510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Intravascular Administration Sets

K203609510(k) TraditionalCleared Sep 30, 2021294-day reviewClass II21 CFR 880.5440Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corporation · FPA Set, Administration, Intravascular

Cleared on Sep 30, 2021 as substantially equivalent to K112893 (Baxter Healthcare Corporation, 2011) after a 294-day review. Its family tree reaches back 2 generations to 2000. 2 of the 7 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 5 times.

Decided
Sep 30, 2021
Review
294 days
Class
Class II
Product code
FPA
Recalls
5
Predicates
1

Its longest line of descent runs back 2 generations to K003225 (2000). 4 later devices descend from it, 4 citing it directly. 2 of the 7 devices in this family, counting this one, have a Class I recall.

  1. Root · 2 generations backK003225 Solution Administration Sets with Capless Luer Activated ValveBaxter Healthcare Corp · 2000 · Recalled
  2. 1 generation backK112893 Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter…Baxter Healthcare Corporation · 2011 · Recalled (Class I)
  3. This deviceK203609 Intravascular Administration SetsBaxter Healthcare Corporation · 2021
  4. Descendant · cites it directly · 0 of its ownK243529 Solution Administration SetsBaxter Healthcare Corporation · 2025 · Recalled (Class I)
  5. Descendant · cites it directly · 0 of its ownK221121 Volumat Polyethylene I.V. Administration SetFresenius Kabi · 2023
  6. Descendant · cites it directly · 0 of its ownK220301 Volumat LineFresenius Kabi · 2023
  7. Descendant · cites it directly · 0 of its ownK220312 Polyfusion IV Administration SetsPoly Medicure Limited · 2023

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K243529Solution Administration SetsBaxter Healthcare Corporation · 2025 · cites it directly0descendants
  • K221121Volumat Polyethylene I.V. Administration SetFresenius Kabi · 2023 · cites it directly0descendants
  • K220301Volumat LineFresenius Kabi · 2023 · cites it directly0descendants
  • K220312Polyfusion IV Administration SetsPoly Medicure Limited · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K003225Solution Administration Sets with Capless Luer Activated ValveBaxter Healthcare Corp200018
−1K112893Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension…Baxter Healthcare Corporation2011Class I
0 (this device)K203609Intravascular Administration SetsBaxter Healthcare Corporation2021—
+1K243529Solution Administration SetsBaxter Healthcare Corporation2025Class I
+1K221121Volumat Polyethylene I.V. Administration SetFresenius Kabi2023—
+1K220301Volumat LineFresenius Kabi2023—
+1K220312Polyfusion IV Administration SetsPoly Medicure Limited2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202330
202511
FIG. 1 — Lineage of K203609, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112893Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension Sets
Summary page 4
…Table 1. Predicate Device Device Company Predicate 510(k) Clearance Date Solution Administration Baxter Healthcare K112893 Oct 18, 2011 Sets Corporation DESCRIPTION OF THE DEVICE: The proposed devices consist of I…
Baxter Healthcare Corporation2011Class I
†K153158Solution Administration Sets with 0.2 Micron FilterBaxter Healthcare Corporation20152
†K150860Nitroglycerin Sets with DUO-VENT SpikeBaxter Healthcare Corporation2015—
†K161323Solution Set for Epidural UseBaxter Healthcare Corporation2016—
†K123868Interlink System Lever Lock Cannula with Check Valve, Secondary Medication Sets…Baxter Healthcare Corp2013—
†K161808Solution Set with 1.2 Micron Air Eliminating FilterBaxter Healthcare Corporation2017—
†K180739Clearlink System Solution SetBaxter Healthcare Corporation20196

Also cleared against K112893: 4 other devices →

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243529Solution Administration Sets Class IBaxter Healthcare Corporation20250
K221121Volumat Polyethylene I.V. Administration SetFresenius Kabi20230
K220301Volumat LineFresenius Kabi20230
K220312Polyfusion IV Administration SetsPoly Medicure Limited20230

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 29, 2025Class IIZ-0193-2026Process controlIV sets may leak.
Aug 29, 2025Class IIZ-0194-2026Process controlIV sets may leak.
Aug 29, 2025Class IIZ-0192-2026Process controlIV sets may leak.
Aug 8, 2025Class IIZ-2594-2025Under Investigation by firmBaxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
Dec 20, 2024Class IIZ-1011-2025Process controlSpecific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K112893Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension SetsBaxter Healthcare Corporation · 2011 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K243529Solution Administration SetsBaxter Healthcare Corporation · 2025 · Cites it directly · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 21 years and passes through 2 companies; a new generation was cleared every 11.0 years on median.
  • 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K243529 (Baxter Healthcare Corporation, 2025) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K112893.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

For the administration of fluids from a container into the patient’s vascular system through a vascular access device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K203609: Intravascular Administration Sets.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203609

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026