510(k) Report

510(k) · Cleared

Cleared

Consolidated HC Steam Sterilizer

K220736510(k) AbbreviatedCleared Aug 1, 2022140-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Consolidated Machine Corp. · FLE Sterilizer, Steam

Cleared on Aug 1, 2022 as substantially equivalent to K183410 (STERIS Corporation, 2019) after a 140-day review. Its family tree reaches back 4 generations to 2001.

Decided
Aug 1, 2022
Review
140 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K010865 (2001). No later device cites it as a predicate yet.

  1. Root · 4 generations backK010865 Amsco Century Medium Sterilizer, Model CenturySTERIS Corporation · 2001 · Recalled
  2. 3 generations backK020747 Amsco Century Medium Sterilizer 26 X 26STERIS Corporation · 2002
  3. 2 generations backK112403 Amsco Chimeron Medium Steam SterlilzerSTERIS Corporation · 2012 · Recalled
  4. 1 generation backK183410 AMSCO 600 Steam SterilizerSTERIS Corporation · 2019
  5. This deviceK220736 Consolidated HC Steam SterilizerConsolidated Machine Corp. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K010865Amsco Century Medium Sterilizer, Model CenturySTERIS Corporation20011
−3K020747Amsco Century Medium Sterilizer 26 X 26STERIS Corporation2002—
−2K112403Amsco Chimeron Medium Steam SterlilzerSTERIS Corporation20121
−1K183410AMSCO 600 Steam SterilizerSTERIS Corporation2019—
0 (this device)K220736Consolidated HC Steam SterilizerConsolidated Machine Corp.2022—
FIG. 1 — Lineage of K220736, 2001–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K183410AMSCO 600 Steam Sterilizer
Summary page 7
…Steam Sterilizer/Autoclave Classification Name: Sterilizer, Steam Classification Number: 21 CFR 880.6880 Product Code: FLE Predicate Device K183410 AMSCO 600 Steam Sterilizer Device Class: Class II Common Name: Steam Sterilizer/Autoclave Classification…
STERIS Corporation2019—

Also cleared against K183410: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 2 companies; a new generation was cleared every 6.9 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K183410.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Consolidated HC Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities using saturated steam as the sterilizing agent. They are equipped with factory programmed cycles. The factory programmed cycles, the chamber and overall dimensions, and the number of standard test items per AAMI ST8:2013 R2018 for all models of the HC Series steam sterilizers are provided below. Table 1: Validated…

Read the full statement (480 words)

The Consolidated HC Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities using saturated steam as the sterilizing agent. They are equipped with factory programmed cycles. The factory programmed cycles, the chamber and overall dimensions, and the number of standard test items per AAMI ST8:2013 R2018 for all models of the HC Series steam sterilizers are provided below.

Table 1: Validated Sterilization Cycles

Cycle Type | Sterilization Temperature | Sterilization Time | Dry Time | Load Type

Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Fabric Packs

Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double Wrapped Instrument Trays Maximum weight per tray: 25lbs (11.3 kg)

Gravity – IUSS | 270°F (132°C) | 3 minutes | 1 minutes | Single Unwrapped Tray Maximum weight per tray: 25lbs (11.3 kg)

Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double Wrapped Instrument Trays Maximum weight per tray: 25lbs (11.3 kg)

Pre-Vacuum – IUSS | 270°F (132°C) | 4 minutes | 1 minutes | Single Unwrapped Tray Maximum weight per tray: 25lbs (11.3 kg)

Pre-Vacuum | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs

Pre-Vacuum | 270°F (132°C) | 4 minutes | 30 minutes | Double Wrapped Instrument Trays Maximum weight per tray: 25lbs (11.3 kg)

Pre-Vacuum | 275°F (135°C) | 3 minutes | 30 minutes | Double Wrapped Instrument Trays Maximum weight per tray: 25lbs (11.3 kg)

Pre-Vacuum | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack and Single Unwrapped Tray Maximum weight per tray: 25lbs (11.3 kg)

Liquids | 250°F (121°C) | 30 minutes | N/A | 1L Liquid Volumes

Bowie Dick Test | 273°F (134°C) | 3.5 minutes | 0 minutes | Bowie Dick Test Pack

Vacuum Leak Test | 270°F (132°C) | N/A | N/A | Empty Chamber

Warm Up | 270°F (132°C) | 3 minutes | 1 minutes | Empty Chamber

Table 2: Load Size

Model | Sterilizer Chamber Size | Double Wrapped Instrument Trays | Fabric Packs2 | Number of Liquid Containers3

26AV-HC | 26” X 26” X 39” | 9, maximum weight 25lb each | 12 | 70, 1 Liter Volume Per Container

26BV-HC | 26” X 26” X 49” | 12, maximum weight 25lb each | 16 | 90, 1 Liter Volume Per Container

3AV-HC | 20” X 20” X 38” | 3, maximum weight 25lb each | 9 | 28, 1 Liter Volume Per Container

1 For instrument trays, the maximum weight per tray is 25lbs (11.3 kg).

2 The weight of each fabric pack should be limited to approximately 3.2 lbs (1.4 kg) with a density of 11.3 lb/ft3 (181 kg/m3) or less.

3 For liquid containers, the maximum volume of liquid per container is 1,000 ml. Please note, liquid loads processed in the sterilizer are inappropriate for direct patient contact.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220736: Consolidated HC Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220736

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026