510(k) · Cleared
Recalled · Class I · 2025Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension Sets
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Baxter Healthcare Corporation · FPA Set, Administration, Intravascular
Cleared on Oct 18, 2011 as substantially equivalent to K003225 (Baxter Healthcare Corp, 2000) after a 15-day review. Its family tree reaches back 1 generation to 2000. 2 of the 13 devices in its family, counting itself, have a Class I recall. 11 later devices descend from it. It has been recalled 8 times, including a Class I recall.
- Decided
- Oct 18, 2011
- Review
- 15 days
- Class
- Class II
- Product code
- FPA
- Recalls
- 8 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 1 generation to K003225 (2000). 11 later devices descend from it, 5 citing it directly. 2 of the 13 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK003225 Solution Administration Sets with Capless Luer Activated ValveBaxter Healthcare Corp · 2000 · Recalled
- This deviceK112893 Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter…Baxter Healthcare Corporation · 2011
- Descendant · cites it directly · 4 of its ownK203609 Intravascular Administration SetsBaxter Healthcare Corporation · 2021 · Recalled
- Descendant · cites it directly · 2 of its ownK192366 Intravascular Extension Sets and AccessoriesBaxter Healthcare Corporation · 2020
- Descendant · 2 generations on · 0 of its ownK243529 Solution Administration SetsBaxter Healthcare Corporation · 2025 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK242339 Intravascular Extension Sets and AccessoriesBaxter Healthcare Corporation · 2025
- Descendant · 2 generations on · 0 of its ownK221121 Volumat Polyethylene I.V. Administration SetFresenius Kabi · 2023
Showing 5 of 11 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K243529 Solution Administration Sets2 generations on · 2025
Most influential descendants
Showing 5 of 11 · See all- K203609Intravascular Administration SetsBaxter Healthcare Corporation · 2021 · cites it directly4descendants
- K192366Intravascular Extension Sets and AccessoriesBaxter Healthcare Corporation · 2020 · cites it directly2descendants
- K243529Solution Administration SetsBaxter Healthcare Corporation · 2025 · 2 generations on0descendants
- K242339Intravascular Extension Sets and AccessoriesBaxter Healthcare Corporation · 2025 · 2 generations on0descendants
- K221121Volumat Polyethylene I.V. Administration SetFresenius Kabi · 2023 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K003225 | Solution Administration Sets with Capless Luer Activated Valve | Baxter Healthcare Corp | 2000 | 18 |
| 0 (this device) | K112893 | Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension… | Baxter Healthcare Corporation | 2011 | — |
| +1 | K203609 | Intravascular Administration Sets | Baxter Healthcare Corporation | 2021 | 5 |
| +1 | K192366 | Intravascular Extension Sets and Accessories | Baxter Healthcare Corporation | 2020 | — |
| +2 | K243529 | Solution Administration Sets | Baxter Healthcare Corporation | 2025 | Class I |
| +2 | K242339 | Intravascular Extension Sets and Accessories | Baxter Healthcare Corporation | 2025 | — |
| +2 | K221121 | Volumat Polyethylene I.V. Administration Set | Fresenius Kabi | 2023 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 3 | 0 |
| 2025 | 2 | 1 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FPA device
- #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K003225 | Solution Administration Sets with Capless Luer Activated ValveSummary page 2…________________________ _________ 510(k) date Modification to Solution Baxter K003225 October 19, Administration Set with Capped Luer H-ealthcare 2000 Activated Device… | 2000 |
What descends from it? (11)
See all 115 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 29, 2025 | Class II | Z-0190-2026 | Process control | IV sets may leak. |
| Aug 29, 2025 | Class II | Z-0191-2026 | Process control | IV sets may leak. |
| Aug 29, 2025 | Class II | Z-0197-2026 | Process control | IV sets may leak. |
| Aug 29, 2025 | Class II | Z-0198-2026 | Process control | IV sets may leak. |
| Aug 29, 2025 | Class II | Z-0199-2026 | Process control | IV sets may leak. |
| Aug 9, 2022 | Class I | Z-1653-2022 | Process design | Firm noted an increase in customer reports of leaks. |
| Sep 4, 2013 | Class II | Z-0099-2014 | Under Investigation by firm | The firm is issuing a voluntary recall due to the potential for pinholes in the packaging. Pinholes are a breach in the sterile barrier and could lead to potential contamination. |
| Mar 28, 2013 | Class II | Z-1271-2013 | Packaging process control | Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. |
The 5 latest of 8 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K243529Solution Administration SetsBaxter Healthcare Corporation · 2025 · 2 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
What does this lineage say?
How does its category compare?
All FPA devicesFPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · FPA
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FPA
225in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FPA
98in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · FPA
34,506in 2025
MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 75 | 17 | 1,499 |
| 2008 | 18 | 89 | 15 | 1,941 |
| 2009 | 10 | 166 | 166 | 2,512 |
| 2010 | 23 | 120 | 58 | 2,537 |
| 2011 | 12 | 87 | 28 | 4,407 |
| 2012 | 21 | 96 | 16 (4) | 4,930 |
| 2013 | 13 | 90 | 64 | 4,651 |
| 2014 | 22 | 89 | 24 | 4,695 |
| 2015 | 17 | 123 | 6 | 6,625 |
| 2016 | 12 | 195 | 10 | 8,470 |
| 2017 | 17 | 240 | 13 | 6,462 |
| 2018 | 4 | 177 | 4 | 6,574 |
| 2019 | 7 | 161 | 13 | 11,227 |
| 2020 | 10 | 327 | 107 | 9,563 |
| 2021 | 13 | 186 | 1 | 10,414 |
| 2022 | 14 | 213 | 9 (5) | 13,393 |
| 2023 | 14 | 340 | 12 (4) | 10,916 |
| 2024 | 7 | 244 | 13 (1) | 47,210 |
| 2025 | 11 | 225 | 98 (1) | 34,506 |
| 2026 | 5 | 266 | 10 (2) | — |
What is it indicated for?
For use with a vascular access device for the administration of drugs and solutions. The Clearlink Luer Activated Valve is an in-line injection site, which can be connected to standard male Luer adapters (e.g., syringes or sets) for continuous or intermittent fluid administration or the withdrawal of fluids.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K112893 · printed Nov 3, 2026
Cite this page
510(k) Report. “K112893: Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension Sets.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K112893
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026