Philips Easydiagnost Eleva

K031535510(k) SpecialCleared Jun 17, 200332-day reviewClass II21 CFR 892.1680Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America Co. · KPR System, X-Ray, Stationary

Cleared on Jun 17, 2003 as substantially equivalent to K970640 (Philips Medical Systems North America, Inc., 1997) after a 32-day review. Its family tree reaches back 2 generations to 1992. 6 later devices descend from it. It has been recalled 15 times.

Decided
Jun 17, 2003
Review
32 days
Class
Class II
Product code
KPR
Recalls
15
Predicates
1

Its longest line of descent runs back 2 generations to K924593 (1992). 6 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2018: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared2023: 1 descendant cleared2024: 1 descendant clearedK924593 · Diagnost 76 Family*R/F Systems - Release 4.4 · Philips Medical Systems, Inc. · 1992 · recalledK970640 · Philips Easy Diagnost · Philips Medical Systems North America, Inc. · 1997 · recalled1990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants K970640 · Philips Easy Diagnost · Philips Medical Systems North America, Inc. · 1997 · recalledK924593 · Diagnost 76 Family*R/F Systems - Release 4.4 · Philips Medical Systems, Inc. · 1992 · recalled0
  1. Root · 2 generations backK924593 Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc. · 1992 · Recalled
  2. 1 generation backK970640 Philips Easy DiagnostPhilips Medical Systems North America, Inc. · 1997 · Recalled
  3. This deviceK031535 Philips Easydiagnost ElevaPhilips Medical Systems North America Co. · 2003
  4. Descendant · cites it directly · 2 of its ownK173433 ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2018 · Recalled
  5. Descendant · cites it directly · 2 of its ownK163210 Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2017 · Recalled
  6. Descendant · 2 generations on · 1 of its ownK212837 ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2021 · Recalled
  7. Descendant · 2 generations on · 1 of its ownK203087 CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2020 · Recalled
  8. Descendant · 3 generations on · 0 of its ownK233945 ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH · 2024 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2017: 12018: 12020: 12021: 12023: 12024: 1
2017high 1 (2017) · low 0 (2019)2026

Most influential descendants

Showing 5 of 6 · See all
  • K173433ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2018 · cites it directly2descendants
  • K163210Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2017 · cites it directly2descendants
  • K212837ProxiDiagnost N90Philips Medical Systems Dmc GmbH · 2021 · 2 generations on1descendant
  • K203087CombiDiagnost R90Philips Medical Systems Dmc GmbH · 2020 · 2 generations on1descendant
  • K233945ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K924593Diagnost 76 Family*R/F Systems - Release 4.4Philips Medical Systems, Inc.19921
−1K970640Philips Easy DiagnostPhilips Medical Systems North America, Inc.19975
0 (this device)K031535Philips Easydiagnost ElevaPhilips Medical Systems North America Co.2003—
+1K173433ProxiDiagnost N90Philips Medical Systems Dmc GmbH20187
+1K163210Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH20178
+2K212837ProxiDiagnost N90Philips Medical Systems Dmc GmbH20215
+2K203087CombiDiagnost R90Philips Medical Systems Dmc GmbH20206
+3K233945ProxiDiagnost N90 / Precision CRF (706110, 706400)Philips Medical Systems Dmc GmbH20242

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201810
202010
202110
202310
202410
FIG. 1 — Lineage of K031535, 1992–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970640Philips Easy Diagnost
Summary page 2
…ally equivalent to, the Philips EasyDIAGNOST manufactured by Philips Medical Systems. The EasyDIAGNOST received 5 10(k) clearance March 2 1, 1997 ( see 5 10(k) K970640). The Eleva control concept was cleared by the FDA for the Philips MultiDiagnost Eleva (see 510(k) K023441). The intend…
Philips Medical Systems North America, Inc.19975

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173433ProxiDiagnost N90Philips Medical Systems Dmc GmbH20182
K163210Philips CombiDiagnost R90Philips Medical Systems Dmc GmbH20172

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 21, 2023Class IIZ-0279-2024Under Investigation by firmPotential for units suspended on the ceiling to fail and fall.
Jun 2, 2020Class IIZ-2523-2020Process controlThermo switches in the main power supply for the system may be incorrectly installed, resulting in a malfunction.
Jul 14, 2017Class IIZ-0311-2018Under Investigation by firmIf the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is…
Aug 19, 2014Class IIZ-0365-2016Device DesignThe hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
Oct 30, 2013Class IIZ-0431-2014Software designThe delivered application software was an outdated version to the model of equipment.
Jan 16, 2013Class IIZ-0767-2013Software design (manufacturing process)The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Jul 11, 2012Class IIZ-2208-2012Equipment maintenanceInclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
May 11, 2012Class IIZ-1821-2012Device DesignIf a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to…
Mar 9, 2012Not classifiedZ-1283-2012Software designUnderexposed image due to lag in switching between auxiliaries
Feb 14, 2012Not classifiedZ-1216-2012Software designMisinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system
Mar 6, 2009Not classifiedZ-0916-2011OtherImages from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will be overwritten with…
Jan 21, 2009Not classifiedZ-0479-2011Software designPatient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked to the incorrect…
Dec 10, 2008Not classifiedZ-0353-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Jul 9, 2008Not classifiedZ-2194-2010OtherPotential for bright artifacts to appear on the patient image from a previous exposure.
Jul 23, 2007Not classifiedZ-0910-2007OtherUnits do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 6.2 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K173433 (Philips Medical Systems Dmc GmbH, 2018) has gone furthest, with 2 descendants of its own.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The Philips EasyDiagnost Eleva intended use is for the following applications: as a multi-functional/universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with pediatric examinations and some more specialized interventional applications. This includes the following examples:

Routine Examinations

Colon examinations

Examinations of the digestive tract

Lung fluoroscopy

Examinations of the gall bladder

Thorax examinations

Skeleton imaging

Pediatric examinations

Special Applications (might require special accessories and technique)

Angiography

Myelography

Phlebography

Arthrography

Bronchography

Tomography

Sialography

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K031535: Philips Easydiagnost Eleva.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031535

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026