushan X-Ray Flat Panel DetectorInnocare Optoelectronics Corp.2019—
0 (this device)K201591Flat Panel DetectorKa Imaging, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220
200310
200410
200810
200910
201010
FIG. 1 — Lineage of K201591, 1998–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MQB device

  1. #1 K201528 Yushan X-Ray Flat Panel Detector with DROCscore 21.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K191939Yushan X-Ray Flat Panel Detector
Summary page 4
…Substantial equivalence to the following predicate devices is as follow: 510(k) # Product Company Reference K191939 Yushan X-Ray Flat Panel InnoCare Primary Detector Optoelec…
Innocare Optoelectronics Corp.2019—
†K150766Carestream DRX-1 SystemCarestream Health, Inc.2015—
†K122454Fujifilm Dual Energy Subtraction (Des) Software OptionFujifilm Medical System U.S.A., Inc.2012—
†K013481Dual Energy and Tissue Equalization Software OptionGE Medical Systems2001—

Also cleared against K191939: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 7 companies; a new generation was cleared every 1.2 years on median.
  • 3 of the 19 ancestors have a recall of any class.
  • 1 other device also cites its predicate K191939.

How does its category compare?

All MQB devices

MQB Solid State X-Ray Imager (Flat Panel/Digital Imager): FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MQB

8in 2025

2007: 132008: 82009: 152010: 112011: 242012: 222013: 212014: 322015: 212016: 222017: 312018: 162019: 222020: 192021: 212022: 312023: 102024: 82025: 82026 (partial year): 3
2007high 32 (2014) · low 8 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQB

127in 2025

2007: 22 days median2008: 15 days median2009: 57 days median2010: 116 days median2011: 102 days median2012: 80 days median2013: 105 days median2014: 102 days median2015: 81 days median2016: 92 days median2017: 30 days median2018: 30 days median2019: 30 days median2020: 28 days median2021: 48 days median2022: 48 days median2023: 30 days median2024: 187 days median2025: 127 days median2026 (partial year): 245 days median
2007high 245 (2026) · low 15 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MQB

1in 2025

2007: 152008: 362009: 192010: 142011: 102012: 82013: 62017: 12018: 42019: 12020: 42021: 12025: 12026 (partial year): 1
2007high 36 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MQB

141in 2025

2007: 252008: 302009: 342010: 392011: 1012012: 2672013: 1592014: 292015: 212016: 262017: 202018: 842019: 452020: 92021: 262022: 132023: 512024: 912025: 141
2007high 267 (2012) · low 9 (2020)2025

MAUDE reports naming product code MQB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200713221525
20088153630
200915571934
2010111161439
20112410210101
201222808267
2013211056159
201432102029
20152181021
20162292026
20173130120
20181630484
20192230145
2020192849
20212148126
20223148013
20231030051
20248187091
202581271141
202632451—

What is it indicated for?

The flat panel detector when used with a radiographic imaging system is intended to generate radiographic images of human anatomy wherever a conventional screen-film, digital radiography (DR) or computed radiography (CR) detector is used for general purposes.

When the dual energy subtraction function is enabled, it is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying or overlying anatomical structures.

This device is not intended for use in mammography applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K201591: Flat Panel Detector.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K201591

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026