510(k) Report

510(k) · Cleared

Cleared

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate

K221626510(k) AbbreviatedCleared Dec 7, 2022184-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jr Engineering & Medical Technologies (M) Sdn. Bhd. · LZA Polymer Patient Examination Glove

Cleared on Dec 7, 2022 as substantially equivalent to K202622 (O&M Halyard, Inc., 2021) after a 184-day review. Its family tree reaches back 7 generations to 2004.

Decided
Dec 7, 2022
Review
184 days
Class
Class I
Product code
LZA
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K042805 (2004). 8 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 7 generations backK042805 Powder Free Nitrile Patient Examination Gloves, Blue Colored…Perusahaan Getah Asas Sdn. Bhd. · 2004
  2. 6 generations backK081089 Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam…Kimberly-Clark Corp. · 2008
  3. 5 generations backK141982 Dermagrip Powder Free Blue Nitrile Patient Examination Gloves…Wrp Asia Pacific Sdn. Bhd. · 2015
  4. 4 generations backK163146 POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy…Hebei Hongsen Plastics Technology Co, Ltd. · 2017
  5. 3 generations backK172525 Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for…Central Medicare Sdn Bhd · 2017
  6. 2 generations backK180646 Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with…Halyard Health · 2018
  7. 1 generation backK202622 Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves…O&M Halyard, Inc. · 2021
  8. This deviceK221626 Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use…Jr Engineering & Medical Technologies (M) Sdn. Bhd. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K042805Powder Free Nitrile Patient Examination Gloves, Blue Colored Non-SterilePerusahaan Getah Asas Sdn. Bhd.2004—
−6K081089Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam Gloves with…Kimberly-Clark Corp.2008—
−5K141982Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile…Wrp Asia Pacific Sdn. Bhd.2015—
−4K163146POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy DrugsHebei Hongsen Plastics Technology Co, Ltd.2017—
−3K172525Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with…Central Medicare Sdn Bhd2017—
−2K180646Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with…Halyard Health2018—
−1K202622Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for…O&M Halyard, Inc.2021—
0 (this device)K221626Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with…Jr Engineering & Medical Technologies (M) Sdn. Bhd.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
FIG. 1 — Lineage of K221626, 2004–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZA device

  1. #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K202622Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with…
Summary page 5
…Tested for use with Chemotherapy Drugs and Fentanyl Citrate 510( K) Number K202622 Regulatory Class 1 Product code LZC, QDO D. DESCRIPTION OF THEDEVICE: The…
O&M Halyard, Inc.2021—

Also cleared against K202622: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 8 companies; a new generation was cleared every 2.0 years on median.
  • None of the 8 ancestors has a recall.
  • 1 other device also cites its predicate K202622.

How does its category compare?

All LZA devices

LZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · LZA

24in 2025

2007: 212008: 182009: 282010: 312011: 262012: 332013: 202014: 272015: 202016: 302017: 442018: 222019: 252020: 302021: 1392022: 1462023: 622024: 302025: 242026 (partial year): 15
2007high 146 (2022) · low 18 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZA

110in 2025

2007: 95 days median2008: 97 days median2009: 112 days median2010: 97 days median2011: 122 days median2012: 110 days median2013: 130 days median2014: 197 days median2015: 112 days median2016: 171 days median2017: 124 days median2018: 178 days median2019: 212 days median2020: 101 days median2021: 111 days median2022: 111 days median2023: 92 days median2024: 89 days median2025: 110 days median2026 (partial year): 103 days median
2007high 212 (2019) · low 89 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZA

3so far in 2026, the most since 2007

2015: 12016: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZA

73in 2025

2007: 72008: 22009: 122010: 122011: 92012: 112013: 102014: 82015: 122016: 252017: 42018: 32019: 212020: 42021: 202022: 192023: 242024: 762025: 73
2007high 76 (2024) · low 2 (2008)2025

MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007219507
2008189702
200928112012
20103197012
20112612209
201233110011
201320130010
20142719708
201520112112
201630171125
20174412404
20182217803
201925212021
20203010104
2021139111020
2022146111019
20236292024
20243089076
202524110073
2026151033—

What is it indicated for?

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by…

Read the full statement (366 words)

Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate is a disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner.

Additionally, the gloves were tested for use with chemotherapy drugs and Fentanyl Citrate in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

The tested chemotherapy drugs and their breakthrough detection times are as follows:

Bendamustine (5 mg/ml) >240 Minutes

Bleomycin (15 mg/ml) >240 Minutes

Busulfan (6 mg/ml) >240 Minutes

Carfilzomib (2 mg/ml) >240 Minutes

Cetuximab (2 mg/ml) >240 Minutes

Cladribine (1 mg/ml) >240 Minutes

Cytarabine (100 mg/ml) >240 Minutes

Daunorubicin HCl (5 mg/ml) >240 Minutes

Decitabine (5 mg/ml) >240 Minutes

Docetaxel (10 mg/ml) >240 Minutes

Epirubicin HCl (2 mg/ml) >240 Minutes

Fludarabine (25 mg/ml) >240 Minutes

Fulvestrant (50 mg/ml) >240 Minutes

Gemcitabine (38 mg/ml) >240 Minutes

Idarubicin HCl (1 mg/ml) >240 Minutes

Irinotecan (20 mg/ml) >240 Minutes

Mechlorethamine (1 mg/ml) >240 Minutes

Melphalan (5 mg/ml) >240 Minutes

Oxaliplatin (5 mg/ml) >240 Minutes

Paraplatin (10 mg/ml) >240 Minutes

Pemetrexed (25 mg/ml) >240 Minutes

Raltitrexed (0.5 mg/ml) >240 Minutes

Rituximab (10 mg/ml) >240 Minutes

Topotecan (1 mg/ml) >240 Minutes

Triclosan (2 mg/ml) >240 Minutes

Trisenox (1 mg/ml) >240 Minutes

Velcade (Bortezomib) (1 mg/ml) >240 Minutes

Vidaza (Azacytidine) (25 mg/ml) >240 Minutes

Vinblastine (1 mg/ml) >240 Minutes

Vinorelbine (10 mg/ml) >240 Minutes

Carmustine (BCNU) (3.3 mg/ml) 35.0 Minutes

Carboplatin (10 mg/ml) >240 Minutes

Cisplatin (1 mg/ml) >240 Minutes

Cyclophosphamide (Cytoxan) (20 mg/ml) >240 Minutes

Dacarbazine (10.0 mg/ml) >240 Minutes

Doxorubicin HCl (2 mg/ml) >240 Minutes

Etoposide (20 mg/ml) >240 Minutes

Fluorouracil (50 mg/ml) >240 Minutes

Ifosfamide (50 mg/ml) >240 Minutes

Methotrexate (25 mg/ml) >240 Minutes

Mitomycin C (0.5 mg/ml) >240 Minutes

Mitoxantrone (2 mg/ml) >240 Minutes

Paclitaxel (6 mg/ml) >240 Minutes

Thiotepa (10 mg/ml) 64.9 Minutes

Vincristine Sulfate (1 mg/ml) >240 Minutes

The following hazardous drugs (opioid) and concentration had no breakthrough detected up to 240 minutes:

Fentanyl Citrate Injection (100 mcg/2ml)

Please note that the following drugs have low permeation times:

Carmustine (BCNU) (3.3 mg/ml) 35.0 Minutes

Thiotepa (10 mg/ml) 64.9 Minutes

Warning: Do not use with Carmustine or Thiotepa.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K221626: Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs and Fentanyl Citrate.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K221626

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026