510(k) · Cleared
ClearedHalyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and and Fentanyl Citrate
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
O&M Halyard, Inc. · LZA Polymer Patient Examination Glove
Cleared on Jan 7, 2021 as substantially equivalent to K180646 (Halyard Health, 2018) after a 119-day review. Its family tree reaches back 6 generations to 2004. 2 later devices descend from it.
- Decided
- Jan 7, 2021
- Review
- 119 days
- Class
- Class I
- Product code
- LZA
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K042805 (2004). 7 ancestors in all. 2 later devices descend from it, 2 citing it directly.
- Root · 6 generations backK042805 Powder Free Nitrile Patient Examination Gloves, Blue Colored…Perusahaan Getah Asas Sdn. Bhd. · 2004
- 5 generations backK081089 Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam…Kimberly-Clark Corp. · 2008
- 4 generations backK141982 Dermagrip Powder Free Blue Nitrile Patient Examination Gloves…Wrp Asia Pacific Sdn. Bhd. · 2015
- 3 generations backK163146 POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy…Hebei Hongsen Plastics Technology Co, Ltd. · 2017
- 2 generations backK172525 Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for…Central Medicare Sdn Bhd · 2017
- 1 generation backK180646 Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with…Halyard Health · 2018
- This deviceK202622 Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves…O&M Halyard, Inc. · 2021
- Descendant · cites it directly · 0 of its ownK221626 Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use…Jr Engineering & Medical Technologies (M) Sdn. Bhd. · 2022
- Descendant · cites it directly · 0 of its ownK213937 Halyard Lavender Nitrile, Low Dermatitis Potential, Powder-Free Exam…O&M Halyard, Inc. · 2022
Most influential descendants
Showing 2 of 2 · See all- K221626Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy…Jr Engineering & Medical Technologies (M) Sdn. Bhd. · 2022 · cites it directly0descendants
- K213937Halyard Lavender Nitrile, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for…O&M Halyard, Inc. · 2022 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K042805 | Powder Free Nitrile Patient Examination Gloves, Blue Colored Non-Sterile | Perusahaan Getah Asas Sdn. Bhd. | 2004 | — |
| −5 | K081089 | Kimberly-Clark Sterling Nitrile & Nitrile-Xtra Powder-Free Exam Gloves with… | Kimberly-Clark Corp. | 2008 | — |
| −4 | K141982 | Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile… | Wrp Asia Pacific Sdn. Bhd. | 2015 | — |
| −3 | K163146 | POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs | Hebei Hongsen Plastics Technology Co, Ltd. | 2017 | — |
| −2 | K172525 | Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with… | Central Medicare Sdn Bhd | 2017 | — |
| −1 | K180646 | Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with… | Halyard Health | 2018 | — |
| 0 (this device) | K202622 | Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for… | O&M Halyard, Inc. | 2021 | — |
| +1 | K221626 | Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with… | Jr Engineering & Medical Technologies (M) Sdn. Bhd. | 2022 | — |
| +1 | K213937 | Halyard Lavender Nitrile, Low Dermatitis Potential, Powder-Free Exam Gloves… | O&M Halyard, Inc. | 2022 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2022 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LZA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LZA device
- #1 K192333 Blue Nitrile Examination Gloves Powder Freescore 71.0 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K180646 | Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with Chemotherapy DrugsSummary page 5…protection glove Predicate Device Halyard Lavender, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs (K180646) Subject Device Description Halyard Lavender Nitrile, Powder-Free Exam Gloves Tested for Use with… | 2018 |
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 17 years and passes through 7 companies; a new generation was cleared every 2.5 years on median.
- None of the 7 ancestors has a recall.
- Of its descendants, K221626 (Jr Engineering & Medical Technologies (M) Sdn. Bhd., 2022) has gone furthest, with 0 descendants of its own.
- 2 other devices also cite its predicate K180646.
How does its category compare?
