510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Symbia VA10A Family

K212604510(k) SpecialCleared Sep 16, 202130-day reviewClass II21 CFR 892.1200Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · KPS System, Tomography, Computed, Emission

Cleared on Sep 16, 2021 as substantially equivalent to K210557 (Siemens Medical Solutions USA, Inc., 2021) after a 30-day review. Its family tree reaches back 12 generations to 1995. 1 of the 30 devices in its family, counting itself, has a Class I recall. 3 later devices descend from it. It has been recalled once.

Decided
Sep 16, 2021
Review
30 days
Class
Class II
Product code
KPS
Recalls
1
Predicates
1

Its longest line of descent runs back 12 generations to K952674 (1995). 26 ancestors in all, the oldest K892358 (1989). 3 later devices descend from it, 1 citing it directly. 1 of the 30 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2017: 1 other ancestor2004: 2 other ancestors1995: 1 other ancestor2005: 1 other ancestor2003: 1 other ancestor2002: 1 other ancestor1999: 1 other ancestor1997: 1 other ancestor2001: 1 other ancestor2000: 2 other ancestors1991: 1 other ancestor1989: 1 other ancestor2023: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant clearedReference device: K200524 · SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim, go. Open Pro, Scan&GO · Siemens Medical Solutions USA, Inc. · 2020 · recalledReference device: K200474 · Symbia 6.7 · Siemens Medical Solutions USA, Inc. · 2020 · recalledReference device: K201202 · MI View&GO · Siemens Medical Solutions USA, Inc. · 2020K952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK023190 · E.Cam Computer / E.Soft Workstation · Siemens Medical Solutions USA, Inc. · 2002K063826 · Syngo MI Applications 2007A · Siemens Medical Solutions USA, Inc. · 2007K072567 · Symbia-E, Model 10275879 · Siemens Medical Solutions USA, Inc. · 2007 · recalledK082506 · Symbia, Version 4.0 · Siemens Medical Solutions USA, Inc. · 2008 · recalledK131634 · Symbia · Siemens Medical Solutions USA, Inc. · 2013 · recalledK142006 · Symbia 6.0 · Siemens Medical Solutions USA, Inc. · 2014 · recalledK162337 · Symbia 6.5 · Siemens Medical Solutions USA, Inc. · 2016 · recalledK200474 · Symbia 6.7 · Siemens Medical Solutions USA, Inc. · 2020 · recalledK210557 · Symbia VA10A Family · Siemens Medical Solutions USA, Inc. · 20211990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor7 generations back: 2 other ancestors8 generations back: 5 other ancestors9 generations back: 6 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K210557 · Symbia VA10A Family · Siemens Medical Solutions USA, Inc. · 2021K200474 · Symbia 6.7 · Siemens Medical Solutions USA, Inc. · 2020 · recalledK162337 · Symbia 6.5 · Siemens Medical Solutions USA, Inc. · 2016 · recalledK142006 · Symbia 6.0 · Siemens Medical Solutions USA, Inc. · 2014 · recalledK131634 · Symbia · Siemens Medical Solutions USA, Inc. · 2013 · recalledK082506 · Symbia, Version 4.0 · Siemens Medical Solutions USA, Inc. · 2008 · recalledK072567 · Symbia-E, Model 10275879 · Siemens Medical Solutions USA, Inc. · 2007 · recalledK063826 · Syngo MI Applications 2007A · Siemens Medical Solutions USA, Inc. · 2007K023190 · E.Cam Computer / E.Soft Workstation · Siemens Medical Solutions USA, Inc. · 2002K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalled−10−50
  1. Root · 12 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 8 generations between
  3. 3 generations backK162337 Symbia 6.5Siemens Medical Solutions USA, Inc. · 2016 · Recalled
  4. 2 generations backK200474 Symbia 6.7Siemens Medical Solutions USA, Inc. · 2020 · Recalled
  5. 1 generation backK210557 Symbia VA10A FamilySiemens Medical Solutions USA, Inc. · 2021
  6. This deviceK212604 Symbia VA10A FamilySiemens Medical Solutions USA, Inc. · 2021
  7. Descendant · cites it directly · 2 of its ownK231102 Symbia Pro.specta VA20A FamilySiemens Medical Solutions USA, Inc. · 2023
  8. Descendant · 2 generations on · 1 of its ownK243186 Symbia Pro.specta Q3 (11364751); Symbia Pro.specta X3 (11364752)…Siemens Medical Solutions USA, Inc. · 2024
  9. Descendant · 3 generations on · 0 of its ownK261315 Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3…Siemens Medical Solutions USA, Inc. · 2026
Show all 12 generations
  1. Root · 12 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 11 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 10 generations backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. 9 generations backK023190 E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc. · 2002
  5. 8 generations backK063826 Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc. · 2007
  6. 7 generations backK072567 Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc. · 2007 · Recalled
  7. 6 generations backK082506 Symbia, Version 4.0Siemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 5 generations backK131634 SymbiaSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  9. 4 generations backK142006 Symbia 6.0Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  10. 3 generations backK162337 Symbia 6.5Siemens Medical Solutions USA, Inc. · 2016 · Recalled
  11. 2 generations backK200474 Symbia 6.7Siemens Medical Solutions USA, Inc. · 2020 · Recalled
  12. 1 generation backK210557 Symbia VA10A FamilySiemens Medical Solutions USA, Inc. · 2021
  13. This deviceK212604 Symbia VA10A FamilySiemens Medical Solutions USA, Inc. · 2021

