510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Visera Hysterovideoscope Olympus HYF Type V

K221557510(k) TraditionalCleared Sep 2, 202294-day reviewClass II21 CFR 884.1690Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corporation · HIH Hysteroscope (And Accessories)

Cleared on Sep 2, 2022 as substantially equivalent to K022445 (Olympus Optical Co., Ltd., 2003) after a 94-day review. Its family tree reaches back 2 generations to 1989. It has been recalled 2 times.

Decided
Sep 2, 2022
Review
94 days
Class
Class II
Product code
HIH
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to K891451 (1989). No later device cites it as a predicate yet.

  1. Root · 2 generations backK891451 Olympus Hyf-P Flexible HysterofiberscopeOlympus Corp. · 1989 · Recalled
  2. 1 generation backK022445 Visera Hysterovideoscope Olympus Hyf Type VOlympus Optical Co., Ltd. · 2003 · Recalled
  3. This deviceK221557 Visera Hysterovideoscope Olympus HYF Type VOlympus Medical Systems Corporation · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K891451Olympus Hyf-P Flexible HysterofiberscopeOlympus Corp.19891
−1K022445Visera Hysterovideoscope Olympus Hyf Type VOlympus Optical Co., Ltd.20031
0 (this device)K221557Visera Hysterovideoscope Olympus HYF Type VOlympus Medical Systems Corporation2022—
FIG. 1 — Lineage of K221557, 1989–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HIH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HIH device

  1. #1 K232003 Endoscopic Video Image Processor, Single-Use…score 16.1 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022445Visera Hysterovideoscope Olympus Hyf Type V
Summary page 5
…Predicate Device 510(k) No. VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V K022445 The predicate device has not been subject to a design-related recall. Product Description This instrument has been designed to…
Olympus Optical Co., Ltd.20031

Also cleared against K022445: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 25, 2023Class IIZ-0336-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Jul 20, 2023Class IIZ-2480-2023Reprocessing ControlsIFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 33 years and passes through 3 companies; a new generation was cleared every 19.6 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • 1 other device also cites its predicate K022445.

How does its category compare?

All HIH devices

HIH Hysteroscope (And Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · HIH

3in 2025

2007: 12008: 22009: 42010: 42011: 32012: 42013: 52014: 22015: 52016: 32017: 22018: 82019: 142020: 72021: 42022: 22023: 42024: 62025: 32026 (partial year): 7
2007high 14 (2019) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HIH

111in 2025

2009: 130 days median2010: 109 days median2011: 118 days median2012: 25 days median2013: 144 days median2015: 52 days median2016: 106 days median2018: 135 days median2019: 91 days median2020: 88 days median2021: 194 days median2023: 200 days median2024: 148 days median2025: 111 days median2026 (partial year): 168 days median
2007high 200 (2023) · low 25 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HIH

3so far in 2026, the most since 2007

2019: 12023: 22026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HIH

1,698in 2025

2007: 72008: 122009: 172010: 312011: 592012: 782013: 1112014: 1252015: 2112016: 1722017: 3362018: 1822019: 2202020: 2532021: 1932022: 2892023: 1,4492024: 1,6112025: 1,698
2007high 1,698 (2025) · low 7 (2007)2025

MAUDE reports naming product code HIH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—07
20082—012
20094130017
20104109031
20113118059
2012425078
201351440111
20142—0125
20155520211
201631060172
20172—0336
201881350182
201914911220
20207880253
202141940193
20222—0289
2023420021,449
2024614801,611
2025311101,698
202671683—

What is it indicated for?

VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis within the uterus. Do not use the instrument for any purpose other than its intended use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K221557: Visera Hysterovideoscope Olympus HYF Type V.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K221557

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026