510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Vortex Surgical Laser Probes, Vortex Surgical Illuminated Laser Probes, Vortex Surgical Endoilluminators, Vortex Surgical Chandeliers

K220263510(k) TraditionalCleared Nov 3, 2022276-day reviewClass II21 CFR 886.4690Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vortex Surgical, Inc. · HQB Photocoagulator And Accessories

Cleared on Nov 3, 2022 as substantially equivalent to K191846 (Vortex Surgical, Inc., 2019) after a 276-day review. Its family tree reaches back 6 generations to 1987. It has been recalled 3 times.

Decided
Nov 3, 2022
Review
276 days
Class
Class II
Product code
HQB
Recalls
3
Predicates
1

Its longest line of descent runs back 6 generations to K870942 (1987). 11 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK870942 MVS Fiberoptic LightpipeAdvanced Surgical Products, Inc. · 1987
  2. 5 generations backK982462 Fiber Optic Endoillumination ProbeMicrovision, Inc. · 1998
  3. 4 generations backK050807 20G Sma Laser Fiber, Model LF100Ophthalmed, LLC · 2005
  4. 3 generations backK121187 Katalyst Laser Probes, Katalyst Illuminated Laser ProbesKatalyst Surgical, LLC · 2012
  5. 2 generations backK142830 Ophthalmed Bending Laser ProbesOphthalmed, LLC · 2015
  6. 1 generation backK191846 MAXReach Laser ProbeVortex Surgical, Inc. · 2019
  7. This deviceK220263 Vortex Surgical Laser Probes, Vortex Surgical Illuminated Laser…Vortex Surgical, Inc. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K870942MVS Fiberoptic LightpipeAdvanced Surgical Products, Inc.1987—
−5K982462Fiber Optic Endoillumination ProbeMicrovision, Inc.1998—
−4K05080720G Sma Laser Fiber, Model LF100Ophthalmed, LLC2005—
−3K121187Katalyst Laser Probes, Katalyst Illuminated Laser ProbesKatalyst Surgical, LLC2012—
−2K142830Ophthalmed Bending Laser ProbesOphthalmed, LLC2015—
−1K191846MAXReach Laser ProbeVortex Surgical, Inc.2019—
0 (this device)K220263Vortex Surgical Laser Probes, Vortex Surgical Illuminated Laser Probes, Vortex…Vortex Surgical, Inc.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199520
200110
200210
201210
FIG. 1 — Lineage of K220263, 1987–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HQB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K191846MAXReach Laser Probe
Summary page 4
…Laser Probe Common Name: Ophthalmic Laser Vortex Branded Name: Direct Flex Laser Probe 510(k) Number: K191846 Regulatory class: II Product Codes: HQB Regulation Number: 21 CF…
Vortex Surgical, Inc.2019—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 16, 2025Class IIZ-1246-2026Employee errorThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Dec 16, 2025Class IIZ-1245-2026Employee errorThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Nov 19, 2025Class IIZ-1004-2026Component design/selectionReason for the voluntary recall is some products in this lot are not passing through a 25ga cannula smoothly.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 5 companies; a new generation was cleared every 7.0 years on median.
  • 1 of the 11 ancestors has a recall of any class.

How does its category compare?

All HQB devices

HQB Photocoagulator And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · HQB

1in 2022

2011: 32012: 12013: 32019: 22022: 1
2007high 3 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HQB

221in 2013

2011: 57 days median2013: 221 days median
2007high 221 (2013) · low 57 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HQB

2in 2025

2013: 282025: 2
2007high 28 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HQB

7in 2025

2011: 32012: 32013: 22014: 12016: 12017: 22018: 32019: 12020: 142021: 12022: 32023: 32024: 32025: 7
2007high 14 (2020) · low 0 (2007)2025

MAUDE reports naming product code HQB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
201135703
20121—03
20133221282
20140—01
20150—00
20160—01
20170—02
20180—03
20192—01
20200—014
20210—01
20221—03
20230—03
20240—03
20250—27
20260—0—

What is it indicated for?

Vortex Surgical Laser Probes and Illuminated Laser Probes are indicated for use in laser endophotocoagulation procedures in the posterior segment of the eye during vitreoretinal surgery at 500nm to 1100nm. Vortex Surgical Illuminated Laser Probes, Endoilluminator, and Chandelier are indicated for illumination during vitreoretinal surgery with visible light.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220263: Vortex Surgical Laser Probes, Vortex Surgical Illuminated Laser Probes, Vortex Surgical Endoilluminators, Vortex Surgical Chandeliers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220263

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026