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K231102Symbia Pro.specta VA20A FamilySiemens Medical Solutions USA, Inc. · 2023 · cites it directly2descendants
  • K243186Symbia Pro.specta Q3 (11364751); Symbia Pro.specta X3 (11364752); Symbia Pro.specta X7…Siemens Medical Solutions USA, Inc. · 2024 · 2 generations on1descendant
  • K261315Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia Pro.specta X7…Siemens Medical Solutions USA, Inc. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−11K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−10K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
−9K023190E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc.2002—
−8K063826Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc.2007—
−7K072567Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc.20072
−6K082506Symbia, Version 4.0Siemens Medical Solutions USA, Inc.20089
−5K131634SymbiaSiemens Medical Solutions USA, Inc.20137
−4K142006Symbia 6.0Siemens Medical Solutions USA, Inc.20149
−3K162337Symbia 6.5Siemens Medical Solutions USA, Inc.20164
−2K200474Symbia 6.7Siemens Medical Solutions USA, Inc.20202
−1K210557Symbia VA10A FamilySiemens Medical Solutions USA, Inc.2021—
0 (this device)K212604Symbia VA10A FamilySiemens Medical Solutions USA, Inc.2021—
+1K231102Symbia Pro.specta VA20A FamilySiemens Medical Solutions USA, Inc.2023—
+2K243186Symbia Pro.specta Q3 (11364751); Symbia Pro.specta X3 (11364752); Symbia…Siemens Medical Solutions USA, Inc.2024—
+3K261315Symbia Pro.specta Q3 (11364751);Symbia Pro.specta X3 (11364752);Symbia…Siemens Medical Solutions USA, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199110
199510
199710
199910
200020
200110
200210
200310
200420
200510
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
202610
FIG. 1 — Lineage of K212604, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K210557Symbia VA10A Family
Summary page 5
…KPS and JAK Classification Panel: Radiology Device Class: Class II 510(k) Number: K210557 Reference Device: Device Name and 510(k) Scan&GO, SOMATOM Go.Platform numbers: Common Name:…
Siemens Medical Solutions USA, Inc.2021—
†K200524SOMATOM X.cite, Scan&GO, SOMATOM go.Platform - go.Up, go.Now, go.Top, go.All, go. Sim…Siemens Medical Solutions USA, Inc.20207
†K200474Symbia 6.7Siemens Medical Solutions USA, Inc.20202
†K201202MI View&GOSiemens Medical Solutions USA, Inc.2020—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231102Symbia Pro.specta VA20A FamilySiemens Medical Solutions USA, Inc.20232

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 12, 2023Class IIZ-2256-2023Device DesignDuring a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · 8 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 3 companies; a new generation was cleared every 2.1 years on median.
  • 17 of the 26 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K231102 (Siemens Medical Solutions USA, Inc., 2023) has gone furthest, with 2 descendants of its own.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The Siemens Symbia VA10A Family is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment…

Read the full statement (270 words)

The Siemens Symbia VA10A Family is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses.

SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV.

CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles.

SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient’s physiological (SPECT) and anatomical (CT) images.

Software: The SPECTsyngo software is an acquisition, display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities.

This CT system can be used for low dose lung cancer screening in high-risk populations.*

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K212604: Symbia VA10A Family.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K212604

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026