Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box)

K201840510(k) TraditionalCleared Nov 4, 2020125-day reviewClass II21 CFR 882.1480Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Rebound Therapeutics · GWG Endoscope, Neurological

Cleared on Nov 4, 2020 as substantially equivalent to K191861 (Rebound Therapeutics, 2019) after a 125-day review. Its family tree reaches back 6 generations to 2000. 1 of the 14 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled once.

Decided
Nov 4, 2020
Review
125 days
Class
Class II
Product code
GWG
Recalls
1
Predicates
1

Its longest line of descent runs back 6 generations to K002931 (2000). 11 ancestors in all, the oldest K830332 (1983). 2 later devices descend from it, 2 citing it directly. 1 of the 14 devices in this family, counting this one, has a Class I recall.

  1. Root · 6 generations backK002931 Metrx SystemMedtronic Sofamor Danek, Inc. · 2000
  2. 5 generations backK043602 Quadrant Retractor SystemMedtronic Sofamor Danek · 2005
  3. 4 generations backK060973 Vycor Surgical Access SystemsVycor Medical, LLC · 2006 · Recalled (Class I)
  4. 3 generations backK120691 Brain PortNico Corporation · 2012 · Recalled
  5. 2 generations backK182211 Aurora Surgiscope SystemRebound Therapeutics Corporation · 2019
  6. 1 generation backK191861 Aurora Surgiscope SystemRebound Therapeutics · 2019
  7. This deviceK201840 Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System…Rebound Therapeutics · 2020
  8. Descendant · cites it directly · 0 of its ownK250752 AURORA® Surgiscope® System (ASX15/60); AURORA® Surgiscope® System…Integra LifeSciences Corporation · 2025
  9. Descendant · cites it directly · 0 of its ownK232618 Aurora Surgiscope SystemRebound Therapeutics Corporation · 2023 · Recalled

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K250752AURORA® Surgiscope® System (ASX15/60); AURORA® Surgiscope® System (ASX15/80)Integra LifeSciences Corporation · 2025 · cites it directly0descendants
  • K232618Aurora Surgiscope SystemRebound Therapeutics Corporation · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K002931Metrx SystemMedtronic Sofamor Danek, Inc.2000—
−5K043602Quadrant Retractor SystemMedtronic Sofamor Danek2005—
−4K060973Vycor Surgical Access SystemsVycor Medical, LLC2006Class I
−3K120691Brain PortNico Corporation20122
−2K182211Aurora Surgiscope SystemRebound Therapeutics Corporation2019—
−1K191861Aurora Surgiscope SystemRebound Therapeutics2019—
0 (this device)K201840Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control…Rebound Therapeutics2020—
+1K250752AURORA® Surgiscope® System (ASX15/60); AURORA® Surgiscope® System (ASX15/80)Integra LifeSciences Corporation2025—
+1K232618Aurora Surgiscope SystemRebound Therapeutics Corporation20231

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198920
199810
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202510
FIG. 1 — Lineage of K201840, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWG device

  1. #1 K211202 1688 4K Camera System, L11 LED Light Source…score 22.5 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K191861Aurora Surgiscope System
Summary page 4
…Self-retaining retractor for neurosurgery Regulatory Class: II Product Code: GWG, GZT PREDICATE DEVICE AURORA Surgiscope System, Rebound Therapeutics, K191861 NICO BrainPath, NICO Corp, K172433 (Reference device) DEVICE DESCRIPTION The Aurora Surgiscope System consists of two components:…
Rebound Therapeutics2019—
†K172433NICO BrainPathNico Corporation2017—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250752AURORA® Surgiscope® System (ASX15/60); AURORA® Surgiscope® System (ASX15/80)Integra LifeSciences Corporation20250
K232618Aurora Surgiscope SystemRebound Therapeutics Corporation20230

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 6, 2025Class IIZ-1295-2025Process controlPossibility for the obturator to break (separate).

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060973Vycor Surgical Access SystemsVycor Medical, LLC · 2006 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 20 years and passes through 6 companies; a new generation was cleared every 4.2 years on median.
  • 4 of the 11 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K250752 (Integra LifeSciences Corporation, 2025) has gone furthest, with 0 descendants of its own.

How does its category compare?

All GWG devices

GWG Endoscope, Neurological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · GWG

2in 2025

2008: 22013: 12014: 22016: 22017: 42018: 42019: 32020: 22021: 32023: 52024: 12025: 22026 (partial year): 4
2007high 5 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWG

269in 2023

2017: 126 days median2018: 128 days median2019: 156 days median2021: 106 days median2023: 269 days median2026 (partial year): 217 days median
2007high 269 (2023) · low 106 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWG

2in 2011

2009: 12011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWG

33in 2025

2007: 32008: 52010: 12011: 42012: 72013: 32014: 192015: 372016: 252017: 122018: 162019: 102020: 192021: 212022: 122023: 92024: 92025: 33
2007high 37 (2015) · low 0 (2009)2025

MAUDE reports naming product code GWG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—03
20082—05
20090—10
20100—01
20110—24
20120—07
20131—03
20142—019
20150—037
20162—025
20174126012
20184128016
20193156010
20202—019
20213106021
20220—012
2023526909
20241—09
20252—033
202642170—

What is it indicated for?

The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K201840: Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K201840

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026