510(k) Report

510(k) · Cleared

Cleared

Acetabular Dome Hole Plug

K220376510(k) SpecialCleared Mar 9, 202227-day reviewClass II21 CFR 888.3358

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Howmedica Osteonics Corp., Dba Stryker Orthopaedics · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Mar 9, 2022 as substantially equivalent to K942809 (Osteonics Corp., 1994) and 1 other predicate after a 27-day review. Its family tree reaches back 10 generations to 1994. 5 later devices descend from it.

Decided
Mar 9, 2022
Review
27 days
Class
Class II
Product code
LPH
Recalls
None
Predicates
2

Its longest line of descent runs back 10 generations to K921308 (1994). 46 ancestors in all, the oldest K903362 (1990). 5 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1994: 6 other ancestors2016: 1 other ancestor1998: 2 other ancestors1992: 1 other ancestor2013: 1 other ancestor2008: 1 other ancestor2001: 2 other ancestors2011: 1 other ancestor2002: 2 other ancestors2000: 5 other ancestors1995: 7 other ancestors2009: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1990: 1 other ancestor2023: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant cleared2026: 1 descendant clearedReference device: K153345 · Stryker Orthopaedics Hip Systems Labeling Update · Stryker Orthopaedics · 2016 · recalledK921308 · HG Multilock Hip Prosthesis W/ TI-Nidium · Zimmer, Inc. · 1994 · recalledK972893 · Partnership Revision Femoral Stems · Howmedica Corp. · 1997K983404 · Type 3 Femoral Components · Howmedica Corp. · 1998 · recalledK013106 · 2 Piece Modular Hip Stem · Howmedica Osteonics Corp. · 2002 · recalledK103479 · Novel Tapered Hip Stem · Stryker Corp. · 2011K120578 · Accolade II Hip Stem · Howmedica Osteonics · 2012 · recalledK122853 · Novel Fit and Fill Stem · Howmedica Osteonics Corp. · 2013K143085 · Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip Instruments for use with the Direct Superior Approach · Stryker Orthopaedics · 2015 · recalledK171768 · Trident® II Acetabular System · Howmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 2017 · recalledK191358 · Trident II Acetabular System · Howmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 20191990200020102020
  1. Root · 10 generations backK921308 HG Multilock Hip Prosthesis W/ TI-NidiumZimmer, Inc. · 1994 · Recalled
  2. 6 generations between
  3. 3 generations backK143085 Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip…Stryker Orthopaedics · 2015 · Recalled
  4. 2 generations backK171768 Trident® II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 2017 · Recalled
  5. 1 generation backK191358 Trident II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 2019
  6. This deviceK220376 Acetabular Dome Hole PlugHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2022
  7. Descendant · cites it directly · 4 of its ownK233498 Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2023
  8. Descendant · 2 generations on · 2 of its ownK243784 Stryker Orthopaedics Hip Devices Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2025
  9. Descendant · 2 generations on · 0 of its ownK261624 Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2026 · Recalled
  10. Descendant · 2 generations on · 0 of its ownK250989 Stryker and Serf hip devicesHowmedica Osteonics Corp (Dba Stryker Orthopaedics) · 2025
  11. Descendant · 2 generations on · 0 of its ownK240418 Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2024
Show all 10 generations
  1. Root · 10 generations backK921308 HG Multilock Hip Prosthesis W/ TI-NidiumZimmer, Inc. · 1994 · Recalled
  2. 9 generations backK972893 Partnership Revision Femoral StemsHowmedica Corp. · 1997
  3. 8 generations backK983404 Type 3 Femoral ComponentsHowmedica Corp. · 1998 · Recalled
  4. 7 generations backK013106 2 Piece Modular Hip StemHowmedica Osteonics Corp. · 2002 · Recalled
  5. 6 generations backK103479 Novel Tapered Hip StemStryker Corp. · 2011
  6. 5 generations backK120578 Accolade II Hip StemHowmedica Osteonics · 2012 · Recalled
  7. 4 generations backK122853 Novel Fit and Fill StemHowmedica Osteonics Corp. · 2013
  8. 3 generations backK143085 Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip…Stryker Orthopaedics · 2015 · Recalled
  9. 2 generations backK171768 Trident® II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 2017 · Recalled
  10. 1 generation backK191358 Trident II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics · 2019
  11. This deviceK220376 Acetabular Dome Hole PlugHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2022

