510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

LinkSymphoKnee System

K202924510(k) TraditionalCleared May 13, 2021226-day reviewClass II21 CFR 888.3560Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Waldemar Link GmbH & Co. KG · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on May 13, 2021 as substantially equivalent to K182872 (Waldemar Link GmbH & Co. KG, 2019) and 7 other predicates after a 226-day review. Its family tree reaches back 6 generations to 1994. 2 later devices descend from it. It has been recalled 11 times.

Decided
May 13, 2021
Review
226 days
Class
Class II
Product code
JWH
Recalls
11
Predicates
8

Its longest line of descent runs back 6 generations to K936079 (1994). 32 ancestors in all, the oldest K810520 (1981). 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2019: 1 other ancestor2015: 1 other ancestor2007: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor1996: 6 other ancestors1995: 4 other ancestors2002: 2 other ancestors1994: 4 other ancestors1993: 1 other ancestor2000: 1 other ancestor1991: 1 other ancestor1984: 1 other ancestor1981: 1 other ancestor2023: 1 descendant cleared2022: 1 descendant clearedK936079 · Ecactech Porous Coated Finned Tibial Tray Compnent · Exactech, Inc. · 1994K011976 · Optetrak Total Knee System Size 0/1 Delta (Line Extension) · Exactech, Inc. · 2001K030686 · Optetrak Total Knee System - Molded Metal-Back Tibial Component · Exactech, Inc. · 2003K100900 · Motivation Knee System · Encore Medical L.P. · 2010 · recalledK122239 · Foundation Knee System with Tinbn Coating Foundation PS Knee System with Tinbn Coating 3DKNEE System with Tinbn Coating · Encore Medical L.P. · 2013K152431 · LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System with PorEx (TiNbN) coating · Waldemar Link GmbH & Co. KG · 2016 · recalled19801990200020102020
  1. Root · 6 generations backK936079 Ecactech Porous Coated Finned Tibial Tray CompnentExactech, Inc. · 1994
  2. 5 generations backK011976 Optetrak Total Knee System Size 0/1 Delta (Line Extension)Exactech, Inc. · 2001
  3. 4 generations backK030686 Optetrak Total Knee System - Molded Metal-Back Tibial ComponentExactech, Inc. · 2003
  4. 3 generations backK100900 Motivation Knee SystemEncore Medical L.P. · 2010 · Recalled
  5. 2 generations backK122239 Foundation Knee System with Tinbn Coating Foundation PS Knee System…Encore Medical L.P. · 2013
  6. 1 generation backK152431 LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled…Waldemar Link GmbH & Co. KG · 2016 · Recalled
  7. This deviceK202924 LinkSymphoKnee SystemWaldemar Link GmbH & Co. KG · 2021
  8. Descendant · cites it directly · 0 of its ownK230471 LinkSymphoKnee – Fixed Bearing Ultracongruent (FB UC) Articulating…Waldemar Link GmbH & Co. KG · 2023
  9. Descendant · cites it directly · 0 of its ownK220628 LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG · 2022

