510(k) Report

510(k) · Cleared

Recalled · Class I · 2025

AutoPulse NXT Resuscitation System

K221700510(k) TraditionalCleared Mar 9, 2023269-day reviewClass II21 CFR 870.5200Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zoll Circulation, Inc. · DRM Compressor, Cardiac, External

Cleared on Mar 9, 2023 as substantially equivalent to K112998 (Zoll Circulation, 2012) after a 269-day review. Its family tree reaches back 6 generations to 1985. 1 of the 9 devices in its family, counting itself, has a Class I recall. It has been recalled once, including a Class I recall.

Decided
Mar 9, 2023
Review
269 days
Class
Class II
Product code
DRM
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 6 generations to K851139 (1985). 8 ancestors in all. No later device cites it as a predicate yet. 1 of the 9 devices in this family, counting this one, has a Class I recall.

  1. Root · 6 generations backK851139 Thumper Cardiopulmonary Resuscitator 1005Michigan Instruments, Inc. · 1985
  2. 5 generations backK022345 Autopulse Resuscitation SystemRevivant Corp. · 2002 · Recalled
  3. 4 generations backK032852 Modification to: Autopulse Resuscitation System Model 100Revivant Corp. · 2003 · Recalled
  4. 3 generations backK040453 Autopulse Resuscitation System, Model 100Revivant Corp. · 2004 · Recalled
  5. 2 generations backK063602 Autopulse Resuscitation System MODEL100Zoll Circulation · 2006 · Recalled
  6. 1 generation backK112998 Autopulse Resuscitation System Model 1000Zoll Circulation · 2012 · Recalled
  7. This deviceK221700 AutoPulse NXT Resuscitation SystemZoll Circulation, Inc. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K851139Thumper Cardiopulmonary Resuscitator 1005Michigan Instruments, Inc.1985—
−5K022345Autopulse Resuscitation SystemRevivant Corp.20022
−4K032852Modification to: Autopulse Resuscitation System Model 100Revivant Corp.20032
−3K040453Autopulse Resuscitation System, Model 100Revivant Corp.20042
−2K063602Autopulse Resuscitation System MODEL100Zoll Circulation20061
−1K112998Autopulse Resuscitation System Model 1000Zoll Circulation20121
0 (this device)K221700AutoPulse NXT Resuscitation SystemZoll Circulation, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200810
FIG. 1 — Lineage of K221700, 1985–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRM device

  1. #1 K211289 RMU-2000 Automated Chest Compression Systemscore -9.3 · 9 generations · Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112998Autopulse Resuscitation System Model 1000
Summary page 4
…Classification Name: External Cardiac Compressor (21 CFR 870.5200) Product Code: DRM 510(k) Number: K112998 (Cleared: March 15, 2012) Device Description The AutoPulse® NXT Resuscitation System (also referred to as the AutoPulse® Model…
Zoll Circulation20121

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 4, 2025Class IZ-1533-2025Process change controlAutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 38 years and passes through 4 companies; a new generation was cleared every 5.2 years on median.
  • 7 of the 8 ancestors have a recall of any class.

How does its category compare?

All DRM devices

DRM Compressor, Cardiac, External: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DRM

1in 2023

2008: 22009: 12010: 12012: 12014: 12016: 22018: 12021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRM

Too few DRM 510(k)s in any year since 2007 to give a median.

Recalls per year · DRM

1in 2022

2008: 12011: 12012: 12022: 12026 (partial year): 1
2007high 1 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRM

690in 2025

2007: 82008: 132009: 92010: 72011: 162012: 4832013: 1,2682014: 6642015: 6742016: 6392017: 6792018: 7832019: 7062020: 8442021: 7552022: 7302023: 7652024: 7102025: 690
2007high 1,268 (2013) · low 7 (2010)2025

MAUDE reports naming product code DRM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—08
20082—113
20091—09
20101—07
20110—116
20121—1483
20130—01,268
20141—0664
20150—0674
20162—0639
20170—0679
20181—0783
20190—0706
20200—0844
20211—0755
20220—1730
20231—0765
20240—0710
20250—0690
20260—1—

What is it indicated for?

The AutoPulse NXT Resuscitation System is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by a lack of spontaneous breathing and pulse. The AutoPulse NXT System must be used only in cases where chest compressions are likely to help the patient.

The AutoPulse NXT System is intended for use as an adjunct to manual CPR when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K221700: AutoPulse NXT Resuscitation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K221700

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026