510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee System with LINK PorEx (TiNbN) inhouse coating

K212742510(k) SpecialCleared Sep 16, 202117-day reviewClass II21 CFR 888.3560Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Waldemar Link GmbH & Co. KG · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Sep 16, 2021 as substantially equivalent to K182872 (Waldemar Link GmbH & Co. KG, 2019) and 1 other predicate after a 17-day review. Its family tree reaches back 6 generations to 1994. 1 later device descends from it. It has been recalled 22 times.

Decided
Sep 16, 2021
Review
17 days
Class
Class II
Product code
JWH
Recalls
22
Predicates
2

Its longest line of descent runs back 6 generations to K936079 (1994). 26 ancestors in all, the oldest K810520 (1981). 1 later device descends from it, 1 citing it directly.

  1. Root · 6 generations backK936079 Ecactech Porous Coated Finned Tibial Tray CompnentExactech, Inc. · 1994
  2. 5 generations backK011976 Optetrak Total Knee System Size 0/1 Delta (Line Extension)Exactech, Inc. · 2001
  3. 4 generations backK030686 Optetrak Total Knee System - Molded Metal-Back Tibial ComponentExactech, Inc. · 2003
  4. 3 generations backK100900 Motivation Knee SystemEncore Medical L.P. · 2010 · Recalled
  5. 2 generations backK122239 Foundation Knee System with Tinbn Coating Foundation PS Knee System…Encore Medical L.P. · 2013
  6. 1 generation backK152431 LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled…Waldemar Link GmbH & Co. KG · 2016 · Recalled
  7. This deviceK212742 GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee…Waldemar Link GmbH & Co. KG · 2021
  8. Descendant · cites it directly · 0 of its ownK220628 LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG · 2022

Most influential descendants

Showing 1 of 1 · See all
  • K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K936079Ecactech Porous Coated Finned Tibial Tray CompnentExactech, Inc.1994—
−5K011976Optetrak Total Knee System Size 0/1 Delta (Line Extension)Exactech, Inc.2001—
−4K030686Optetrak Total Knee System - Molded Metal-Back Tibial ComponentExactech, Inc.2003—
−3K100900Motivation Knee SystemEncore Medical L.P.20104
−2K122239Foundation Knee System with Tinbn Coating Foundation PS Knee System with Tinbn…Encore Medical L.P.2013—
−1K152431LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System…Waldemar Link GmbH & Co. KG201616
0 (this device)K212742GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee System with…Waldemar Link GmbH & Co. KG2021—
+1K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198410
199110
199310
199440
199530
199640
200010
200220
201510
201910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K212742, 1994–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K182872LINK GEMINI SL Total Knee System
Summary page 5
…Class II, Orthopedic / 87 Predicate Devices: Primary Predicate: GEMINI SL Total Knee System, K182872 Endo-Model Knee System with PorEx (TiNbN) coating, Sled…
Waldemar Link GmbH & Co. KG2019—
*K152431LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System with PorEx…
Summary page 5
…Endo-Model Knee System with PorEx (TiNbN) coating, Sled Knee System with PorEx (TiNbN) coating, K152431 Page 1 of 5…
Waldemar Link GmbH & Co. KG201616
†K202924LinkSymphoKnee SystemWaldemar Link GmbH & Co. KG202111

Also cleared against K182872: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 22, 2024Class IIZ-2410-2024Process designBlind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
May 22, 2024Class IIZ-2411-2024Process designBlind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
Mar 21, 2023Class IIZ-1544-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1549-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1552-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1543-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1551-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1550-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1541-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1554-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1540-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1553-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1548-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1545-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1539-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1547-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1542-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
Mar 21, 2023Class IIZ-1546-2023Labeling designAdditional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the…
May 24, 2022Class IIZ-1380-2022Nonconforming Material/ComponentThere is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
May 24, 2022Class IIZ-1377-2022Nonconforming Material/ComponentThere is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
May 24, 2022Class IIZ-1378-2022Nonconforming Material/ComponentThere is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
May 24, 2022Class IIZ-1379-2022Nonconforming Material/ComponentThere is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 3 companies; a new generation was cleared every 5.4 years on median.
  • 14 of the 26 ancestors have a recall of any class.
  • Of its descendants, K220628 (Waldemar Link GmbH & Co. KG, 2022) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K182872.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

The GEMINI SL Total Knee System is indicated for patients suffering from disability due to: 1) Degenerative, post-traumatic or rheumatoid arthritis; 2) Avascular necrosis of the femoral condyle; 3) Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; 4) Moderate valgus, varus or flexion deformities. This device may also be indicated in the salvage of previously failed surgical…

Read the full statement (233 words)

The GEMINI SL Total Knee System is indicated for patients suffering from disability due to:

1) Degenerative, post-traumatic or rheumatoid arthritis;

2) Avascular necrosis of the femoral condyle;

3) Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy;

4) Moderate valgus, varus or flexion deformities.

This device may also be indicated in the salvage of previously failed surgical attempts.

The device is indicated for cemented use. Only cementless labeled modular stems are indicated for uncemented use.

The Endo-Model and Sled Knee Systems with LINK PorEx (TiNbN) coating are indicated for patients with severe joint diseases with limitation of mobility due to degenerative, rheumatoid or post-traumatic arthrosis or arthritis. Joint fractures which disallow an osteosynthetic reconstruction. These devices are intended for cemented use only.

The Endo-Model Rotating Hinge and Modular Rotating Hinge Knee System with LINK PorEx (TiNbN) coating are indicated for the following conditions:

1) Bone necroses

2) Bicondylar arthrosis by partly damaged collateral ligaments

3) Revision after primary total knee replacement

4) Revision surgery after hinge knee or rotational knee joint

5) Revision surgery by insufficient / inadequate bone mass

6) Arthrosis of patella flange

7) Valgus/Varus deformities <10°

8) Valgus/Varus deformities 10-15°

9) Valgus/Varus deformities 15-20°

The Sled Knee System with LINK PorEx (TiNbN) coating is indicated for the following conditions:

1) Unicondylar arthrosis by intact ligaments including both cruciate ligaments

2) Valgus/Varus deformities <10°

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K212742: GEMINI SL Total Knee System, Endo-Model Knee System, and Sled Knee System with LINK PorEx (TiNbN) inhouse coating.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K212742

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026