Most influential descendants

Showing 2 of 2 · See all
  • K230471LinkSymphoKnee – Fixed Bearing Ultracongruent (FB UC) Articulating SurfaceWaldemar Link GmbH & Co. KG · 2023 · cites it directly0descendants
  • K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K936079Ecactech Porous Coated Finned Tibial Tray CompnentExactech, Inc.1994—
−5K011976Optetrak Total Knee System Size 0/1 Delta (Line Extension)Exactech, Inc.2001—
−4K030686Optetrak Total Knee System - Molded Metal-Back Tibial ComponentExactech, Inc.2003—
−3K100900Motivation Knee SystemEncore Medical L.P.20104
−2K122239Foundation Knee System with Tinbn Coating Foundation PS Knee System with Tinbn…Encore Medical L.P.2013—
−1K152431LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System…Waldemar Link GmbH & Co. KG201616
0 (this device)K202924LinkSymphoKnee SystemWaldemar Link GmbH & Co. KG2021—
+1K230471LinkSymphoKnee – Fixed Bearing Ultracongruent (FB UC) Articulating SurfaceWaldemar Link GmbH & Co. KG2023—
+1K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198410
199110
199310
199440
199540
199660
200010
200220
200410
200510
200710
201510
201910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202310
FIG. 1 — Lineage of K202924, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (8)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K182872LINK GEMINI SL Total Knee System
Summary page 4
…cation and Panel: Class II, Orthopedic / 87 Predicate Devices: Primary Predicate: Waldemar Link LINK GEMINI SL Total Knee System: K182872 Additional Predicates: Smith & Nephew Legion Knee System: K072531, K0411…
Waldemar Link GmbH & Co. KG2019—
*K072531Legion Stem with Holes
Summary page 4
…aldemar Link LINK GEMINI SL Total Knee System: K182872 Additional Predicates: Smith & Nephew Legion Knee System: K072531, K041106, K043440; K953274; K951987…
Smith & Nephew, Inc.2007—
*K041106Revision Knee System
Summary page 4
…ink LINK GEMINI SL Total Knee System: K182872 Additional Predicates: Smith & Nephew Legion Knee System: K072531, K041106, K043440; K953274; K951987…
Smith & Nephew, Inc.20041
*K043440Smith & Nephew, Inc. Revision Knee System
Summary page 4
…GEMINI SL Total Knee System: K182872 Additional Predicates: Smith & Nephew Legion Knee System: K072531, K041106, K043440; K953274; K951987…
Smith & Nephew, Inc.20057
*K953274Genesis Porous Tibial Trays & Revision Knee System
Summary page 4
…Additional Predicates: Smith & Nephew Legion Knee System: K072531, K041106, K043440; K953274; K951987 K202924 Page 2 of 4…
Smith & Nephew Richards, Inc.1996—
*K951987Genesis II Total Knee System
Summary page 4
…Additional Predicates: Smith & Nephew Legion Knee System: K072531, K041106, K043440; K953274; K951987 K202924 Page 2 of 4…
Smith & Nephew Richards, Inc.19956
*K143179The LINK® Endo-Model® Knee System
Summary page 5
…= Page 5 === K202924 Page 2 of 4 Waldemar Link Endo-Model Knee System: K143179 Waldemar Link Endo-Model Knee System with PorEx (TiNbN) coating: K152431 Device Desc…
Waldemar Link GmbH & Co. KG20157
*K152431LINK Endo-Model Knee System with PorEx (TiNbN) coating, Link Sled Knee System with PorEx…
Summary page 5
…Waldemar Link Endo-Model Knee System: K143179 Waldemar Link Endo-Model Knee System with PorEx (TiNbN) coating: K152431 Device Description: The LinkSymphoKnee System is available in multiple versions with different applicati…
Waldemar Link GmbH & Co. KG201616

Also cleared against K182872: 3 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230471LinkSymphoKnee – Fixed Bearing Ultracongruent (FB UC) Articulating SurfaceWaldemar Link GmbH & Co. KG20230
K220628LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 31, 2025Class IIZ-2541-2025Under Investigation by firmSome modular LinkSymphoKnee TiNbN Tibial Components were manufactured with two uncoated security screws, which is incorrect.
Apr 3, 2025Class IIZ-1701-2025Nonconforming Material/ComponentThe device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Apr 3, 2025Class IIZ-1702-2025Nonconforming Material/ComponentThe device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Apr 3, 2025Class IIZ-1699-2025Nonconforming Material/ComponentThe device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Apr 3, 2025Class IIZ-1700-2025Nonconforming Material/ComponentThe device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Feb 14, 2025Class IIZ-1373-2025Labeling Change ControlA problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
Feb 14, 2025Class IIZ-1372-2025Labeling Change ControlA problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
Feb 14, 2025Class IIZ-1374-2025Labeling Change ControlA problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
Apr 25, 2023Class IIZ-1674-2023Under Investigation by firmContains the incorrect implant in the packaging. The packaging incorrectly contains the opposite side of the implant (Lateral-Right/Medial-Left instead of Medial-Right/Lateral-Left).
Dec 23, 2022Class IIZ-2067-2023Component change controlFixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).
Dec 23, 2022Class IIZ-2066-2023Component change controlFixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 3 companies; a new generation was cleared every 5.0 years on median.
  • 18 of the 32 ancestors have a recall of any class.
  • Of its descendants, K230471 (Waldemar Link GmbH & Co. KG, 2023) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K182872.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

General:

The LinkSymphoKnee System is intended for primary and revision total knee replacement in skeletally mature patients with the following conditions.

This device is intended for cemented use only unless a cementless modular stem is indicated for use.

Indications:

• Primary degenerative arthritis / osteoarthritis

• Secondary arthritis resulting from rheumatoid arthritis

• Fracture

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K202924: LinkSymphoKnee System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K202924

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026