510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

CombiDiagnost R90

K203087510(k) TraditionalCleared Dec 3, 202051-day reviewClass II21 CFR 892.1650Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Dmc GmbH · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Dec 3, 2020 as substantially equivalent to K163210 (Philips Medical Systems Dmc GmbH, 2017) after a 51-day review. Its family tree reaches back 4 generations to 1992. 1 later device descends from it. It has been recalled 6 times.

Decided
Dec 3, 2020
Review
51 days
Class
Class II
Product code
JAA
Recalls
6
Predicates
1

Its longest line of descent runs back 4 generations to K924593 (1992). 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK924593 Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc. · 1992 · Recalled
  2. 3 generations backK970640 Philips Easy DiagnostPhilips Medical Systems North America, Inc. · 1997 · Recalled
  3. 2 generations backK031535 Philips Easydiagnost ElevaPhilips Medical Systems North America Co. · 2003 · Recalled
  4. 1 generation backK163210 Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2017 · Recalled
  5. This deviceK203087 CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2020
  6. Descendant · cites it directly · 0 of its ownK232910 CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2023 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K232910CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K924593Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc.19921
−3K970640Philips Easy DiagnostPhilips Medical Systems North America, Inc.19975
−2K031535Philips Easydiagnost ElevaPhilips Medical Systems North America Co.200315
−1K163210Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH20178
0 (this device)K203087CombiDiagnost R90Philips Medical Systems Dmc GmbH2020—
+1K232910CombiDiagnost R90Philips Medical Systems Dmc GmbH20231

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K203087, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K163210Philips CombiDiagnost R90
Summary page 4
…Philips Medical Systems DMC Manufacturer: K163210 – January 31, 2017 510(k) Clearance: Image…
Philips Medical Systems Dmc GmbH20178
†K202564DigitalDiagnost C90Philips Medical Systems Dmc GmbH20208

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K232910CombiDiagnost R90Philips Medical Systems Dmc GmbH20230

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 21, 2023Class IIZ-0271-2024Under Investigation by firmPotential for units suspended on the ceiling to fail and fall.
Mar 3, 2023Class IIZ-1589-2023Radiation Control for Health and Safety ActPhilips has identified a compliance issue where ProxiDiagnost N90 R.1.0, CombiDiagnost R90 R.1.0 and CombiDiagnost R90 R1.1 systems are missing a certification label required by the Food and Drug Administration (FDA).
Feb 11, 2022Class IIZ-1088-2024Software designWhile performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image…
Oct 21, 2021Class IIZ-1355-2023Device DesignIf the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from…
Sep 22, 2021Class IIZ-0241-2022Radiation Control for Health and Safety ActThe Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control…
Jul 22, 2020Class IIZ-2357-2021Radiation Control for Health and Safety ActPhilips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 4 companies; a new generation was cleared every 6.2 years on median.
  • 4 of the 4 ancestors have a recall of any class.
  • Of its descendants, K232910 (Philips Medical Systems Dmc GmbH, 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

CombiDiagnost R90 1s a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K203087: CombiDiagnost R90.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K203087

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026