Most influential descendants

Showing 5 of 5 · See all
  • K233498Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2023 · cites it directly4descendants
  • K243784Stryker Orthopaedics Hip Devices Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2025 · 2 generations on2descendants
  • K261624Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2026 · 2 generations on0descendants
  • K250989Stryker and Serf hip devicesHowmedica Osteonics Corp (Dba Stryker Orthopaedics) · 2025 · 2 generations on0descendants
  • K240418Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K921308HG Multilock Hip Prosthesis W/ TI-NidiumZimmer, Inc.19941
−9K972893Partnership Revision Femoral StemsHowmedica Corp.1997—
−8K983404Type 3 Femoral ComponentsHowmedica Corp.19981
−7K0131062 Piece Modular Hip StemHowmedica Osteonics Corp.200237
−6K103479Novel Tapered Hip StemStryker Corp.2011—
−5K120578Accolade II Hip StemHowmedica Osteonics20121
−4K122853Novel Fit and Fill StemHowmedica Osteonics Corp.2013—
−3K143085Stryker Orthopaedics Acetabular Shells, Femoral Stems and Total Hip…Stryker Orthopaedics20152
−2K171768Trident® II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics201712
−1K191358Trident II Acetabular SystemHowmedica Osteonics Corp. A.K.A. Stryker Orthopaedics2019—
0 (this device)K220376Acetabular Dome Hole PlugHowmedica Osteonics Corp., Dba Stryker Orthopaedics2022—
+1K233498Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics2023—
+2K243784Stryker Orthopaedics Hip Devices Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics2025—
+2K261624Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert…Howmedica Osteonics Corp., Dba Stryker Orthopaedics20262
+2K250989Stryker and Serf hip devicesHowmedica Osteonics Corp (Dba Stryker Orthopaedics)2025—
+2K240418Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199210
199460
199570
199630
199720
199820
200050
200120
200220
200810
200910
201110
201310
201610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
202520
202610
FIG. 1 — Lineage of K220376, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K942809Acetabular Dome Hole Plugs
Summary page 5
…Product Codes: LPH, KWZ, LZO, MEH, JDI Legally Marketed Predicate Device to Which Substantial Equivalence is Claimed: • Acetabular Dome Hole Plugs (K942809, K191358) 5-2…
Osteonics Corp.1994—
*K191358Trident II Acetabular System
Summary page 5
…ct Codes: LPH, KWZ, LZO, MEH, JDI Legally Marketed Predicate Device to Which Substantial Equivalence is Claimed: • Acetabular Dome Hole Plugs (K942809, K191358) 5-2…
Howmedica Osteonics Corp. A.K.A. Stryker Orthopaedics2019—
†K153345Stryker Orthopaedics Hip Systems Labeling UpdateStryker Orthopaedics20161

Also cleared against K942809: 2 other devices →

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233498Stryker Orthopaedics Hip Systems Labeling UpdateHowmedica Osteonics Corp., Dba Stryker Orthopaedics20234

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 8 companies; a new generation was cleared every 2.4 years on median.
  • 19 of the 46 ancestors have a recall of any class.
  • Of its descendants, K233498 (Howmedica Osteonics Corp., Dba Stryker Orthopaedics, 2023) has gone furthest, with 4 descendants of its own.
  • 2 other devices also cite its predicate K942809.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—

What is it indicated for?

In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous. Indications for use when used with the Trident II Acetabular Shells. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. Clinical management…

Read the full statement (179 words)

In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.

Indications for use when used with the Trident II Acetabular Shells.

Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.

Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.

When used with MDM Liners

Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Dislocation risks

When used with Constrained Liners

The Trident® Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.

The Trident II Acetabular Shells are indicated for cementless use only.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K220376: Acetabular Dome Hole Plug.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220376

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026