510(k) · Cleared
ClearedNeodent EasyGuide Kit Cases
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Jjgc Industria E Comercio DE Materiais Dentarios S.A. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Cleared on Mar 9, 2021 as substantially equivalent to K192670 (Jjgc Industria E Comercio DE Materiais Dentarios S.A., 2019) after an 88-day review. Its family tree reaches back 11 generations to 1984. 1 later device descends from it.
- Decided
- Mar 9, 2021
- Review
- 88 days
- Class
- Class II
- Product code
- KCT
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 11 generations to K844002 (1984). 14 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 11 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
- 7 generations between
- 3 generations backK171713 Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2018
- 2 generations backK182865 Neodent Instruments KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
- 1 generation backK192670 Neodent Instrument Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
- This deviceK203618 Neodent EasyGuide Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2021
- Descendant · cites it directly · 0 of its ownK241492 Guided Surgery Kit CasesJJGC Indústria e Comércio de Materiais Dentários S.A. · 2024
Show all 11 generations
- Root · 11 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
- 10 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
- 9 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
- 8 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
- 7 generations backK023658 Sterilization Cases and TraysCarr Metal Products, Inc. · 2003
- 6 generations backK052992 Conmed Linvatec Instrument Sterilization TrayConmed Linvatec · 2006
- 5 generations backK142519 InterActive Complete Surgical TrayImplant Direct Sybron Manufacturing, LLC · 2015
- 4 generations backK160730 Instrument kitsAnthogyr · 2017
- 3 generations backK171713 Neodent Instrument KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2018
- 2 generations backK182865 Neodent Instruments KitsJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
- 1 generation backK192670 Neodent Instrument Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2019
- This deviceK203618 Neodent EasyGuide Kit CasesJjgc Industria E Comercio DE Materiais Dentarios S.A. · 2021
Most influential descendants
Showing 1 of 1 · See all- K241492Guided Surgery Kit CasesJJGC Indústria e Comércio de Materiais Dentários S.A. · 2024 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −11 | K844002 | HU-Friedy Ims Cassette System | Hu-Friedy Mfg. Co., Inc. | 1984 | — |
| −10 | K980065 | C/T Med-Systems Cassette System | C/T Med System, Ltd. | 1998 | — |
| −9 | K990761 | Tuttnauer Mini Container Cassette, Tuttnauer Standard Case Cassette | Tuttnauer U.S.A Co, Ltd. | 1999 | — |
| −8 | K012105 | Polyvac Surgical Instrument Delivery System | Symmetry Medical, Inc. | 2002 | — |
| −7 | K023658 | Sterilization Cases and Trays | Carr Metal Products, Inc. | 2003 | — |
| −6 | K052992 | Conmed Linvatec Instrument Sterilization Tray | Conmed Linvatec | 2006 | — |
| −5 | K142519 | InterActive Complete Surgical Tray | Implant Direct Sybron Manufacturing, LLC | 2015 | — |
| −4 | K160730 | Instrument kits | Anthogyr | 2017 | — |
| −3 | K171713 | Neodent Instrument Kits | Jjgc Industria E Comercio DE Materiais Dentarios S.A. | 2018 | — |
| −2 | K182865 | Neodent Instruments Kits | Jjgc Industria E Comercio DE Materiais Dentarios S.A. | 2019 | — |
| −1 | K192670 | Neodent Instrument Kit Cases | Jjgc Industria E Comercio DE Materiais Dentarios S.A. | 2019 | — |
| 0 (this device) | K203618 | Neodent EasyGuide Kit Cases | Jjgc Industria E Comercio DE Materiais Dentarios S.A. | 2021 | — |
| +1 | K241492 | Guided Surgery Kit Cases | JJGC Indústria e Comércio de Materiais Dentários S.A. | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
| 1997 | 1 | 0 |
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KCT device
- #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K192670 | Neodent Instrument Kit CasesSummary page 5…ssification Panel General Hospital Reviewing Branch Infection Control Devices Branch PREDICATE DEVICE INFORMATION Primary Predicate Device K192670 – Neodent Instrument Kit Cases, JJGC Indústria e Comércio de Materiais Dentários S.A… | 2019 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K241492 | Guided Surgery Kit Cases | 2024 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 37 years and passes through 9 companies; a new generation was cleared every 1.2 years on median.
- None of the 14 ancestors has a recall.
- Of its descendants, K241492 (JJGC Indústria e Comércio de Materiais Dentários S.A., 2024) has gone furthest, with 0 descendants of its own.
- 3 other devices also cite its predicate K192670.
How does its category compare?
All KCT devicesKCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · KCT
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KCT
205in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KCT
17in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KCT
50in 2025
MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 199 | 0 | 2 |
| 2008 | 6 | 182 | 0 | 8 |
| 2009 | 6 | 89 | 0 | 5 |
| 2010 | 12 | 202 | 1 | 3 |
| 2011 | 6 | 117 | 2 | 3 |
| 2012 | 10 | 236 | 0 | 2 |
| 2013 | 8 | 131 | 1 | 6 |
| 2014 | 11 | 169 | 1 | 5 |
| 2015 | 15 | 151 | 0 | 6 |
| 2016 | 12 | 182 | 3 | 16 |
| 2017 | 10 | 211 | 1 | 3 |
| 2018 | 15 | 192 | 0 | 22 |
| 2019 | 18 | 187 | 0 | 15 |
| 2020 | 14 | 134 | 0 | 66 |
| 2021 | 18 | 185 | 1 | 26 |
| 2022 | 11 | 137 | 17 | 8 |
| 2023 | 10 | 94 | 0 | 30 |
| 2024 | 8 | 106 | 0 | 18 |
| 2025 | 11 | 205 | 0 | 50 |
| 2026 | 9 | 122 | 0 | — |
What is it indicated for?
Indications for Use for GM EasyGuide Surgical Kit Case Narrow/Regular Diam Implants: Neodent Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instruments Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instruments Kits require the use of FDA-cleared wrap to maintain the sterility of enclosed devices. The kits are to be enclose in a…
Read the full statement (338 words)
Indications for Use for GM EasyGuide Surgical Kit Case Narrow/Regular Diam Implants:
Neodent Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instruments Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instruments Kits require the use of FDA-cleared wrap to maintain the sterility of enclosed devices.
The kits are to be enclose in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:
Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20 minutes dry time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minutes dry-time.
Neodent Instruments Kits are intended for sterilization of non-porous loads.
The combined weight of GM EasyGuide Surgical Kit Case Narrow/Regular Diam Implants and the associated instruments is 310,18 g. The weight of the empty Kit Case is 263,63 g
Neodent GM EasyGuide Kit Cases should not be stacked during sterilization
Indications for Use for GM EasyGuide Surgical Kit Case Regular/Large Diam Implants:
Neodent Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instruments Kits are intended to allow sterilization of the enclosed medical devices. Neodent Instruments Kits require the use of FDA-cleared wrap to maintain the sterility of enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:
Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20 minutes dry-time.
Gravity displacement - Exposure at 132 °C for 15 minutes, 45-minutes dry-time.
Neodent Instruments Kits are intended for sterilization of non-porous loads.
The combined weight of the GM EasyGuide Surgical Kit Case Regular/Large Diam Implants and the associated instruments is 346,45 g. The weight of the empty Kit Case is 264,12 g.
Neodent GM EasyGuide Kit Cases should not be stacked during sterilization.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K203618 · printed Nov 6, 2026
Cite this page
510(k) Report. “K203618: Neodent EasyGuide Kit Cases.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K203618
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026