All LZA devicesLZA Polymer Patient Examination Glove: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · LZA
24in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LZA
110in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LZA
3so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LZA
73in 2025
MAUDE reports naming product code LZA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 21 | 95 | 0 | 7 |
| 2008 | 18 | 97 | 0 | 2 |
| 2009 | 28 | 112 | 0 | 12 |
| 2010 | 31 | 97 | 0 | 12 |
| 2011 | 26 | 122 | 0 | 9 |
| 2012 | 33 | 110 | 0 | 11 |
| 2013 | 20 | 130 | 0 | 10 |
| 2014 | 27 | 197 | 0 | 8 |
| 2015 | 20 | 112 | 1 | 12 |
| 2016 | 30 | 171 | 1 | 25 |
| 2017 | 44 | 124 | 0 | 4 |
| 2018 | 22 | 178 | 0 | 3 |
| 2019 | 25 | 212 | 0 | 21 |
| 2020 | 30 | 101 | 0 | 4 |
| 2021 | 139 | 111 | 0 | 20 |
| 2022 | 146 | 111 | 0 | 19 |
| 2023 | 62 | 92 | 0 | 24 |
| 2024 | 30 | 89 | 0 | 76 |
| 2025 | 24 | 110 | 0 | 73 |
| 2026 | 15 | 103 | 3 | — |
What is it indicated for?
The Halyard Lavender Nitrile, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM -D6978-05 : The following chemotherapy drugs and concentration had NO breakthrough detected up…
Read the full statement (302 words)
The Halyard Lavender Nitrile, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM -D6978-05 :
The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:
Azacitidine (25 mg/ml)
Bendamustine HCl (5 mg/ml)
Bleomycin Sulfate (15 mg/ml)
Bortezomib (1 mg/ml)
Busulfan (6 mg/ml)
Capecitabine (26 mg/ml)
Carboplatin (10 mg/ml)
Carlzomib (2 mg/ml)
Cetuximab (2 mg/ml)
Chloroquine (50 mg/ml)
Cisplatin (1 mg/ml)
Cladribine (1 mg/ml)
Cyclophosphamide (20 mg/ml)
Cyclosporin A (100 mg/ml)
Cytarabine (Cytosine) (100 mg/ml)
Cytovene (Ganciclovir) (10 mg/ml)
Dacarbazine (DTIC) (10 mg/ml)
Dactinomycin (0.5 mg/ml)
Daunorubicin HCl (5 mg/ml)
Decitabine (5 mg/ml)
Docetaxel (10 mg/ml)
Doxorubicin HCl (2 mg/ml)
Epirubicin HCl (Ellence) (2 mg/ml)
Etoposide (Toposar) (20 mg/ml)
Fludarabine (25 mg/ml)
5-Fluorouracil (50 mg/ml)
Fulvestrant (50 mg/ml)
Gemcitabine (38 mg/ml)
Idarubicin (1 mg/ml)
Ifosfamide (50 mg/ml)
Irinotecan HCl (20 mg/ml)
Leuprolide Acetate Salt (5 mg/ml)
Mechlorethamine HCl (1 mg/ml)
Melphalan (5 mg/ml)
Methotrexate (25 mg/ml)
Mitomycin C (0.5 mg/ml)
Mitoxantrone (2 mg/ml)
Oxaliplatin (5 mg/ml)
Paclitaxel (6 mg/ml)
Pemetrexed (25 mg/ml)
Raltitrexed (0.5 mg/ml)
Retrovir (10 mg/ml)
Rituximab (10 mg/ml)
Temsirolimus (25 mg/ml)
Topotecan HCl (1 mg/ml)
Triclosan (2 mg/ml)
Trisenox (1 mg/ml)
Vinblastine Sulfate (1 mg/ml)
Vincristine (1 mg/ml)
Vinorelbine (10 mg/ml)
Zoledronic Acid (0.8 mg/ml)
The following chemotherapy drugs and concentration showed breakthrough detected in less than 60 minutes:
Carmustine (3.3 mg/ml) No breakthrough up to 0.3 minutes.
Thiotepa (10 mg/ml) No breakthrough up to 30.9 minutes.
The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:
Fentanyl Citrate Injection (100 mcg/2 ml)
Warning: Not for Use With: Carmustine, ThioTEPA
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K202622 · printed Nov 6, 2026
Cite this page
510(k) Report. “K202622: Halyard Lavender, Low Dermatitis Potential, Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs and and Fentanyl Citrate.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K202622
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026