510(k) · Cleared
Recalled · Class I · 2025MAGNETOM Vida, MAGNETOM Sola, MAGNETOM Lumina, MAGNETOM Altea
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solutions USA, Inc. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Mar 31, 2021 as substantially equivalent to K192924 (Siemens Medical Solutions USA, Inc., 2020) after a 128-day review. Its family tree reaches back 17 generations to 1997. 14 of the 60 devices in its family, counting itself, have a Class I recall. 20 later devices descend from it. It has been recalled 2 times, including a Class I recall.
- Decided
- Mar 31, 2021
- Review
- 128 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 2 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 17 generations to K971684 (1997). 39 ancestors in all, the oldest K932271 (1994). 20 later devices descend from it, 7 citing it directly. 14 of the 60 devices in this family, counting this one, have a Class I recall.
- Root · 17 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 13 generations between
- 3 generations backK181433 MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2018
- 2 generations backK183254 Magnetom VidaSiemens Medical Solutions USA, Inc. · 2019
- 1 generation backK192924 MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida FitSiemens Medical Solutions USA, Inc. · 2020 · Recalled (Class I)
- This deviceK203443 MAGNETOM Vida, MAGNETOM Sola, MAGNETOM Lumina, MAGNETOM AlteaSiemens Medical Solutions USA, Inc. · 2021
- Descendant · cites it directly · 13 of its ownK213693 MAGNETOM Vida with syngo MR XA50ASiemens Medical Solutions USA, Inc. · 2022
- Descendant · 2 generations on · 9 of its ownK221733 MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola FitSiemens Medical Solutions USA, Inc. · 2022
- Descendant · 3 generations on · 4 of its ownK232535 MAGNETOM Sola; MAGNETOM AlteaSiemens Medical Solutions USA, Inc. · 2023
- Descendant · 3 generations on · 4 of its ownK232482 MAGNETOM Viato.MobileSiemens Medical Solutions USA, Inc. · 2023
- Descendant · 4 generations on · 3 of its ownK240608 MAGNETOM Viato.MobileSiemens Medical Solutions USA, Inc. · 2024
Show all 17 generations
- Root · 17 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 16 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
- 15 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 14 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
- 13 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
- 12 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
- 11 generations backK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008
- 10 generations backK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
- 9 generations backK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
- 8 generations backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
- 7 generations backK141977 Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with…Siemens Medical Solutions USA, Inc. · 2014
- 6 generations backK151579 MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit…Siemens Medical Solutions USA, Inc. · 2015 · Recalled (Class I)
- 5 generations backK153343 MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medi Cal Solutions, Inc. · 2016 · Recalled (Class I)
- 4 generations backK170396 MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2017 · Recalled (Class I)
- 3 generations backK181433 MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2018
- 2 generations backK183254 Magnetom VidaSiemens Medical Solutions USA, Inc. · 2019
- 1 generation backK192924 MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida FitSiemens Medical Solutions USA, Inc. · 2020 · Recalled (Class I)
- This deviceK203443 MAGNETOM Vida, MAGNETOM Sola, MAGNETOM Lumina, MAGNETOM AlteaSiemens Medical Solutions USA, Inc. · 2021
Showing 5 of 20 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 20 · See all- K213693MAGNETOM Vida with syngo MR XA50ASiemens Medical Solutions USA, Inc. · 2022 · cites it directly13descendants
- K221733MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola FitSiemens Medical Solutions USA, Inc. · 2022 · 2 generations on9descendants
- K232535MAGNETOM Sola; MAGNETOM AlteaSiemens Medical Solutions USA, Inc. · 2023 · 3 generations on4descendants
- K232482MAGNETOM Viato.MobileSiemens Medical Solutions USA, Inc. · 2023 · 3 generations on4descendants
- K240608MAGNETOM Viato.MobileSiemens Medical Solutions USA, Inc. · 2024 · 4 generations on3descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −17 | K971684 | Magnetom Project 047 System (Numaris 3.5 Software) | Siemens Medical Solutions USA, Inc. | 1997 | 4 |
| −16 | K032428 | Magnetom Avanto MR System | Siemens Medical Solutions USA, Inc. | 2003 | 4 |
| −15 | K041112 | Magnetom Espree System | Siemens Medical Solutions USA, Inc. | 2004 | 4 |
| −14 | K052164 | Magnetom Systems with Software Syngo MR 2006A | Siemens Medical Solutions USA, Inc. | 2005 | — |
| −13 | K062454 | Magnetom Systems - I-Class and T-Class Releases | Siemens Medi Cal Solutions, Inc. | 2006 | — |
| −12 | K072237 | Magnetom Verio | Siemens Medical Solutions USA, Inc. | 2007 | Class I |
| −11 | K082427 | Syngo MR B17 | Siemens Medical Solutions USA, Inc. | 2008 | — |
| −10 | K101347 | Magnetom Aera; Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2010 | Class I |
| −9 | K111242 | Magnetom Aera and Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2011 | 3 |
| −8 | K121434 | Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio | Siemens Medi Cal Solutions, Inc. | 2012 | Class I |
