510(k) · Cleared
ClearedCannulated Compression Device System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Orthopedic Designs North America, Inc. · HWC Screw, Fixation, Bone
Cleared on Apr 13, 2022 as substantially equivalent to K000080 (Howmedica Osteonics Corp., 2000) and 2 other predicates after a 447-day review. Its family tree reaches back 1 generation to 2000.
- Decided
- Apr 13, 2022
- Review
- 447 days
- Class
- Class II
- Product code
- HWC
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 1 generation to K000080 (2000). 3 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK000080 Asnis Iii Cannulated Screw SystemHowmedica Osteonics Corp. · 2000 · Recalled
- This deviceK210146 Cannulated Compression Device SystemOrthopedic Designs North America, Inc. · 2022
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K000080 | Asnis Iii Cannulated Screw System | Howmedica Osteonics Corp. | 2000 | 1 |
| 0 (this device) | K210146 | Cannulated Compression Device System | Orthopedic Designs North America, Inc. | 2022 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
| 2008 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HWC device
- #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K000080 | Asnis Iii Cannulated Screw SystemSummary page 4…HWC Proposed Regulatory Class: Class II Predicate Devices: Primary: Stryker® Asnis® III Cannulated Screw System (K000080) Additional: Synthes® (USA) 6.5 mm Cannulated Screw (K021932) Synthes® (USA) 4.5 and 6.5 mm He… | 2000 |
| * | K021932 | Synthes 6.5 MM Cannulated ScrewSummary page 4…evices: Primary: Stryker® Asnis® III Cannulated Screw System (K000080) Additional: Synthes® (USA) 6.5 mm Cannulated Screw (K021932) Synthes® (USA) 4.5 and 6.5 mm Headless Compression Screws (K080943) Device De… | 2002 |
| * | K080943 | Synthes 4.5 MM and 6.5 MM Headless Compression ScrewsSummary page 4…Additional: Synthes® (USA) 6.5 mm Cannulated Screw (K021932) Synthes® (USA) 4.5 and 6.5 mm Headless Compression Screws (K080943) Device Description: The Cannulated Compression Device System is used to aid in the alignment and stabilization of bon… | 2008 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- 2 of the 3 ancestors have a recall of any class.
- 20 other devices also cite its predicate K000080.
How does its category compare?
All HWC devicesHWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.
FDA decisions per year · HWC
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HWC
111in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HWC
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HWC
2,893in 2025
MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 77 | 25 | 669 |
| 2008 | 33 | 76 | 33 | 660 |
| 2009 | 27 | 113 | 29 | 651 |
| 2010 | 23 | 110 | 7 | 765 |
| 2011 | 35 | 88 | 12 | 1,104 |
| 2012 | 20 | 99 | 30 | 3,339 |
| 2013 | 25 | 90 | 14 | 3,850 |
| 2014 | 31 | 88 | 2 | 2,549 |
| 2015 | 41 | 85 | 5 | 2,711 |
| 2016 | 26 | 131 | 4 | 2,285 |
| 2017 | 40 | 127 | 4 | 2,561 |
| 2018 | 20 | 106 | 1 | 3,662 |
| 2019 | 30 | 108 | 5 | 5,463 |
| 2020 | 28 | 75 | 4 | 4,820 |
| 2021 | 26 | 84 | 2 | 4,446 |
| 2022 | 26 | 167 | 13 | 3,934 |
| 2023 | 24 | 104 | 0 | 3,718 |
| 2024 | 17 | 90 | 0 | 3,575 |
| 2025 | 15 | 111 | 1 | 2,893 |
| 2026 | 21 | 125 | 2 | — |
What is it indicated for?
The Cannulated Compression Device System is indicated for fracture fixation of small and long bones. The system is not intended for spinal use.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K210146 · printed Nov 6, 2026
Cite this page
510(k) Report. “K210146: Cannulated Compression Device System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K210146
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026