OASIS MRI System

K202030510(k) TraditionalCleared Oct 21, 202091-day reviewClass II21 CFR 892.1000Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Healthcare Americas · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Oct 21, 2020 as substantially equivalent to K192851 (Hitachi Healthcare Americas, 2020) after a 91-day review. Its family tree reaches back 5 generations to 1995. 2 later devices descend from it. It has been recalled once.

Decided
Oct 21, 2020
Review
91 days
Class
Class II
Product code
LNH
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K930265 (1995). 6 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 5 generations backK930265 Hydrogen Spectroscopy Option-Probe #M1040HS/HTGE Medical Systems · 1995
  2. 4 generations backK991568 MR Spectroscopy PackagePhilips Medical Systems (Cleveland), Inc. · 1999
  3. 3 generations backK071506 Echelon Spectroscopy PackageHitachi Medical Systems America, Inc. · 2007
  4. 2 generations backK093044 Oasis Mri SystemHitachi Medical Systems America, Inc. · 2010 · Recalled
  5. 1 generation backK192851 OASIS MRI SystemHitachi Healthcare Americas · 2020
  6. This deviceK202030 OASIS MRI SystemHitachi Healthcare Americas · 2020
  7. Descendant · cites it directly · 1 of its ownK211406 OASIS MRI SystemHitachi Healthcare Americas · 2021
  8. Descendant · 2 generations on · 0 of its ownK240571 OASIS MRI SystemFujifilm Healthcare Corporation · 2024

Most influential descendants

Showing 2 of 2 · See all
  • K211406OASIS MRI SystemHitachi Healthcare Americas · 2021 · cites it directly1descendant
  • K240571OASIS MRI SystemFujifilm Healthcare Corporation · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K930265Hydrogen Spectroscopy Option-Probe #M1040HS/HTGE Medical Systems1995—
−4K991568MR Spectroscopy PackagePhilips Medical Systems (Cleveland), Inc.1999—
−3K071506Echelon Spectroscopy PackageHitachi Medical Systems America, Inc.2007—
−2K093044Oasis Mri SystemHitachi Medical Systems America, Inc.20101
−1K192851OASIS MRI SystemHitachi Healthcare Americas2020—
0 (this device)K202030OASIS MRI SystemHitachi Healthcare Americas2020—
+1K211406OASIS MRI SystemHitachi Healthcare Americas2021—
+2K240571OASIS MRI SystemFujifilm Healthcare Corporation2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202410
FIG. 1 — Lineage of K202030, 1995–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K192851OASIS MRI System
Summary page 4
…H Class II Panel Radiology Predicate Device Name Predicate Device(s): OASIS MRI System (K192851) Regulation Number: 21 CFR 892.1000 Regulation Name: System, Nuclear Magnetic Resonance I…
Hitachi Healthcare Americas2020—
†K073310Oasis CTL Coil, OD Flexible Body Coil (L and XL), and Rapid Foot CoilHitachi Medical Corp.2008—
†K080062Oasis Rapid Breast Coil, Model MR-Ibc-120 and Oasis Rapid NV Coil, Model MR-NVC-120Hitachi Medical Systems America, Inc.2008—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211406OASIS MRI SystemHitachi Healthcare Americas20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 29, 2021Class IIZ-2339-2021Vendor change controlThere was no label on the pad packaging or notification in the user manual that the table pads contained dry natural rubber which is required by the Code of Federal Regulations section 21CFR801.437.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 4.2 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K211406 (Hitachi Healthcare Americas, 2021) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The OASIS MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen…

Read the full statement (135 words)

The OASIS MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities

Nucleus excited: Proton

Diagnostic uses: T1, T2, proton density weighted imaging

Diffusion weighted imaging

MR Angiography

Image processing

Spectroscopy

Whole Body

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K202030: OASIS MRI System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K202030

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026