| −7 | K141977 | Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo… | Siemens Medical Solutions USA, Inc. | 2014 | — |
| −6 | K151579 | MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit, MAGNETOM… | Siemens Medical Solutions USA, Inc. | 2015 | Class I |
| −5 | K153343 | MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fit | Siemens Medi Cal Solutions, Inc. | 2016 | Class I |
| −4 | K170396 | MAGNETOM Vida | Siemens Medical Solutions USA, Inc. | 2017 | Class I |
| −3 | K181433 | MAGNETOM Vida | Siemens Medical Solutions USA, Inc. | 2018 | — |
| −2 | K183254 | Magnetom Vida | Siemens Medical Solutions USA, Inc. | 2019 | — |
| −1 | K192924 | MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida Fit | Siemens Medical Solutions USA, Inc. | 2020 | Class I |
| 0 (this device) | K203443 | MAGNETOM Vida, MAGNETOM Sola, MAGNETOM Lumina, MAGNETOM Altea | Siemens Medical Solutions USA, Inc. | 2021 | — |
| +1 | K213693 | MAGNETOM Vida with syngo MR XA50A | Siemens Medical Solutions USA, Inc. | 2022 | — |
| +2 | K221733 | MAGNETOM Sola, MAGNETOM Altea, MAGNETOM Sola Fit | Siemens Medical Solutions USA, Inc. | 2022 | — |
| +3 | K232535 | MAGNETOM Sola; MAGNETOM Altea | Siemens Medical Solutions USA, Inc. | 2023 | — |
| +3 | K232482 | MAGNETOM Viato.Mobile | Siemens Medical Solutions USA, Inc. | 2023 | — |
| +4 | K240608 | MAGNETOM Viato.Mobile | Siemens Medical Solutions USA, Inc. | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2000 | 2 | 0 |
| 2001 | 3 | 0 |
| 2002 | 2 | 0 |
| 2004 | 1 | 0 |
| 2005 | 3 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2022 | 6 | 2 |
| 2023 | 6 | 3 |
| 2024 | 3 | 0 |
| 2025 | 3 | 1 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
14 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K192924 | MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida FitSummary page 5…Primary: LNH Secondary: LNI, MOS 4. Legally Marketed Predicate Device Trade name: MAGNETOM Vida 510(k) Number: K192924 Classification Name: Magnetic Resonance Diagnostic Device (MRDD) Classification Panel: Radiology CFR Code:… | 2020 |
| † | K192496 | Magnetom Sola, Magnetom Altea and Magnetom Sola Fit | 2020 |
| † | K202014 | Magnetom Aera, Magnetom Skyra, Magnetom Prisma/Prisma fit | 2020 |
What descends from it? (20)
See all 207 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K213693 | MAGNETOM Vida with syngo MR XA50A | 2022 | 13 |
| K220589 | MAGNETOM Skyra Fit Class I | 2022 | 3 |
| K220151 | MAGNETOM Avanto Fit | 2022 | 3 |
| K241069 | iLoop Interventional Coil 0.55T | 2024 | 0 |
| K213805 | MAGNETOM Vida | 2023 | 0 |
| K220939 | MAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50A Class I | 2022 | 0 |
| K220425 | MAGNETOM Vida, MAGNETOM Sola | 2022 | 0 |
Has anything in its family been recalled? (14 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 28, 2025 | Class I | Z-2658-2025 | Under Investigation by firm | There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within… |
| Aug 16, 2021 | Class II | Z-2447-2021 | Software design | Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS). |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K192924MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida FitSiemens Medical Solutions USA, Inc. · 2020 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
- K170396MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2017 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K153343MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medi Cal Solutions, Inc. · 2016 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K151579MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medical Solutions USA, Inc. · 2015 · 6 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 7 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · 9 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 9 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K220939MAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50ASiemens Medical Solutions USA, Inc. · 2022 · Cites it directly · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
- K220589MAGNETOM Skyra FitSiemens Medical Solutions USA, Inc. · 2022 · Cites it directly · ◆ in the 2022 descendants bar of FIG. 12 Class I recalls
- K231587MAGNETOM Cima.XSiemens Medical Solutions USA, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
- K232494MAGNETOM Avanto fit; MAGNETOM Skyra fitSiemens Medical Solutions USA, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
- K231560MAGNETOM Vida; MAGNETOM Lumina; MAGNETOM Aera; MAGNETOM Skyra; MAGNETOM Prisma; MAGNETOM Prisma fitSiemens Medical Solutions USA, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 14 Class I recalls
Nearest 12 of 13 shown.
Nearest 5 of 13; all are on the web page.
What does this lineage say?
- The line of descent spans 24 years and passes through 2 companies; a new generation was cleared every 1.1 years on median.
- 18 of the 39 ancestors have a recall of any class; 7 have a Class I recall.
- Of its descendants, K213693 (Siemens Medical Solutions USA, Inc., 2022) has gone furthest, with 13 descendants of its own.
- 3 other devices also cite its predicate K192924.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These…
Read the full statement (120 words)
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K203443 · printed Nov 6, 2026
Cite this page
510(k) Report. “K203443: MAGNETOM Vida, MAGNETOM Sola, MAGNETOM Lumina, MAGNETOM Altea.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203443
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026